FDA finalized its guidance on the FDA Predetermined Change Control Plan (PCCP) — "Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions" — on 4 December 2024, then reissued it with updates on 18 August 2025, both under docket FDA-2022-D-2628; the plan itself lets a manufacturer of an AI-enabled medical device pre-authorize a defined set of future AI modifications as part of its original marketing submission, so later updates within that plan's scope don't each need a brand-new submission.
Key takeaways
- The PCCP is a statutory tool, not only a guidance concept.
- The guidance applies across the De Novo, premarket approval (PMA), and premarket notification (510(k)) pathways — whichever route a given AI-enabled device software function uses to reach market.
- On 24 October 2023, FDA, Health Canada and the UK MHRA jointly published five guiding principles for PCCPs for machine learning-enabled medical devices.
- FDA has started to write PCCPs into classification regulations.
What problem is a PCCP meant to solve?
AI-enabled device development is iterative by nature — a manufacturer may want to retrain or tune a model's performance after it's already on the market, which under FDA's normal process would mean a new marketing submission for each meaningful change. FDA's guidance states a PCCP is intended to let manufacturers plan for such modifications while still providing a reasonable assurance of safety and effectiveness, "without necessitating additional marketing submissions for implementing each modification described in the PCCP" — authorizing the change process itself, up front, rather than authorizing each individual future change one at a time.
What does the law actually say about PCCPs?
The PCCP is a statutory tool, not only a guidance concept. Section 3308 of the Food and Drug Omnibus Reform Act of 2022 (Pub. L. 117-328, enacted 29 December 2022) added section 515C to the FD&C Act, codified at 21 U.S.C. 360e-4, as the Federal Register notice for the final guidance recounts. The statute has four parts a regulatory team should know by heart:
| Provision | What it says |
|---|---|
| 515C(a), approved (PMA) devices | No PMA supplement is needed for a change consistent with an approved PCCP; FDA may approve the plan if the device remains safe and effective without any change |
| 515C(b), cleared (510(k)) devices | No new 510(k) is needed for a change consistent with a cleared PCCP; the device must remain safe and effective without the change and substantially equivalent to its predicate |
| 515C(a)(3) and (b)(3), scope | FDA may require the plan to include labeling for safe use as the device changes, notification requirements if the device does not function as intended, and performance requirements for changes made under the plan |
| 515C(c), predicates | FDA does not compare a new device against a version changed under a PCCP; only the version originally cleared or approved can serve as a predicate |
Two consequences follow. First, the original, unchanged device has to stand on its own evidence — a PCCP cannot be used to clear a device that only becomes adequate after its first modification. Second, a competitor (or the manufacturer's own next product) cannot cite the improved, post-PCCP performance as its substantial-equivalence baseline.
What must a PCCP actually contain?
FDA's guidance recommends a PCCP comprise at least three sections:
| Section | What it covers |
|---|---|
| Description of modifications | The specific AI-DSF changes the manufacturer intends to make under the plan |
| Modification protocol | The methodology used to develop, validate and implement each described modification |
| Impact assessment | An assessment of the impact each modification could have on the device's safety and effectiveness |
A PCCP is reviewed and authorized as part of the same marketing submission as the device itself — FDA does not authorize a PCCP separately during the pre-submission (Q-Submission) process, though manufacturers are encouraged to use a Q-Submission to get FDA feedback on a proposed PCCP before filing it formally.
Which submission pathways and devices does it apply to?
The guidance applies across the De Novo, premarket approval (PMA), and premarket notification (510(k)) pathways — whichever route a given AI-enabled device software function uses to reach market. It replaces an April 2023 draft guidance that covered only machine-learning-enabled devices; the final version broadens scope to all AI-enabled device software functions (AI-DSFs), aligning FDA's "artificial intelligence" and "machine learning" definitions with Executive Order 14110 (30 October 2023) and FDA's own September 2024 Digital Health and Artificial Intelligence Glossary. FDA is explicit, though, that most PCCPs it has reviewed in practice remain for the machine-learning subset of AI-DSFs specifically — devices that benefit from preapproved modifications tied to continued learning on new data.
The statute itself is not limited to AI. On 22 August 2024 FDA published a separate draft guidance, "Predetermined Change Control Plans for Medical Devices" (docket FDA-2024-D-2338), which describes a PCCP as "the documentation describing what modifications will be made to a device and how the modifications will be assessed" for devices in general. That document is a draft; for AI-enabled functions, the final AI-DSF guidance above is the one to follow.
What makes a PCCP reviewable?
On 24 October 2023, FDA, Health Canada and the UK MHRA jointly published five guiding principles for PCCPs for machine learning-enabled medical devices. They are not binding requirements, but they read as a reviewer's checklist:
| Principle | What the regulators describe | What to show in the plan |
|---|---|---|
| Focused and bounded | Specific changes, limited to modifications within the device's intended use, with methods to verify and validate them and mechanisms to "detect and revert or stop implementation of a change that fails to meet specified performance criteria" | A closed list of modifications, acceptance criteria per modification, and a rollback path |
| Risk based | Design and implementation driven by risk management, so that "individual and cumulative changes remain appropriate over time" | Hazard analysis per modification and for the combined effect of several, tied to the ISO 14971 risk file |
| Evidence based | Methods and metrics that are scientifically and clinically justified, and evidence of benefits and risks before and after each change | Pre-specified test methods and datasets, not chosen after results are known |
| Transparent | Information for users on data used, testing, and device characteristics before and after each change, plus monitoring of performance deviations | Labeling text, version history and the field-monitoring plan |
| Total product lifecycle | Use of existing regulatory, quality and risk measures across the lifecycle to monitor, report and respond to safety concerns | Links into the QMS, complaint handling and postmarket surveillance |
What does a codified PCCP device type look like?
FDA has started to write PCCPs into classification regulations. A final order published on 17 June 2026 created 21 CFR 892.2055, "radiological machine learning-based quantitative imaging software with predetermined change control plan", a class II device type whose design specifications include modifications made under an established PCCP (the order states the classification was applicable from 24 February 2023). Its special controls show in concrete terms what FDA expects:
- Training and test data: a detailed description of training data, including annotation methods and important cohorts, and a test dataset "independent from data used in training/development".
- Modification methodology: documentation of the intended modifications and of the methodology for developing, verifying and validating them against the plan's performance requirements.
- Risk management: identification and assessment of the risks of the modifications, with mitigations.
- PCCP labeling: a statement that the device has a PCCP; a description of modifications implemented, with current performance, inputs, validation requirements and evidence; and "a version history, a description of how device modification(s) will be implemented, and a description of how users will be informed".
Later 510(k) devices of that type must meet the same special controls, so the codified text is a useful template even for devices of other types.
Worked example: how would a connected ECG patch scope its PCCP?
Synthetic example. The device, numbers and decisions below are invented for illustration and reflect SoftWSP's reading of the statute and the guiding principles, not an FDA determination.
A manufacturer is preparing a 510(k) for a wearable ECG patch with an on-device arrhythmia classifier and a Bluetooth link to a phone app. The regulatory team sorts each change it expects in the next three years:
| Expected change | In the PCCP? | Reason |
|---|---|---|
| Retrain the atrial-fibrillation classifier on additional labelled recordings, same inputs and outputs | Yes | Within intended use; can be tested against a fixed, independent test set with pre-set sensitivity and specificity floors |
| Accept recordings from a second patch hardware revision with the same sampling rate and lead placement | Yes, with its own acceptance tests | An input modification that can be bounded and validated |
| Add detection of a new arrhythmia type | No | A new output and indication; outside "focused and bounded", so it needs its own submission |
| Upgrade the inference runtime and Bluetooth stack to fix CVEs | Not an AI modification | Ordinary software change, handled through the change-control and Section 524B patch processes |
For the two in-scope modifications, the modification protocol names the data sources and labelling method, the frozen test set, the pass/fail thresholds, how the update is rolled out and reverted, and how users see the new version in labeling. The impact assessment covers each modification alone and both together.
Why the runtime line matters. Teams often try to fold every future firmware change into the PCCP. A dependency upgrade does not change the AI-DSF's intended behaviour, and listing it there blurs the "focused and bounded" scope reviewers look for. Keep it in the cybersecurity package, where FDA expects patching to happen anyway.
How does a PCCP fit alongside FDA's other cybersecurity submission duties?
A PCCP addresses intended AI modifications; it does not substitute for the separate cybersecurity content Section 524B of the FD&C Act already requires in a cyber device's premarket submission — an SBOM, a vulnerability-monitoring plan, and a secure-update process, covered in our guide to FDA premarket cybersecurity for medical devices. An AI-enabled connected device is very often a cyber device under that separate definition too, which means a manufacturer can need both a PCCP (for the AI modification pathway) and the full Section 524B cybersecurity package (SBOM, disclosure process, patch plan) in the same submission, each reviewed against its own guidance. FDA's older postmarket cybersecurity guidance and its QMSR under 21 CFR Part 820 continue to apply after clearance regardless of whether a PCCP is in place.
The statutory text of section 524B (21 U.S.C. 360n-2) makes the overlap concrete: the sponsor must provide postmarket updates and patches "on a reasonably justified regular cycle" for known unacceptable vulnerabilities and "as soon as possible out of cycle" for critical ones, plus a software bill of materials covering commercial, open-source and off-the-shelf components. Every model update shipped under a PCCP is also a new software release, so its firmware SBOM and vulnerability review need to move with it.
What does labeling need to say if a device has an authorized PCCP?
FDA's guidance reiterates that labeling for an AI/ML-enabled device with an authorized PCCP should clearly state that the device incorporates machine learning and operates under an authorized PCCP — so a clinician or user understands the device may receive software updates, implemented automatically ("continuous learning"), manually, or a mix of both, that change its performance, inputs or use after the device first reaches the field. This is a continuing-transparency obligation that runs alongside the PCCP's technical content, not a one-time disclosure made only at clearance.
What are the common pitfalls?
- An open-ended modification list. "Performance improvements" without named inputs, outputs and limits is not "focused and bounded". Each modification needs its own acceptance criteria.
- Test data that leaks from training. The 21 CFR 892.2055 special controls require a test set independent from training and development data; reusing or refreshing it with training data defeats the protocol.
- No revert path. The guiding principles expect mechanisms to stop or roll back a change that misses its criteria. Document how a fielded device returns to the previous model.
- Ignoring cumulative effects. Two modifications that each pass may interact. Assess them together.
- Relying on the modified device as a predicate. Section 515C(c) rules this out for the next 510(k).
- Letting labeling drift. The version history and "how users will be informed" text must be updated with every implemented modification.
- Assuming the PCCP covers the cybersecurity file. It does not; SBOM, monitoring and patching stay under Section 524B.
How does this compare outside the US?
Manufacturers selling an AI-enabled medical device into both the US and EU face a structurally similar but legally separate problem on the EU side: the EU MDR's own cybersecurity expectations under Annex I, plus the EU AI Act's cybersecurity-specific requirements for high-risk AI systems covered in our guide to the EU AI Act's cybersecurity requirements. Neither EU regime has a direct PCCP equivalent today — an intended post-market AI modification in the EU still runs through MDR's own significant-change assessment rather than a pre-authorized change-control envelope — so a PCCP-covered US submission does not automatically clear the equivalent EU bar for the same intended modification.
Where does SoftWSP's own AI use fit alongside an AI-enabled device's PCCP?
A PCCP governs intended changes to a device's own AI; SoftWSP's AI plays no part in that — it does not design, validate or assess the device's machine-learning modifications, and has no role in the modification protocol or impact assessment a PCCP itself describes. What SoftWSP's pipeline supports is the separate Section 524B cybersecurity package many AI-enabled connected devices also need: extracting the firmware's software components and matching them against known CVE, KEV and EPSS data, with AI-assisted analysis drafting the resulting finding narrative for an engineer to review before submission. That package documents the device's cybersecurity posture; it says nothing about the AI model's own clinical performance.
In practice, SoftWSP extracts an SBOM from the firmware image, writes it as CycloneDX 1.6 JSON, and matches its components against NVD, CISA KEV, FIRST EPSS and CISA's CSAF ICS advisories; the AI-drafted report cites the record behind each finding for an engineer to check. Each new release is scanned when its image is uploaded (earlier SBOMs are not re-matched automatically), and watchlist alerts fire on a CVE ID or vendor/product you follow. See device & firmware security services and PSIRT as a service.
This guide is general information, not legal or regulatory advice. It does not constitute FDA regulatory or legal counsel; confirm the current guidance document and submission requirements directly with FDA before relying on them.
Put this guide to work
Frequently asked questions
What is an FDA Predetermined Change Control Plan (PCCP)?
A PCCP is a plan, submitted and authorized as part of an AI-enabled device's marketing submission, that describes modifications the manufacturer intends to make to the device's AI after clearance, authorization or approval, how those modifications will be developed and validated, and an assessment of their impact — so FDA can authorize a defined envelope of future changes up front instead of requiring a brand-new submission each time the manufacturer wants to update the AI. FDA's guidance requires a PCCP to contain, at minimum, three sections: a description of modifications, a modification protocol, and an impact assessment.
When did FDA finalize this guidance, and has it changed since?
FDA finalized 'Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions' on 4 December 2024 (Federal Register notice 89 FR 96259, document 2024-28361, docket FDA-2022-D-2628), replacing an April 2023 draft that covered only machine-learning-enabled devices. FDA reissued the same guidance with updates on 18 August 2025, still under docket FDA-2022-D-2628; that August 2025 version is the current final guidance.
Which FDA submission pathways does a PCCP apply to?
FDA's guidance applies PCCPs across the De Novo, premarket approval (PMA) and premarket notification (510(k)) pathways — any pathway an AI-enabled device software function could use to reach market. A PCCP is reviewed and authorized as part of that same marketing submission; FDA does not authorize a PCCP during the separate pre-submission (Q-Submission) process, though manufacturers are encouraged to use Q-Submissions to get FDA feedback on a proposed PCCP before filing.
Does the guidance cover all AI-enabled devices, or only machine-learning ones?
The final guidance broadened scope from the 2023 draft: it covers all AI-enabled device software functions (AI-DSFs), not only the machine-learning subset the draft focused on exclusively, and FDA updated its 'artificial intelligence' and 'machine learning' definitions to align with Executive Order 14110 (30 October 2023) and FDA's own September 2024 Digital Health and Artificial Intelligence Glossary. FDA notes most PCCPs it has actually reviewed remain for machine-learning devices — specifically ones that benefit from preapproved modifications based on continued learning — so the broadened AI-DSF scope is forward-looking as much as it reflects current submissions.
What must a device's labeling say if it has an authorized PCCP?
FDA's guidance reiterates that labeling for an AI/ML-enabled device with an authorized PCCP should clearly state that the device incorporates machine learning and has an authorized PCCP, so users understand the device may receive software updates that change its performance, inputs or use after it's already in the field — a continuing-transparency duty that sits alongside, not instead of, the PCCP's technical content.
Sources
- FDA: Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions (guidance page)
- Federal Register 89 FR 96259 (document 2024-28361): Availability notice, docket FDA-2022-D-2628 (4 December 2024)
- FDA: Digital Health and Artificial Intelligence Glossary (September 2024)
- Ropes & Gray: FDA Finalizes Guidance on Predetermined Change Control Plans for AI-Enabled Medical Device Software
- Foley & Lardner: AI-Enabled Device Software Functions — FDA's Final Guidance for Predetermined Change Control Plans
- 21 U.S.C. 360e-4: Predetermined change control plans for devices (FD&C Act section 515C, Cornell LII)
- 21 U.S.C. 360n-2: Ensuring cybersecurity of devices (FD&C Act section 524B, Cornell LII)
- FDA, Health Canada, MHRA: Predetermined change control plans for machine learning-enabled medical devices: guiding principles (24 October 2023)
- Federal Register 2024-18828: Predetermined Change Control Plans for Medical Devices, draft guidance (22 August 2024)
- Federal Register 2026-12166: Classification of radiological machine learning-based quantitative imaging software with PCCP, 21 CFR 892.2055 (17 June 2026)