ISO 14971 ("Medical devices — Application of risk management to medical devices") is the international standard that defines how a medical device manufacturer must identify hazards, estimate and evaluate the resulting risks, control them, and monitor whether those controls actually work, across a device's full life cycle. The current Third Edition, published 9 December 2019, is recognized by FDA as a consensus standard and explicitly excludes two things manufacturers sometimes assume it covers: clinical-procedure decisions and general business risk. It is the foundational safety risk management standard that software (IEC 62304), networked-device (IEC 80001-1) and security-specific (AAMI SW96) standards all build on rather than duplicate.
Key takeaways
- What does each clause produce for the risk management file?
- Two exclusions are stated directly in both ISO's and FDA's own descriptions, and both trip up manufacturers who assume the standard is broader than it is.
- EN ISO 14971:2019 as amended by EN ISO 14971:2019/A11:2021.
- As a connected, parallel assessment that feeds the safety file.
What problem does ISO 14971 solve?
Before ISO 14971, medical device risk management guidance was fragmented across national and sector-specific documents — its own drafting history traces back to EN 1441, a 1997 European risk-analysis standard, and ISO/IEC Guide 51 and Guide 63's general work on safety aspects in standards. ISO 14971 consolidated that into a single, auditable process: a manufacturer runs the same risk analysis → evaluation → control → residual-risk review cycle for every device, regardless of what kind of harm is in scope. Per ISO's own catalogue listing, covered risk categories explicitly include biocompatibility, data and systems security, electricity, moving parts, radiation, and usability — the standard treats a cybersecurity vulnerability that could cause patient harm as just another hazard type feeding the same process, which is why AAMI SW96's security-specific risk management explicitly operates "within the risk management framework defined by ISO 14971" rather than building a separate parallel system.
What does the Third Edition (2019) actually require?
Per the FDA's own recognized-standard record, ISO 14971 requires manufacturers to "identify the hazards associated with medical devices, including in vitro diagnostic (IVD) medical devices, to estimate and evaluate the associated risks, to control these risks, and to monitor the effectiveness of the controls." That requirement runs across 10 main clauses (per the standard's published clause structure): Clause 4 sets general requirements for the risk management system itself; Clause 5 is risk analysis; Clause 6, risk evaluation; Clause 7, risk control; Clause 8, evaluation of overall residual risk (not just risk-by-risk, but the combined picture); Clause 9, a risk management review; and Clause 10, production and post-production activities — the requirement to keep monitoring a device's real-world risk profile after it ships, not just at design time. Three informative annexes (rationale, process overview, fundamental concepts) explain the clauses but impose no separate obligations of their own.
What does each clause produce for the risk management file?
Each main clause leaves a record behind, and the EU MDR asks for the same records in its own words. The middle column below lists the record a reviewer usually looks for; the right column points to the closest item in EU MDR Annex I Section 3, which lists six duties labelled (a) to (f).
| ISO 14971:2019 clause | Typical record in the risk management file | Closest EU MDR Annex I point |
|---|---|---|
| 4 General requirements | Risk management plan with risk acceptability criteria; the file itself | Section 3(a): a risk management plan for each device |
| 5 Risk analysis | Intended use, reasonably foreseeable misuse, hazards, hazardous situations, risk estimates | Section 3(b) and 3(c) |
| 6 Risk evaluation | Each estimated risk compared with the acceptability criteria | Section 3(c) |
| 7 Risk control | Control measures, proof they were implemented and work, residual risk per hazard | Section 3(d) and Section 4 |
| 8 Overall residual risk | One conclusion for the whole device, weighed against its benefits | Section 4 (overall residual risk) and Section 1 |
| 9 Risk management review | Review before commercial release that the plan was carried out | Annex II Section 5 (results go into the technical documentation) |
| 10 Production and post-production | Collection and review of field information, feedback into the file | Section 3(e) and 3(f) |
Clause titles per ISO's catalogue listing; the mapping is SoftWSP's reading, not an official correspondence table.
What does ISO 14971 explicitly not cover?
Two exclusions are stated directly in both ISO's and FDA's own descriptions, and both trip up manufacturers who assume the standard is broader than it is. First, ISO 14971 "does not apply to decisions on the use of a medical device in the context of any particular clinical procedure" — it governs the device's designed risk profile, not a clinician's case-by-case judgment calls. Second, it explicitly excludes business risk management — commercial, reputational or financial risk to the manufacturer is out of scope; this is purely a patient/user/third-party harm framework. FDA's guidance adds a practical caveat on top of the standard's own exclusions: because ISO 14971 is process-oriented rather than prescriptive, a bare declaration of conformity to it is often not sufficient supporting evidence on its own in a premarket submission — reviewers still expect to see the actual risk management file's substance.
What happened to the 2007 edition, and does FDA recognize the current one?
FDA's recognized-consensus-standards database lists ISO 14971 Third Edition 2019-12 (FR Recognition Number 5-125, FR Recognition List Number 053, Date of Entry 12/23/2019) as identically adopted by ANSI/AAMI as ANSI/AAMI/ISO 14971:2019. That recognition arrived as part of a larger batch — more than 100 other consensus standards from ISO, AAMI, ANSI, ASTM International and USP were recognized in the same action, per contemporaneous reporting. FDA set a firm transition deadline for the standard it replaced: declarations of conformity to the superseded ISO 14971:2007 were accepted only "until 25 December 2022. After this transition period, declarations of conformity to [ISO 14971:2007] will not be accepted." Any current FDA submission relying on ISO 14971 conformity needs to cite the 2019 edition.
Which edition counts under the EU MDR?
EN ISO 14971:2019 as amended by EN ISO 14971:2019/A11:2021. Commission Implementing Decision (EU) 2022/757 of 11 May 2022 added that combination to the list of harmonised standards for the MDR. Its recitals explain that CEN and Cenelec adopted the A11 amendment "to adapt them to the requirements of Regulation (EU) 2017/745," and that devices conforming to a harmonised standard whose reference is published in the Official Journal are presumed to meet the requirements it covers (MDR Article 8(1)).
The MDR itself sets the risk management duties the standard supports. Annex I Section 3 requires manufacturers to "establish, implement, document and maintain a risk management system," understood as "a continuous iterative process throughout the entire lifecycle of a device." Annex I Section 4 then fixes the order for choosing controls:
- eliminate or reduce risks as far as possible through safe design and manufacture;
- where appropriate, take adequate protection measures, including alarms, for risks that cannot be eliminated;
- provide information for safety and, where appropriate, training to users.
Residual risk must be judged acceptable both per hazard and overall, and users must be told about it. Annex II Section 5 puts "the solutions adopted and the results of the risk management" into the technical documentation. For software, Annex I 17.2 adds that risk management must include information security. The EU MDR Annex I guide covers the cybersecurity points in detail.
How does cybersecurity fit into an ISO 14971 process?
As a connected, parallel assessment that feeds the safety file. FDA and the EU's Medical Device Coordination Group describe the same interface in different words:
| Question | FDA premarket guidance (February 2026, Section V.A) | MDCG 2019-16 rev. 1 (Section 3.2) |
|---|---|---|
| One process or two? | "Performing security risk management is distinct from performing safety risk management as described in ISO 14971" | Security risks sit inside the risk management scope "to avoid any misunderstanding that a separate process would be needed"; specific methods are used for them |
| What is assessed? | Exploitability, not probability; FDA notes the non-probabilistic approach is not the fundamental approach of ISO 14971 safety risk management | The same elements as safety: analysis, evaluation, control, residual risk and reporting, in a security risk management plan |
| How do the files connect? | Controls for one type of risk must not introduce new risks of the other type; AAMI TIR57 and ANSI/AAMI SW96 describe the interface | Security risks or controls with a safety impact go into the safety risk assessment, and the reverse (Annex IV) |
| What does the safety file list? | A separate security risk assessment accompanies the safety assessment (ISO 13485 subclause 7.1) | Generic security hazards such as denial of service, code execution, memory corruption, information disclosure and privilege gain, not every attack vector |
| What goes to the regulator? | A security risk management plan and report, with traceability between threat model, risk assessment, SBOM and testing | The risk management results in the technical documentation |
Sources: FDA premarket cybersecurity guidance, Section V.A and V.A.2; MDCG 2019-16 rev. 1, Sections 3.2 and 3.3. Both read 2026-11-10.
The FDA premarket cybersecurity guide lists the other documents that sit next to the security risk management report in a submission, and the MDCG 2019-16 guide covers the EU guidance as a whole.
Can a security control create a safety risk?
Yes, which is why the two files must reference each other. MDCG 2019-16 gives the standard example: blanking a screen can be a sound control against disclosure of personal data, but on a device used for interventional procedures or for displaying vital signs, a blank screen is a safety concern, so the control should not be implemented there. FDA makes the reverse point for testing: a functional defect may be left unfixed when its impact is small, but a security anomaly may still need mitigation because of how it could be exploited.
How does it relate to ISO/TR 24971 and AAMI SW96?
ISO/TR 24971 is the practical how-to companion — a Technical Report, not a requirements standard — that FDA's database lists as Second Edition 2020-06, "Medical devices — Guidance on the application of ISO 14971," adopted identically as AAMI/ISO TIR24971:2020. It explains how to actually run the clauses (worked examples, guidance on risk acceptability criteria ISO 14971 itself deliberately leaves undefined) without adding new obligations. AAMI SW96 sits differently: it is a normative security-specific standard, not guidance, and it is explicitly scoped to extend ISO 14971 rather than duplicate it — see our AAMI SW96 guide for its security risk management process, plan and file structure, and our guides to IEC 62304 and IEC 80001-1 for how the software-lifecycle and networked-device standards likewise plug into the same ISO 14971 risk management file rather than running a separate process.
Worked example: how does one firmware CVE travel through both files?
Synthetic example: the device, component and scores are invented to show the hand-offs; no real product or CVE is described.
A manufacturer's networked syringe pump has a communication module that sends status and alarms to a central station. A newly published CVE affects the TLS library in module firmware 4.1, with a CVSS v3.1 base score of 7.5; it is not in CISA's Known Exploited Vulnerabilities catalog.
IdentifyThe SBOM for firmware 4.1 lists the library and version, so the match is found within a day of publication. Assess in the security fileThe threat model already assumes, as FDA recommends, that an adversary controls the hospital network. The TLS endpoint is reachable without authentication, so exploitability is rated high. Before control, the security risk is unacceptable under the plan's criteria. Hand off to the safety fileThe security finding maps to the generic hazard "denial of service". The hazardous situation is that alarm forwarding to the central station stops while the local alarm still sounds. The safety team estimates the harm under its own ISO 14971 severity scale. Choose controls in priority orderFirst a design fix: firmware 4.2 replaces the library (Annex I Section 4, point (a)). Until sites install it, an advisory and updated instructions tell them to restrict the port to the central station's address (point (c)). Check the controlVerification shows the update keeps infusion settings and alarm limits, so the security control adds no new safety risk. Close the loopThe safety file records the new control and re-evaluates overall residual risk (Clauses 7 and 8). The security report links threat model, SBOM entry and test report. Post-production records feed back under Clause 10 and Annex I Section 3(e) and (f).
What are the common pitfalls?
- Copying every attack vector into the safety analysis. MDCG 2019-16 recommends generic security hazards instead, so the safety file stays readable.
- Scoring exploits by probability. FDA expects exploitability-based assessment for security risk, not the probabilistic estimates used for random failures.
- Treating a declaration of conformity as evidence. FDA expects the substance of the risk management file, not only the declaration.
- Citing the wrong edition. FDA stopped accepting ISO 14971:2007 declarations after 25 December 2022, and an EU file should cite EN ISO 14971:2019 with A11:2021.
- Mixing business risk into the safety file. ISO 14971 excludes it, and FDA notes that business and reputational risks fall outside its safety and effectiveness assessment.
- Closing the file at clearance. Clause 10 and MDR Annex I Section 3(e) require field information to keep flowing in, including new vulnerabilities in devices that are aging out of support.
SoftWSP's platform produces one input to this process: it extracts an SBOM from device firmware as CycloneDX 1.6 JSON, matches the components against NVD, CISA KEV, FIRST EPSS and CISA ICS advisories, and drafts a cited report for an engineer to review. It does not judge exploitability, severity or residual risk; those decisions stay in your security and safety files. To check a severity vector before it goes into the security risk assessment, use the free CVSS calculator. See device & firmware security services for how that scope covers security risk inputs alongside a conventional firmware assessment.
This guide is based on FDA's own recognized-consensus-standards database record for ISO 14971:2019, ISO's own catalogue listing for the standard (scope, clause structure, page count), RAPS's contemporaneous reporting on FDA's December 2019 recognition action and the 2007-edition transition deadline, and, read on 2026-11-10, FDA's February 2026 premarket cybersecurity guidance, the EU MDR, Implementing Decision (EU) 2022/757 and MDCG 2019-16 rev. 1. ISO 14971's full normative text is paywalled and was not independently read for this guide; clause-level wording beyond what is quoted here is not drawn from that purchased text.
Put this guide to work
Frequently asked questions
What is ISO 14971?
ISO 14971 is the international standard titled "Medical devices — Application of risk management to medical devices." Per the FDA's own recognized-standard record, it establishes a risk management process for manufacturers to "identify the hazards associated with medical devices, including in vitro diagnostic (IVD) medical devices, to estimate and evaluate the associated risks, to control these risks, and to monitor the effectiveness of the controls." ISO's own catalogue page describes the current (2019) edition as specifying "terminology, principles and a process for risk management of medical devices, including software as a medical device and in vitro diagnostic medical devices." It is the baseline safety risk management standard that nearly every other medical device standard — including software (IEC 62304), networked-device (IEC 80001-1) and security-specific (AAMI SW96) standards — is written to sit inside.
What does the current (2019) edition cover, and what does it explicitly exclude?
ISO 14971:2019 is the Third Edition, published 9 December 2019, superseding the 2007 Second Edition. Per the FDA's recognition record, it "applies across all life-cycle stages of medical devices but excludes clinical decision-making and does not specify acceptable risk levels." ISO's catalogue listing is explicit about two further exclusions: the standard "does not apply to decisions on the use of a medical device in the context of any particular clinical procedure; or business risk management." A quality management system is not required to apply it, though ISO notes risk management may be integrated into one (commonly ISO 13485). Within scope, ISO's listing covers risk categories including biocompatibility, data and systems security, electricity, moving parts, radiation, and usability, across the device's full life from "initial conception" through "ultimate decommissioning and disposal."
What is the ten-clause risk management process ISO 14971 sets out?
Per the standard's own published clause structure (confirmed against its international-standards catalogue listing), ISO 14971:2019 runs 10 main clauses after its scope/references/definitions front matter: Clause 4, general requirements for a risk management system; Clause 5, risk analysis; Clause 6, risk evaluation; Clause 7, risk control; Clause 8, evaluation of overall residual risk; Clause 9, risk management review; and Clause 10, production and post-production activities. Three informative annexes (A: rationale for requirements, B: risk management process overview, C: fundamental risk concepts) support the normative clauses but do not themselves impose requirements.
Does FDA recognize ISO 14971:2019, and what happened to the 2007 edition?
Yes. The FDA's own recognized-consensus-standards database lists ISO 14971 Third Edition 2019-12 under FR Recognition Number 5-125, FR Recognition List Number 053, Date of Entry 12/23/2019, with "Identical Adoption" as ANSI AAMI ISO 14971:2019. FDA recognized it in a batch alongside more than 100 other consensus standards from ISO, AAMI, ANSI, ASTM International and USP. FDA set a transition deadline for the superseded edition: declarations of conformity to ISO 14971:2007 were accepted "until 25 December 2022. After this transition period, declarations of conformity to [ISO 14971:2007] will not be accepted." FDA's own guidance on recognized standards separately cautions that ISO 14971 is process-oriented enough that a bare declaration of conformity is often not sufficient supporting evidence on its own in a submission.
How does ISO 14971 relate to ISO/TR 24971 and AAMI SW96?
ISO/TR 24971 is the companion guidance document — not a requirements standard — that explains how to apply ISO 14971's clauses in practice; FDA's recognized-standards database lists its Second Edition (2020-06), "Medical devices — Guidance on the application of ISO 14971," adopted identically as AAMI/ISO TIR24971:2020. AAMI SW96 (2023) is a separate, security-specific standard that deliberately operates inside ISO 14971's existing risk management framework rather than replacing it — per AAMI's own scope language, SW96 provides requirements for "security risk management for device manufacturers... within the risk management framework defined by ISO 14971." See our [AAMI SW96 guide](/guides/aami-sw96-medical-device-security-risk-management/) for how the two fit together in practice.
Sources
- FDA: Recognized Consensus Standards detail record — ISO 14971 Third Edition 2019-12 (standard identification no. 41349)
- ISO: ISO 14971:2019 — Medical devices — Application of risk management to medical devices (catalogue page)
- RAPS (Regulatory Affairs Professionals Society): "FDA Recognizes New Version of ISO 14971, More Than 100 Other Consensus Standards" (16 January 2020)
- iTeh Standards catalogue: ISO 14971:2019 record (page count, clause structure, annexes)
- FDA: Cybersecurity in Medical Devices — Quality Management System Considerations and Content of Premarket Submissions (guidance, February 2026), Section V.A
- Regulation (EU) 2017/745 on medical devices (EU MDR), Annex I Sections 1–4 and 17.2, Annex II Section 5
- Commission Implementing Decision (EU) 2022/757 of 11 May 2022 — harmonised standards incl. EN ISO 14971:2019/A11:2021 under the MDR
- MDCG 2019-16 rev. 1: Guidance on Cybersecurity for medical devices (July 2020), Sections 3.2 and 3.3
About this guide
SoftWSP Regulation Desk
Covers the Medical & regulated devices cluster: 30 guides.

In this cluster
Questions about this page?