A US medical device sponsor cannot hand FDA a cybersecurity narrative written to its own outline anymore — since 1 October 2023, nearly every 510(k) has had to be built inside FDA's own electronic Submission Template And Resource (eSTAR), a structured, dynamic PDF that collects cybersecurity documentation in the fields FDA's reviewers already expect, not wherever a sponsor's own document template happens to put it.
For a connected ("cyber") device under Section 524B of the FD&C Act, that structure is not cosmetic: eSTAR's own completeness logic, not a human reviewer's first read, decides whether a submission with thin cybersecurity content even reaches review.
Key takeaways
- FDA's premarket cybersecurity duties for a cyber device come from section 524B of the FD&C Act, which took effect on 29 March 2023.
- The eSTAR 7.1 template answers this in its own help dialog.
- The cybersecurity section of the nIVD eSTAR 7.1 template sits under IMDRF table-of-contents heading CH3.05.05.11.
- Section 524B(b)(3) requires an SBOM "including commercial, open-source, and off-the-shelf software components".
What is eSTAR, in FDA's own words?
FDA's eSTAR Program page describes the tool directly: eSTAR is "an interactive PDF form that guides applicants through the process of preparing a comprehensive medical device submission." Fields unlock or close based on earlier answers, so declaring that a device has software, connectivity or cybersecurity risk opens the sections that content requires. FDA's FAQ on the same page is explicit that "only relevant sections and requests will be displayed based on the applicant's answers."
On the page's most recent update (content current as of 21 September 2026), three templates are current: the Non-In Vitro Diagnostic (nIVD) eSTAR version 7.1, the In Vitro Diagnostic (IVD) eSTAR version 7.1, and the Early Submission Requests eSTAR (PreSTAR) version 3.1 for Q-Submissions, IDEs and 513(g) requests. A major version follows a final guidance or rule; a sponsor on an older major version "may receive additional information requests related to the major changes." Download the current template from the eSTAR Program page rather than reusing a saved copy from an earlier submission.
When did eSTAR become mandatory, submission type by submission type?
eSTAR's mandatory-use dates arrived separately for each submission type, through separate final guidance documents:
| Submission type | Status | Source |
|---|---|---|
| 510(k) | Mandatory since 1 October 2023 | FDA final guidance "Electronic Submission Template for Medical Device 510(k) Submissions" (Docket FDA-2021-D-0872) |
| De Novo request | Mandatory since 1 October 2025 | FDA final guidance "Electronic Submission Template for Medical Device De Novo Requests" (Docket FDA-2023-D-3788, August 2024) |
| PMA application / supplement | Voluntary | FDA's eSTAR Program page |
| Q-Submission, IDE, 513(g) | Voluntary (PreSTAR) | FDA's eSTAR Program page |
A sponsor preparing a PMA today is not required to use eSTAR, but FDA has already moved two submission types from voluntary to mandatory eSTAR use on its own timetable, so a PMA sponsor should not assume voluntary status is permanent.
How does eSTAR enforce completeness without a classic RTA letter?
FDA's premarket cybersecurity duties for a cyber device come from section 524B of the FD&C Act, which took effect on 29 March 2023. FDA backed that statute with a dedicated refuse-to-accept (RTA) policy, announced in the Federal Register on 30 March 2023 (88 FR 19148), under which FDA said it would generally not issue RTA decisions based solely on 524B information before 1 October 2023. See FDA premarket cybersecurity for medical devices for how that policy works.
eSTAR adds a separate completeness gate. The eSTAR Program page states that "eSTAR submissions are not anticipated to undergo a Refuse to Accept (RTA) process". Instead, FDA runs a virus scan and a technical screening. The CDRH Portal will not accept an eSTAR whose status line still reads "eSTAR INCOMPLETE". If a submitted eSTAR fails screening (for example, attachments that are not relevant to the question, or an inaccurate answer), FDA emails a notice and the submission sits on a technical screening hold for up to 180 days; if no replacement eSTAR arrives in that window, FDA "will consider the submission withdrawn."
FDA's cybersecurity FAQ page ties this directly to cybersecurity: "an eSTAR will be put on a Technical Screening hold if it does not contain accurate responses and relevant attachments in the Cybersecurity section of eSTAR." A thin cybersecurity section stops the clock before substantive review starts.
Which devices have to complete the cybersecurity section?
The eSTAR 7.1 template answers this in its own help dialog. It asks for cybersecurity and interoperability documentation only if the device or system meets any of these conditions:
- can communicate with other devices or systems;
- has a network or server connection, or the potential for one that should be disabled;
- has the potential for unused ports for network connections;
- uses any form of wireless communication (Bluetooth, Wi-Fi, cellular, RF, inductive);
- has a USB port or physical media access (memory card, JTAG);
- allows software or firmware downloads, such as updates or patches;
- allows cloud storage or cloud services.
The template then states: "Providing documentation regarding the Cybersecurity is required if the submission is a 510(k), PMA, De Novo, or HDE, and any of the above conditions are met."
Separately, section 524B(c) defines a cyber device as one that "(1) includes software validated, installed, or authorized by the sponsor as a device or in a device; (2) has the ability to connect to the internet; and (3) contains any such technological characteristics … that could be vulnerable to cybersecurity threats." FDA's February 2026 guidance (Section VII.B) reads this broadly. Firmware and programmable logic count as software. A USB, Ethernet or serial connector, Bluetooth Low Energy or magnetic inductive link counts as the "ability to connect to the internet". A device serviced over USB is "considered to have the ability to connect to the internet". Most connected firmware-based devices are therefore cyber devices.
What does the eSTAR 7.1 cybersecurity section ask for?
The cybersecurity section of the nIVD eSTAR 7.1 template sits under IMDRF table-of-contents heading CH3.05.05.11. It mostly asks for attachments, and several questions ask you to cite the page number where an item is covered. Each block maps to the Cybersecurity in Medical Devices guidance (February 2026, Docket FDA-2021-D-1158), which superseded the 27 June 2025 version:
| eSTAR 7.1 block | What the template asks you to attach or answer | Guidance / statute behind it |
|---|---|---|
| General information | Whether the submission is a modification and whether cybersecurity is affected; the answer decides which questions follow | Section VII.D |
| Risk management: report | A security risk management report from "a separate, parallel, and interconnected" process, with a traceability matrix of risks to controls | Section V.A (AAMI TIR57, ANSI/AAMI SW96) |
| Risk management: threat model | A threat model covering "all the end-to-end elements of the system", the methodology used (e.g. STRIDE, attack trees), and whether it includes the architecture views | Sections V.A.1, V.B.2 |
| Risk management: risk assessment | The assessment, with page citations for methodology and acceptance criteria; a yes/no on using exploitability instead of probability (CVSS, IEEE 11073-40101, NIST SP 800-30 given as examples) | Section V.A.2 |
| Risk management: SBOM | The SBOM; support level and end-of-support date per component; supported operating systems; a safety and security assessment of known vulnerabilities | 524B(b)(3); Section V.A.4 |
| Vulnerabilities with uncontrolled risk | Whether any critical vulnerabilities could cause uncontrolled risks now, and any remediated since the last authorization | 524B(b)(2); Section VII.D |
| Unresolved anomalies, metrics, controls | Anomaly security assessment; cybersecurity metrics (or a justification); security controls A–H with page citations | Sections V.A.5, V.A.6, Appendix 1 |
| Architecture views and testing | Global system, multi-patient harm, updatability/patchability and security use case views; test reports | Sections V.B.2, V.C |
| Labeling and management plan | Page citations for each cybersecurity labeling element; the cybersecurity management plan | 524B(b)(1); Sections VI.A, VI.B, VII.C.1 |
Two prompts catch teams out. The operating system field warns that listing an OS that is "no longer supported or nearing end of support … will generally be considered an inaccurate response". And the metrics prompt asks for three measures: the percentage of identified vulnerabilities updated or patched, time from identification to patch, and time from patch availability to deployment in the field.
What belongs in the SBOM attachment?
Section 524B(b)(3) requires an SBOM "including commercial, open-source, and off-the-shelf software components". The guidance (Section V.A.4.b) and the eSTAR help text add the detail:
- Machine-readable format. eSTAR names SPDX or CycloneDX, serialized as JSON, XML or YAML.
- NTIA baseline attributes. The October 2021 NTIA framing document lists eight: author name, timestamp, supplier name, component name, version string, component hash, unique identifier and relationship. See our SBOM minimum elements guide.
- Two FDA additions per component: the level of support (actively maintained, no longer maintained, abandoned) and the end-of-support date. These may sit in the SBOM or in an addendum.
- Known vulnerabilities, "including those identified in CISA's Known Exploited Vulnerabilities Catalog", how each was discovered, a safety and security risk assessment, and the risk controls that address it. Our KEV and EPSS prioritization guide covers the KEV side.
- A justification wherever an element cannot be determined.
For devices where you only hold a binary, see firmware SBOM, and IMDRF N73's SBOM principles for the international view.
What cybersecurity testing evidence does FDA expect?
Section V.C of the guidance, repeated in the eSTAR help text, lists four groups:
Security requirements:evidence each design input was implemented, plus boundary analysis and its rationale. Threat mitigation:tests showing each risk control works against the threat model and the four architecture views. Vulnerability testing(the guidance cites ANSI/ISA 62443-4-1; see IEC 62443-4-1): abuse and misuse cases, malformed inputs, robustness, fuzz testing, attack surface analysis, vulnerability chaining, closed-box known-vulnerability scanning, software composition analysis of binary executable files, and static and dynamic analysis including checks for hard-coded or default credentials. Penetration testing, with a report stating tester independence and expertise, scope, duration, methods and findings.
FDA asks for the original report from any third-party tester, your assessment of every finding, and tool names, versions and settings. For findings you defer, the submission should carry the remediation timeline and say whether devices shipped in the meantime will receive the fix.
Worked example: how would a small BLE device fill in the section?
Synthetic example, labelled as such: the device, counts and CVEs below are invented for illustration and describe no real product or customer.
A company submits a Traditional 510(k) for a Bluetooth Low Energy dosing-log cap. It pairs with a phone app that syncs to a cloud service, and its firmware updates over BLE through the app.
TriggerWireless communication, firmware downloads and cloud services all appear on the eSTAR list, so the section opens and is required. Under Section VII.B the device is a cyber device. SBOMA CycloneDX JSON SBOM lists 64 components: the RTOS, BLE stack, crypto library, bootloader and the app's open-source libraries. An addendum gives support level and end-of-support date for each. Known vulnerabilitiesMatching component versions against NVD finds 7 CVEs; 1 is in CISA KEV. The attachment states how they were found (binary composition analysis plus database matching), assesses each for patient harm and exploitability, and names the controls. The KEV entry is patched in the submitted firmware. Threat model and viewsSTRIDE across device, app, cloud and update path. One global system view, one multi-patient harm view (the cloud reaches every device), one updatability view (signed images over BLE) and security use case views for pairing, log sync and firmware update. TestingBLE GATT fuzzing, a third-party penetration test of app, cloud API and device (original report attached), and static analysis showing no hard-coded credentials. MetricsThe product has no field history, so the metrics prompt gets a justification and the cybersecurity management plan states the targets the company will measure against. LabelingThe instructions list BLE as the only interface and describe the update procedure.
What changes for a modification to a cleared device?
The first eSTAR cybersecurity question asks whether the submission is a modification and whether cybersecurity is affected. FDA gives examples: changes to "authentication or encryption algorithms, new connectivity features, or changing software update process/mechanisms" may affect cybersecurity and need the full documentation. For changes unlikely to affect it (materials, sterilization, an algorithm change without architecture or connectivity change), Section VII.D allows a lighter set: the 524B(b)(1) plan or a reference to the earlier one, a statement on current critical vulnerabilities with uncontrolled risk and on any remediated since the last authorization, and an SBOM. Section 524B still applies, so a premarket submission for a modified cyber device always carries an SBOM.
Why does the quality system, and not only the device, matter here?
eSTAR's cybersecurity content sits downstream of the quality system obligations that FDA's QMSR (21 CFR Part 820) governs. The February 2026 guidance points to ISO 13485 subclauses 7.3.6 and 7.3.7 for security verification and validation, and to 7.3.10 and 4.2.3 for keeping the SBOM in the design and device files. It also recommends a security risk assessment separate from, and alongside, the ISO 14971 safety assessment (see ANSI/AAMI SW96).
What are the common pitfalls?
- Attaching a generic document to every slot. Attachments that are not relevant to the question are an explicit trigger for a technical screening hold.
- A human-readable component list instead of an SBOM. eSTAR asks for a machine-readable SPDX or CycloneDX file.
- No support or end-of-support data. FDA asks for both per component, or a justification.
- Listing an outdated operating system. eSTAR warns that an OS that is no longer supported or nearing end of support is generally an inaccurate response.
- Missing architecture views. Appendix 4 of the guidance recommends them for every premarket submission of a device with cybersecurity risk, IDEs included.
- A plan without a patch cycle. The 524B(b)(1) plan should state the regular update cycle with its justification and the out-of-cycle route for critical vulnerabilities; see FDA postmarket cybersecurity.
For a threat-modeling method that fits the end-to-end scope eSTAR asks for, see OWASP threat modeling.
Can AI help fill in eSTAR's cybersecurity fields?
SoftWSP's firmware analysis extracts the software components from a firmware image into a CycloneDX 1.6 JSON SBOM and matches them against NVD, CISA KEV, FIRST EPSS and CISA ICS advisories. That covers the machine-readable SBOM and the known-vulnerability list, including KEV entries, that the eSTAR SBOM block asks for. AI then drafts a plain-language report that cites the component, version and CVE, KEV or EPSS record behind each finding, for an engineer to review before anything goes into the submission.
It does not supply support levels or end-of-support dates, write VEX statements, analyse reachability, draft the threat model, the architecture views, the security-control justifications or the test reports, or decide whether an eSTAR field is complete. Those stay with the sponsor. See device & firmware security, and PSIRT as a service for the coordinated-disclosure side of the 524B(b)(1) plan.
This guide is general information, not legal or regulatory advice. It reflects the nIVD eSTAR version 7.1 template and the February 2026 guidance as published on FDA's website on 13 November 2026; FDA revises both on its own schedule.
Put this guide to work
Frequently asked questions
What is FDA's eSTAR?
The electronic Submission Template And Resource (eSTAR) is, in FDA's own words on its eSTAR Program page, "an interactive PDF form that guides applicants through the process of preparing a comprehensive medical device submission." It is a structured, dynamic PDF that builds a 510(k), De Novo, or (voluntarily) PMA submission section by section, rather than a free-form document a sponsor assembles from scratch.
When did eSTAR become mandatory, and for which submission types?
For 510(k)s: FDA's final guidance, "Electronic Submission Template for Medical Device 510(k) Submissions" (Docket FDA-2021-D-0872, issued October 2023), states "As of October 1, 2023, FDA will require that 510(k) electronic submissions be provided as described in this guidance." For De Novo requests: FDA's final guidance "Electronic Submission Template for Medical Device De Novo Requests" (Docket FDA-2023-D-3788, issued August 2024) states "As of October 1, 2025, FDA will require that De Novo Request electronic submissions be provided as described in this guidance." Per FDA's eSTAR Program page, use remains voluntary for PMA applications and supplements, Q-Submissions, IDEs, and 513(g) requests.
Does an incomplete eSTAR cybersecurity section trigger a refuse-to-accept decision?
Not in the same sense as a classic refuse-to-accept (RTA) letter. FDA's eSTAR Program page states that "eSTAR submissions are not anticipated to undergo a Refuse to Accept (RTA) process" at all — instead, FDA runs a technical screening process, and if eSTAR is incomplete, FDA notifies the submitter by email and the submission is placed, and remains, on a hold of up to 180 days until a complete replacement eSTAR is submitted. Functionally this still blocks review start on a submission missing required cybersecurity content, but the mechanism is a technical-screening hold built into the template's own logic, not the separate statutory RTA policy FDA issued for Section 524B cyber devices.
What cybersecurity content does eSTAR actually ask for?
eSTAR's software/cybersecurity area is built to collect the documentation FDA's current premarket cybersecurity guidance describes, including a software bill of materials, a threat model, a cybersecurity risk assessment, security control justifications, cybersecurity testing evidence, and a plan to monitor and address postmarket vulnerabilities. FDA's guidance for that content is "Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions" (Docket FDA-2021-D-1158), reissued February 2026; confirm the current eSTAR template's exact field list and version number on FDA's own eSTAR Program page before relying on a specific field name, since FDA updates the template periodically.
Is eSTAR mandatory for a PMA submission?
No. Per FDA's eSTAR Program page, PMA applications and supplements are listed under voluntary, not mandatory, eSTAR use, alongside Q-Submissions, IDEs, and 513(g) requests. Confirm current status directly on FDA's eSTAR Program page before assuming otherwise, since FDA has moved other submission types from voluntary to mandatory eSTAR use before (510(k) in 2023, De Novo in 2025) and could do the same for PMA.
Sources
- FDA: eSTAR Program (current template versions, mandatory vs. voluntary use, technical screening and hold process)
- FDA: Electronic Submission Template for Medical Device 510(k) Submissions (final guidance, Docket FDA-2021-D-0872)
- FDA: Electronic Submission Template for Medical Device De Novo Requests (final guidance, Docket FDA-2023-D-3788)
- FDA: Cybersecurity in Medical Devices – Quality Management System Considerations and Content of Premarket Submissions (guidance, Docket FDA-2021-D-1158)
- Federal Register 88 FR 19148: Refuse to Accept Policy for Cyber Devices Under Section 524B (Docket FDA-2023-D-1030)
- FDA: Cybersecurity in Medical Devices guidance, February 2026 (full text PDF: Sections V.A.4, V.C, VII and Appendices 1 and 4)
- FDA: Cybersecurity resources for medical device manufacturers — guidance, resources and FAQs (section 524B dates, eSTAR technical screening hold)
- 21 U.S. Code § 360n-2 (FD&C Act section 524B), Ensuring cybersecurity of devices — Cornell Legal Information Institute
- NTIA: Framing Software Component Transparency — Establishing a Common SBOM, Second Edition (21 October 2021)