CVE-2023-30560 is an improper authentication flaw in the BD Alaris Point-of-Care Unit (PCU) Model 8015, the control unit of the BD Alaris infusion system: anyone with a physical connection to the PCU can modify its configuration without authenticating. NVD and BD both score it 6.8. CISA says the flaw is still present in PCU version 12.3.1, so BD relies on compensating controls.
Vulnerability summary
CVSS base score
6.8
Medium severity
EPSS score (FIRST)
0.3%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2023-30560 · Source: NVD
Affected components
| Component | CPE | Version range |
|---|---|---|
| BD Alaris Point-of-Care Unit (PCU) Model 8015 | cpe:2.3:o:bd:alaris_8015_pcu_firmware:*:*:*:*:*:*:*:* | 12.1.3 and prior (BD CNA record, NVD CPE up to 12.1.3 inclusive). CISA ICSMA-23-194-01 Update A (2023-10-26) says the flaw is still present in PCU version 12.3.1; no fixed version is named there |
What does this mean for affected devices?
BD (Becton, Dickinson and Company) is the CNA, and its description is short: "The configuration from the PCU can be modified without authentication using physical connection to the PCU." The CVE record lists the BD Alaris Point-of-Care Unit (PCU) Model 8015, version 12.1.3 and prior, and was published on 2023-07-13 alongside CISA's ICS medical advisory ICSMA-23-194-01, which covers eight CVEs in the BD Alaris System with Guardrails Suite MX (CVE record, CISA CSAF ICSMA-23-194-01).
Why does a physical-access flaw score 6.8?
NVD's Primary score is CVSS v3.1 6.8 (CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H), identical to BD's, and the CISA CSAF file uses the same values as a v3.0 vector. The physical vector keeps the number down; the impact metrics are all high because the PCU is the unit that drives the attached infusion modules, so a changed configuration can affect how therapy is delivered. CISA's SSVC entry on the CVE record reads exploitation none, automatable no, technical impact total. Our CVSS v4 vs v3.1 guide uses this CVE to show how CVSS v4.0 records harm to a patient through the Subsequent System and Safety metrics.
Is there a fixed version?
No fixed version is named as of 2026-11-23. CISA's Update A (2023-10-26) says BD released BD Alaris System 12.3 and updated the PCU Model 8015 to version 12.3.1, and then states: "The following CVEs are still present on the BD Alaris PCU Model 8015 version 12.3.1: CVE-2023-30559, CVE-2023-30560, and CVE-2023-30561." BD still recommends moving to System 12.3 where regulatory authorization allows, because it remediates other CVEs in the same advisory; customers arrange that through their BD Account Executive.
The CSAF vulnerability note names only the PCU, but its product status block for this CVE also lists the Guardrails Editor, Systems Manager, CQI Reporter and Calculation Services. Read the note and the CVE record, which name only the PCU, as the scope of the flaw.
What should hospitals do in the meantime?
From the mitigations BD gives in the CISA advisory, the ones that address a physical configuration change are:
- Follow chapter 1 of the Alaris System Maintenance Software User Manual to enable an authentication challenge password for network configuration changes, and manage those credentials as described in its Network Settings section.
- Periodically inspect BD Alaris System components to confirm they run the correct software versions.
- Inspect the system before use for signs of tampering, as described in BD's FIPS 140-2 compliance instructions.
- Apply physical protection and access control practices, which BD ties to NIST SP 800-171 Rev. 2.
The other controls in the advisory are network ones: the PCU needs only DNS, DHCP and the Systems Manager on port 3613, accepts no unsolicited inbound traffic, and BD recommends a separate VLAN and MAC filtering. The authentication bypass fix guide covers the weakness class for device makers, and the FDA postmarket cybersecurity guide covers how manufacturers are expected to handle flaws that stay open. Related advisories: CVE-2023-0888, a flaw in another infusion pump's communication module, and CVE-2018-10933, an authentication bypass in an embedded SSH library.
CVE-2023-30560 is not in the CISA KEV catalog as of 2026-11-23 (catalog version 2026.11.20, 1,790 entries), and CISA's advisory reports no known public exploitation specifically targeting these vulnerabilities. FIRST EPSS gives it 0.00303, the 23.5th percentile, in the 2026-11-22 dataset (NVD, CISA KEV, FIRST EPSS).
Evidence
- NVD (read 2026-11-23): status Modified, published 2023-07-13; NVD Primary CVSS v3.1 6.8 AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H; BD Secondary 6.8 same vector; CWE-287 (NVD and BD); CPE cpe:2.3:o:bd:alaris_8015_pcu_firmware up to 12.1.3 inclusive. — NVD
- CVE record (checked 2026-11-23): state PUBLISHED, CNA BD, published 2023-07-13; product BD Alaris Point-of-Care Unit (PCU) Model 8015, versions up to 12.1.3; CVSS v3.1 6.8; CISA-ADP SSVC exploitation none, automatable no, technical impact total. — CVE.org
- CISA CSAF ICSMA-23-194-01, revision 2 Update A (2023-10-26), read 2026-11-23: CVE-2023-30560 CWE-287, CVSS v3.0 6.8; note names PCU Model 8015 v12.1.3 and prior; BD released Alaris System 12.3 and PCU 12.3.1; CVE-2023-30559, CVE-2023-30560 and CVE-2023-30561 still present in PCU 12.3.1; mitigations include authentication challenge password for network configuration changes, version inspection, tamper inspection, NIST SP 800-171 Rev. 2 practices, port 3613, VLAN and MAC filtering; no known public exploitation reported to CISA. — CISA CSAF ICSMA-23-194-01
- Not in the CISA KEV catalog, version 2026.11.20 (1,790 entries), checked 2026-11-23. — CISA KEV
- EPSS 0.00303, percentile 0.23474, dataset 2026-11-22, read 2026-11-23. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which BD Alaris version fixes CVE-2023-30560?
None is named as of 2026-11-23. CISA's ICSMA-23-194-01 Update A says CVE-2023-30560 is still present in BD Alaris PCU Model 8015 version 12.3.1, the PCU release that came with BD Alaris System 12.3. The CVE record lists version 12.1.3 and prior as affected.
What does an attacker need to exploit it?
A physical connection to the PCU. The CVE description says the configuration can then be modified without authentication. That is why the CVSS vector is physical, while confidentiality, integrity and availability impact are all rated high.
How can hospitals reduce the risk without a fix?
BD's mitigations in the CISA advisory include enabling the authentication challenge password for network configuration changes, periodically checking software versions, inspecting units for tampering before use, and physical access control. Network controls such as a separate VLAN and MAC filtering address the other CVEs in the advisory.
Is CVE-2023-30560 exploited in the wild?
It is not in the CISA KEV catalog as of 2026-11-23 (catalog version 2026.11.20, 1,790 entries), and CISA reported no known public exploitation specifically targeting it. Its EPSS score is 0.00303, the 23.5th percentile, in the 2026-11-22 dataset.
Questions about this page?