Security Advisory

CVE-2020-6961: GE CARESCAPE Telemetry SSH Key Exposure

Last updated:
  • CVE-2020-6961

CVE-2020-6961 is a credential-storage flaw in GE patient-monitoring servers: ApexPro and CARESCAPE Telemetry Servers, Clinical Information Center and CARESCAPE Central Station keep an SSH private key in configuration files an attacker can reach. NVD and CISA both score it 10.0. GE's remedy is strict isolation of the monitoring networks.

Vulnerability summary

CVSS base score

10

Critical severity

EPSS score (FIRST)

1.6%

Probability of exploitation in the wild in the next 30 days

CISA KEV catalog

No

Not listed

Published

Vulnerability ID: CVE-2020-6961 · Source: NVD

CVSS v3.1 severity rating scale: None 0.0, Low 0.1 to 3.9, Medium 4.0 to 6.9, High 7.0 to 8.9, Critical 9.0 to 10.0
How a CVSS v3.1 base score maps to a severity rating. Source: FIRST CVSS v3.1 specification, section 5.

Affected components

ComponentCPEVersion range
GE ApexPro Telemetry Servercpe:2.3:o:gehealthcare:apexpro_telemetry_server_firmware:*:*:*:*:*:*:*:*4.2 and earlier (CVE record, NVD, CISA ICSMA-20-023-01); GE remedy: MC/IX network isolation per its network configuration guides
GE CARESCAPE Telemetry Servercpe:2.3:o:gehealthcare:carescape_telemetry_server_mp100r_firmware:*:*:*:*:*:*:*:*4.2 and earlier, and 4.3 (CVE record, NVD, CISA)
GE Clinical Information Center (CIC)cpe:2.3:o:gehealthcare:clinical_information_center_mp100d_firmware:5.0:*:*:*:*:*:*:*4.X and 5.X (CVE record, CISA); NVD CPEs name MP100D and MP100R firmware 4.0 and 5.0
GE CARESCAPE Central Station (CSCS)cpe:2.3:o:gehealthcare:carescape_central_station_mai700_firmware:1.0:*:*:*:*:*:*:*1.X (CVE record, CISA); NVD CPEs name MAI700 and MAS700 firmware 1.0

What does this mean for affected devices?

The CVE record from CISA says that in ApexPro Telemetry Server versions 4.2 and prior, CARESCAPE Telemetry Server 4.2 and prior and 4.3, Clinical Information Center (CIC) versions 4.X and 5.X and CARESCAPE Central Station (CSCS) versions 1.X, "a vulnerability exists in the affected products that could allow an attacker to obtain access to the SSH private key in configuration files." NVD's Primary score is 10.0 (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H, CVSS 2.0 7.5) with weakness CWE-522, insufficiently protected credentials. CISA gives the same 10.0 vector in CVSS 3.0 but maps it to CWE-256, plaintext storage of a password; the CVE record carries no CVSS score (NVD, CVE record).

Which GE patient-monitoring products are affected?

CISA's medical advisory ICSMA-20-023-01, released on 2020-01-23, covers six flaws in GE CARESCAPE, ApexPro and Clinical Information Center systems, in the Healthcare and Public Health sector, deployed worldwide. Elad Luz of CyberMDX reported them. Besides the servers above, the advisory lists CARESCAPE Central Station 2.X and the CARESCAPE B450, B650 and B850 bedside monitors for the set. The siblings are an input validation flaw in the web-based configuration utility (CVE-2020-6962), hard-coded SMB credentials (CVE-2020-6963), weak encryption for remote desktop control (CVE-2020-6966), all 10.0 on NVD, an update-package file upload flaw (CVE-2020-6965, NVD 9.9) and unauthenticated keyboard switching (CVE-2020-6964, NVD 8.6). CISA warns that exploitation could cause a loss of monitoring or alarms during active patient monitoring, expose PHI and allow changes to alarm settings on connected monitors.

How do you fix CVE-2020-6961?

The advisory names no fixed software version for this CVE. GE's remedy, as CISA records it, is to confirm that the mission critical (MC) and information exchange (IX) networks are configured and isolated as the Patient Monitoring Network Configuration Guide, the CARESCAPE Network Configuration Guide and the product technical and service manuals require; GE provides these through its customer support portal. On a properly isolated network, GE says, an attacker needs physical access to exploit the flaw. GE also said it was developing software updates with further security enhancements and pointed users to its product security website for notifications.

An older GE Healthcare flaw on this site is the Centricity PACS default passwords issue, and a later infusion-system advisory is BD Alaris CVE-2020-25165. For device makers, the hard-coded credentials fix guide covers per-device keys instead of shared ones, and the FDA postmarket guide covers how such flaws are handled after release.

Is CVE-2020-6961 exploited?

CVE-2020-6961 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.01619, the 76.44th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS). CISA reported no known public exploits specifically targeting these vulnerabilities.

Three signals on every advisory: CVSS says how severe the flaw is, EPSS gives the probability of exploitation in the wild in the next 30 days, and a CISA KEV listing means the flaw has been exploited in the wild
CVSS rates severity, EPSS estimates the likelihood of exploitation, and CISA KEV records exploitation seen in the wild. Source: FIRST EPSS; CISA KEV catalog.
How to fix this class of issueNo dedicated SoftWSP guide covers this weakness class; the fix is usually the vendor patch or firmware update referenced above. Our related guide explains how to plan and verify it: NIST SP 800-40 Rev. 4: enterprise patch management planning, explained.
From advisory to verified fix in four steps: read the advisory, find the affected component in the firmware SBOM, apply the fixed version or vendor mitigation, then re-scan the image to confirm the vulnerable version is gone
Read the advisory, find the component in your firmware SBOM, apply the fix, then re-scan to confirm it.

Evidence

  • NVD (read 2026-11-27): status Modified; published 2020-01-24; [email protected] Primary CVSS 3.1 10.0 AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H, CVSS 2.0 7.5; NVD CWE-522, ICS-CERT CWE-256; CPEs apexpro_telemetry_server_firmware <= 4.2, carescape_telemetry_server_mp100r_firmware <= 4.2 and 4.3, clinical_information_center_mp100d/mp100r_firmware 4.0 and 5.0, carescape_central_station_mai700/mas700_firmware 1.0. — NVD
  • CVE record (CNA icscert, published 2020-01-24): ApexPro Telemetry Server <= 4.2, CARESCAPE Telemetry Server <= 4.2 and 4.3, CIC 4.X and 5.X, CSCS 1.X; SSH private key in configuration files; no CNA CVSS. — CVE.org
  • CISA ICSMA-20-023-01 (CSAF, 2020-01-23): six CVEs 2020-6961..6966; CVE-2020-6961 CWE-256 CVSS 3.0 10.0; products incl. B450/B650/B850 monitors and CSCS 2.X; risk: loss of monitoring/alarms, PHI, alarm-setting changes; GE remedy: confirm MC/IX network isolation per Patient Monitoring Network Configuration Guide and CARESCAPE Network Configuration Guide (via GE support portal); physical access needed on isolated network; GE developing updates; reported by Elad Luz (CyberMDX); no known public exploits. — CISA CSAF
  • NVD Primary scores of siblings (read 2026-11-27): CVE-2020-6962, -6963, -6966 10.0; CVE-2020-6965 9.9; CVE-2020-6964 8.6. — NVD
  • CISA KEV catalog version 2026.11.25, 1,805 entries: CVE-2020-6961 not listed; checked 2026-11-27. — CISA KEV
  • EPSS 0.01619, percentile 0.76441, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories

Sources

  1. NVD
  2. CISA ICSMA-20-023-01: GE CARESCAPE, ApexPro, and Clinical Information Center systems (CSAF)
  3. CISA ICSMA-20-023-01 advisory page
  4. CVE.org record CVE-2020-6961
  5. CISA KEV
  6. FIRST EPSS

Frequently asked questions

Which GE products are affected by CVE-2020-6961?

ApexPro Telemetry Server 4.2 and earlier, CARESCAPE Telemetry Server 4.2 and earlier and 4.3, Clinical Information Center 4.X and 5.X, and CARESCAPE Central Station 1.X, per the CVE record and CISA ICSMA-20-023-01.

Is there a software update for CVE-2020-6961?

CISA's advisory names no fixed version. GE's remedy is to isolate and configure the MC and IX networks as its network configuration guides require; GE said it was developing software updates with further security enhancements.

Why is CVE-2020-6961 scored 10.0?

NVD and CISA both use AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H: network attack vector, low complexity, no privileges or user interaction, changed scope and high impact on confidentiality, integrity and availability. The CVE record itself carries no CVSS score.

Is CVE-2020-6961 in the CISA KEV catalog?

No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.01619, the 76.44th percentile, in the 2026-11-26 dataset.