Security Advisory

CVE-2020-25150: B. Braun SpaceCom Path Traversal to RCE

Last updated:
  • CVE-2020-25150

CVE-2020-25150 is a path traversal flaw in B. Braun's SpaceCom communication device (U61/L81 and earlier), the Battery Pack SP with Wi-Fi and the Data module compactplus A10/A11. A user with service privileges can place files outside the intended folder and run commands on the device. NVD scores it 8.8, CISA 7.6. B. Braun fixed it in U62/L82 and A12.

Vulnerability summary

CVSS base score

8.8

High severity

EPSS score (FIRST)

1.5%

Probability of exploitation in the wild in the next 30 days

CISA KEV catalog

No

Not listed

Published

Vulnerability ID: CVE-2020-25150 · Source: NVD

CVSS v3.1 severity rating scale: None 0.0, Low 0.1 to 3.9, Medium 4.0 to 6.9, High 7.0 to 8.9, Critical 9.0 to 10.0
How a CVSS v3.1 base score maps to a severity rating. Source: FIRST CVSS v3.1 specification, section 5.

Affected components

ComponentCPEVersion range
B. Braun SpaceCom (SpaceStation with SpaceCom 2)cpe:2.3:o:bbraun:spacecom:*:*:*:*:*:*:*:*U61 and earlier (United States), L81 and earlier (rest of world) (CISA ICSMA-20-296-02, B. Braun 10/2020, NVD); fixed in U62 / L82 or later
B. Braun Battery Pack SP with Wi-Fi—U61 and earlier, L81 and earlier (CISA, B. Braun, CVE.org); fixed in U62 / L82 or later
B. Braun Data module compactpluscpe:2.3:o:bbraun:datamodule_compactplus:a11:*:*:*:*:*:*:*A10 and A11, not distributed in the United States (CISA, B. Braun, NVD); fixed in A12 or later

What does this mean for affected devices?

The CVE record from CISA says: "A relative path traversal attack in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with service user privileges to upload arbitrary files. By uploading a specially crafted tar file an attacker can execute arbitrary commands." NVD maps it to CWE-22 and CISA to CWE-23, relative path traversal; NVD published the record on 2022-04-14 (NVD, CVE record).

Why does NVD score it 8.8 and CISA 7.6?

Both require a logged-in service user (PR:L) over the network with low complexity. NVD's Primary score is 8.8 with CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H, treating command execution as full loss of confidentiality, integrity and availability of the device. CISA and B. Braun score 7.6 with CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:L: high integrity impact, low confidentiality and availability impact.

Which B. Braun devices are affected and what fixes it?

CISA's ICSMA-20-296-02, released on 2020-10-22, and B. Braun's 10/2020 advisory list SpaceCom and Battery Pack SP with Wi-Fi U61 and earlier (United States) or L81 and earlier (rest of world), and Data module compactplus A10 and A11, which is not sold in the United States. B. Braun's fix is U62 or L82 or later and Data module compactplus A12 or later. The CISA advisory page (Update A) names the US and Canada builds: Battery Pack SP with Wi-Fi software 028U00093 or 054U00093 depending on serial number, and SpaceStation with SpaceCom 2 software 012U000093; outside those countries, 027L000093, 053L00093 and 011L000093. B. Braun states the flaws are in communication devices separated in hardware and software from the pumps, patient safety is not affected, and it had no reports of exploitation. Facilities should still keep pumps in a separate network zone, firewalled or on a VLAN, not reachable from the internet, with encrypted Wi-Fi and IDS or IPS.

The advisory has ten sibling CVEs from ERNW researchers in the BSI ManiMed project, including CVE-2020-25158 (cross-site scripting, 7.6), CVE-2020-25162 (XPath injection, 7.5), CVE-2020-25156 (active debug code, 7.2), CVE-2020-25166 (firmware signature check, 6.8) and CVE-2020-25168 (hard-coded credentials, 3.3). A later Battery Pack SP flaw is CVE-2023-0888; see also CVE-2021-23233 (Fresenius Kabi Agilia). For fixing the class, see the path traversal fix guide; for hospital-side handling, the FDA postmarket guidance and HHS 405(d) practices.

Is CVE-2020-25150 exploited?

CVE-2020-25150 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.01529, the 75.05th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).

Three signals on every advisory: CVSS says how severe the flaw is, EPSS gives the probability of exploitation in the wild in the next 30 days, and a CISA KEV listing means the flaw has been exploited in the wild
CVSS rates severity, EPSS estimates the likelihood of exploitation, and CISA KEV records exploitation seen in the wild. Source: FIRST EPSS; CISA KEV catalog.
How to fix this class of issueThis advisory's weakness class is Path traversal. Our practical guide covers how to detect, fix, verify and disclose it: How to fix path traversal in embedded firmware.
From advisory to verified fix in four steps: read the advisory, find the affected component in the firmware SBOM, apply the fixed version or vendor mitigation, then re-scan the image to confirm the vulnerable version is gone
Read the advisory, find the component in your firmware SBOM, apply the fix, then re-scan to confirm it.

Evidence

  • NVD (read 2026-11-27): status Modified; published 2022-04-14; NVD Primary CVSS v3.1 8.8 CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H; ICS-CERT Secondary 7.6 CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:L; CWE-22 (NVD Primary), CWE-23 (ICS-CERT); CPEs spacecom up to l81, datamodule_compactplus a10 and a11. — NVD
  • CISA ICSMA-20-296-02 CSAF (released 2020-10-22, read 2026-11-27): SpaceCom and Battery pack with Wi-Fi <= U61 / <= L81, Data module compactplus A10, A11; CVE-2020-25150 CWE-23 7.6; vendor fix builds 028U00093, 054U00093, 012U000093 (US/Canada) and 027L000093, 053L00093, 011L000093 (elsewhere); network zone, encrypted Wi-Fi, IDS/IPS; 11 CVEs; reported by ERNW via BSI ManiMed. — CISA ICSMA-20-296-02
  • B. Braun advisory 10/2020 (dated 2020-10-26, read 2026-11-27): CVE-2020-25150 relative path traversal CWE-23, 7.6; affected SpaceCom and Battery pack with WiFi U61/L81 and earlier, Data module compactplus A10, A11 (not distributed in the US); fix U62/L82 or later and A12 or later; patient safety not affected; no reports of exploitation. — B. Braun
  • CISA advisory page (read 2026-11-27): title includes Update A; same affected versions and fixed software builds; CVE-2020-25150 7.6. — CISA
  • CVE record (read 2026-11-27): state PUBLISHED, CNA ICS-CERT, published 2022-04-14; affected SpaceCom, Battery pack with Wi-Fi (<= U61, <= L81), Data module compactplus A10, A11; CNA CVSS v3.1 7.6; solution U62/L82, A12. — CVE.org
  • Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
  • EPSS 0.01529, percentile 0.75051, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories

Sources

  1. NVD
  2. CISA ICSMA-20-296-02: B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus (CSAF)
  3. CISA ICSMA-20-296-02 advisory page
  4. B. Braun security advisory 10/2020: SpaceCom, Battery Pack SP with WiFi, Data module compactplus
  5. CVE.org record CVE-2020-25150
  6. CISA KEV
  7. FIRST EPSS

Frequently asked questions

Which versions fix CVE-2020-25150?

B. Braun lists SpaceCom and Battery Pack SP with Wi-Fi U62 (United States) or L82 (rest of world) or later, and Data module compactplus A12 or later. CISA's Update A names the exact software builds per region and serial number range.

Can this flaw change a running infusion?

B. Braun states the flaws are in communication devices separated in hardware and software from the infusion pumps and that patient safety is not affected. It received no reports of exploitation.

Why is the NVD score higher than CISA's?

Both need a service-level login over the network. NVD rates confidentiality and availability impact high (8.8); CISA and B. Braun rate them low with high integrity impact (7.6).

Is CVE-2020-25150 in the CISA KEV catalog?

No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.01529 (75.05th percentile) in the 2026-11-26 dataset.