Security Advisory

CVE-2018-7518: BeaconMedaes TotalAlert Credential Exposure

Last updated:
  • CVE-2018-7518

CVE-2018-7518 is a credential exposure flaw in the TotalAlert web application of BeaconMedaes Scroll Medical Air Systems, equipment that supplies medical air in healthcare facilities. Anyone with network access to the integrated web server could retrieve default or user-defined credentials. NVD scores it 9.8, CISA 7.5. The fix is update 4107600010.24 or later.

Vulnerability summary

CVSS base score

9.8

Critical severity

EPSS score (FIRST)

1.3%

Probability of exploitation in the wild in the next 30 days

CISA KEV catalog

No

Not listed

Published

Vulnerability ID: CVE-2018-7518 · Source: NVD

CVSS v3.1 severity rating scale: None 0.0, Low 0.1 to 3.9, Medium 4.0 to 6.9, High 7.0 to 8.9, Critical 9.0 to 10.0
How a CVSS v3.1 base score maps to a severity rating. Source: FIRST CVSS v3.1 specification, section 5.

Affected components

ComponentCPEVersion range
BeaconMedaes TotalAlert Scroll Medical Air Systems (TotalAlert web application)cpe:2.3:o:beaconmedaes:scroll_medical_air_systems_firmware:*:*:*:*:*:*:*:*Software 4107600010.23 and prior (CISA ICSMA-18-144-01); NVD CPE: before 4107600010.23; fixed in update 4107600010.24 or later

What does this mean for affected devices?

The CVE record from CISA says: "In TotalAlert Web Application in BeaconMedaes Scroll Medical Air Systems prior to v4107600010.23, an attacker with network access to the integrated web server could retrieve default or user defined credentials stored and transmitted in an insecure manner." NVD and CISA map it to CWE-522, insufficiently protected credentials (NVD, CVE record).

Why does NVD score it 9.8 and CISA 7.5?

Both see the same attack path: network, low complexity, no privileges, no user interaction. NVD's Primary score is 9.8 with CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, treating stolen credentials as a route to full control; its CVSS v2 score is 5.0. CISA scores 7.5 with CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N, counting only the disclosure. CISA's risk note backs the narrower view: an attacker could view and possibly modify some device information and web application settings, patient health information is not in reach, and BeaconMedaes states that the system keeps delivering medical air in line with the NFPA 99 standard for healthcare facilities.

Which systems are affected and what fixes it?

CISA's ICSMA-18-144-01 lists TotalAlert Scroll Medical Air Systems running software 4107600010.23 and prior; the CVE text and the NVD CPE say "prior to 4107600010.23", so treat 4107600010.23 itself as affected. BeaconMedaes created update 4107600010.24 and asks users to install it or a later release, which customers request directly from the vendor. The same advisory lists two sibling CVEs in the web application: CVE-2018-7510, passwords kept in plain text in a file readable without authentication (7.5), and CVE-2018-7526, information reachable at a specific URL without login (5.3). Until the update is in, CISA's general advice applies: keep the TotalAlert web server off the internet, behind a firewall and separate from the business network. Hospitals can track the update under the FDA postmarket guidance and the HHS 405(d) practices.

Other hospital-equipment pages: CVE-2021-23233 (Fresenius Kabi Agilia Link+) and CVE-2020-25150 (B. Braun SpaceCom).

Is CVE-2018-7518 exploited?

CVE-2018-7518 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries), and CISA's advisory noted no known public exploit. FIRST EPSS gives it 0.01283, the 70.12th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).

Three signals on every advisory: CVSS says how severe the flaw is, EPSS gives the probability of exploitation in the wild in the next 30 days, and a CISA KEV listing means the flaw has been exploited in the wild
CVSS rates severity, EPSS estimates the likelihood of exploitation, and CISA KEV records exploitation seen in the wild. Source: FIRST EPSS; CISA KEV catalog.
How to fix this class of issueNo dedicated SoftWSP guide covers this weakness class; the fix is usually the vendor patch or firmware update referenced above. Our related guide explains how to plan and verify it: NIST SP 800-40 Rev. 4: enterprise patch management planning, explained.
From advisory to verified fix in four steps: read the advisory, find the affected component in the firmware SBOM, apply the fixed version or vendor mitigation, then re-scan the image to confirm the vulnerable version is gone
Read the advisory, find the component in your firmware SBOM, apply the fix, then re-scan to confirm it.

Evidence

  • NVD (read 2026-11-27): status Modified; published 2018-05-24; NVD Primary CVSS v3.0 9.8 CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H; NVD Primary CVSS v2 5.0 AV:N/AC:L/Au:N/C:P/I:N/A:N; CWE-522 (NVD Primary and ICS-CERT); CPE beaconmedaes:scroll_medical_air_systems_firmware before 4107600010.23. — NVD
  • CISA ICSMA-18-144-01 CSAF (read 2026-11-27): TotalAlert Scroll Medical Air Systems running software 4107600010.23 and prior; CISA CVSS v3 7.5 AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N; BeaconMedaes created update 4107600010.24; no patient health information exposed and NFPA 99 compliance unaffected per the vendor; Healthcare and Public Health, worldwide, HQ U.S.; no known public exploits. — CISA ICSMA-18-144-01
  • CISA advisory page (read 2026-11-27 via a page fetch): CVE-2018-7518 CWE-522 7.5; CVE-2018-7510 CWE-256 7.5 (plaintext passwords in a file accessible without authentication); CVE-2018-7526 CWE-284 5.3 (information reachable at a specific URL without authentication); update to 4107600010.24 or later, via BeaconMedaes. — CISA
  • CVE record (read 2026-11-27): state PUBLISHED, CNA ICS-CERT, published 2018-05-24; affected TotalAlert Scroll Medical Air Systems web application, all versions prior to 4107600010.23; no CNA CVSS in the record. — CVE.org
  • Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
  • EPSS 0.01283, percentile 0.70124, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories

Sources

  1. NVD
  2. CISA ICSMA-18-144-01: BeaconMedaes TotalAlert Scroll Medical Air Systems (CSAF)
  3. CISA ICSMA-18-144-01 advisory page
  4. CVE.org record CVE-2018-7518
  5. CISA KEV
  6. FIRST EPSS

Frequently asked questions

Which TotalAlert version fixes CVE-2018-7518?

Update 4107600010.24 or a later release, which BeaconMedaes supplies directly to customers. CISA ICSMA-18-144-01 lists software 4107600010.23 and prior as affected.

Can CVE-2018-7518 stop the supply of medical air?

CISA's advisory says an attacker could view and possibly modify some device information and web application settings, but not patient health information, and BeaconMedaes states the system keeps delivering medical air in line with NFPA 99.

Why do NVD and CISA give different scores?

Both rate the attack as network-based with no privileges or user interaction. NVD treats stolen credentials as leading to full loss of confidentiality, integrity and availability (9.8); CISA counts only the disclosure (7.5).

Is CVE-2018-7518 in the CISA KEV catalog?

No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.01283 (70.12th percentile) in the 2026-11-26 dataset.