CVE-2021-23233 is a missing authentication check in the Fresenius Kabi Agilia Connect infusion system: Agilia Link+ v3.0 D15 and earlier serves sensitive endpoints to requests without a session, so an unauthenticated user on the network can trigger critical actions or change configuration. NVD scores it 9.8, CISA 7.3. The fix is Link+ v3.0 D16 or later.
Vulnerability summary
CVSS base score
9.8
Critical severity
EPSS score (FIRST)
1.0%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2021-23233 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| Fresenius Kabi Agilia Link+ (rack for Agilia infusion pumps) | cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:3.0:d15:*:*:*:*:*:* | v3.0 D15 and earlier (CISA ICSMA-21-355-01, NVD); fixed in Link+ v3.0 D16 or later |
| Fresenius Kabi Agilia Connect pump Wi-Fi module | cpe:2.3:o:fresenius-kabi:agilia_connect_firmware:*:*:*:*:*:*:*:* | D25 and earlier (CISA, NVD); fixed in D29 or later |
| Fresenius Kabi Agilia Partner maintenance software | cpe:2.3:a:fresenius-kabi:agilia_partner_maintenance_software:*:*:*:*:*:*:*:* | 3.3.0 and earlier (CISA, NVD); fixed in v3.3.2 or later |
| Fresenius Kabi Vigilant Software Suite (Centerium, Insight, MasterMed) | cpe:2.3:a:fresenius-kabi:vigilant_centerium:1.0:*:*:*:*:*:*:* | v1.0 (CISA advisory page, NVD); fixed in VSS v1.0.3 or later |
What does this mean for affected devices?
The CVE record from CISA says: "Sensitive endpoints in Fresenius Kabi Agilia Link+ v3.0 and prior can be accessed without any authentication information such as the session cookie. An attacker can send requests to sensitive endpoints as an unauthenticated user to perform critical actions or modify critical configuration parameters." CISA classes it as improper access control, CWE-284. NVD published the record on 2022-01-21 and maps it to CWE-798 instead (NVD, CVE record).
Why does NVD score it 9.8 and CISA 7.3?
Both agree on the attack path: network, low complexity, no privileges, no user interaction. They differ on impact. NVD's Primary score is 9.8 with CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, treating confidentiality, integrity and availability as fully lost. CISA, the CNA, scores 7.3 with CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L, rating each impact as low. CISA's risk note says the flaws sit in system accessories, which is the reading behind the lower impact values.
Which Agilia products are affected and what fixes it?
CISA's ICSMA-21-355-01, released on 2021-12-21 and updated on 2022-01-27 (Update A), covers the Agilia Connect pump Wi-Fi module D25 and earlier, Agilia Link+ v3.0 D15 and earlier, Agilia Partner maintenance software 3.3.0 and earlier, and, on the CISA advisory page, Vigilant Software Suite v1.0. Fresenius Kabi released Link+ v3.0 D16 or later, VSS v1.0.3 or later, Wi-Fi module D29 or later and Agilia Partner v3.3.2 or later, and began telling customers in April 2021. Around 1,200 early Link+ units need a hardware change to run D16; until they are replaced, Fresenius Kabi points users to CISA's network measures: keep the devices off the internet, behind firewalls and separate from the business network, with remote access over an up-to-date VPN. That replacement case is a legacy-device problem of the kind covered in the IMDRF legacy device guide, and hospitals can track it under the FDA postmarket guidance and HHS 405(d) practices. For the code-level fix, see the authentication bypass fix guide and, for NVD's mapping, the hard-coded credentials fix guide.
The same advisory lists 12 sibling CVEs found by ERNW researchers in the BSI ManiMed project, among them CVE-2021-23236 (resource exhaustion, 7.5), CVE-2020-35340 (7.5), CVE-2021-41835, CVE-2021-23196 and CVE-2021-43355 (each 7.3), CVE-2021-31562 (6.5), CVE-2021-44464 (hard-coded credentials, 6.3) and CVE-2021-33848 (cross-site scripting, 5.4). Other infusion-system pages: CVE-2023-30560 (BD Alaris PCU) and CVE-2023-0888 (B. Braun Battery Pack SP with Wi-Fi).
Is CVE-2021-23233 exploited?
CVE-2021-23233 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.01006, the 61.97th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).


Evidence
- NVD (read 2026-11-27): status Modified; published 2022-01-21; NVD Primary CVSS v3.1 9.8 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H; ICS-CERT Secondary 7.3 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L; CWE-798 (NVD Primary), CWE-284 (ICS-CERT); CPEs Link+ firmware up to 3.0 D15, Agilia Connect firmware up to D25, Agilia Partner maintenance software up to 3.3.0, Vigilant Centerium/Insight/MasterMed 1.0. — NVD
- CISA ICSMA-21-355-01 CSAF (released 2021-12-21, Update A 2022-01-27, read 2026-11-27): Agilia Connect Wi-Fi module <= D25, Agilia Partner maintenance software <= 3.3.0, Agilia Link+ <= 3.0 D15; CVE-2021-23233 CWE-284, 7.3; fixes Link+ v3.0 D16+, VSS v1.0.3+, Wi-Fi module D29+, Agilia Partner v3.3.2+; about 1,200 early Link+ devices need a hardware change; 13 CVEs in total; Healthcare and Public Health, worldwide, HQ Germany; reported by ERNW via BSI ManiMed. — CISA ICSMA-21-355-01
- CISA advisory page (read 2026-11-27): title Fresenius Kabi Agilia Connect Infusion System (Update A), also lists Vigilant Software Suite v1.0; CVE-2021-23233 7.3. — CISA
- CVE record (read 2026-11-27): state PUBLISHED, CNA ICS-CERT, published 2022-01-21; affected Agilia Link+ below 3.0; CNA CVSS v3.1 7.3; solution lists the same fixed versions. — CVE.org
- Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
- EPSS 0.01006, percentile 0.61969, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which Agilia Link+ version fixes CVE-2021-23233?
Link+ v3.0 D16 or later. In the same advisory Fresenius Kabi also lists VSS v1.0.3 or later, Agilia Connect Wi-Fi module D29 or later and Agilia Partner v3.3.2 or later (CISA ICSMA-21-355-01).
Why do NVD and CISA give different scores?
Both use network access with no privileges or user interaction. NVD rates confidentiality, integrity and availability impact high (9.8); CISA rates each low (7.3), since the flaws are in system accessories rather than the pumps.
What if a Link+ unit cannot take D16 firmware?
Fresenius Kabi says about 1,200 early Link+ devices need a hardware change. Until they are replaced, keep them off the internet, behind a firewall and separate from the business network, with remote access only over an up-to-date VPN.
Is CVE-2021-23233 in the CISA KEV catalog?
No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.01006 (61.97th percentile) in the 2026-11-26 dataset.
Questions about this page?