CVE-2020-25165 is an authentication weakness between the BD Alaris PC Unit Model 8015 infusion system and BD Alaris Systems Manager. An attacker on the network can tamper with session data and cut the PC Unit's wireless link, forcing manual operation. NVD scores it 7.5, CISA 6.5. Systems Manager 12.0.1 and later releases address it.
Vulnerability summary
CVSS base score
7.5
High severity
EPSS score (FIRST)
1.7%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2020-25165 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| BD Alaris PC Unit Model 8015 | cpe:2.3:o:bd:alaris_8015_pcu_firmware:*:*:*:*:*:*:*:* | 9.33.1 and earlier (CISA ICSMA-20-317-01, NVD); BD stated a later PC Unit software version would address it |
| BD Alaris Systems Manager | cpe:2.3:a:bd:alaris_systems_manager:*:*:*:*:*:*:*:* | 4.33 and earlier (CISA, NVD); versions 12.0.1, 12.0.2, 12.1.0 and 12.1.2 address it |
What does this mean for affected devices?
The CVE record from CISA says that BD Alaris PC Unit Model 8015, versions 9.33.1 and earlier, and BD Alaris Systems Manager, versions 4.33 and earlier, "are vulnerable to a network session authentication vulnerability" in the authentication process between them, and that an attacker "could perform a denial-of-service attack on the BD Alaris PC Unit by modifying the configuration headers of data in transit", leading to "a drop in the wireless capability of the BD Alaris PC Unit, resulting in manual operation of the PC Unit." NVD's Primary score is 7.5 (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H, CVSS 2.0 5.0). CISA scores it 6.5 (CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:L) in its advisory, and the CVE record carries no CVSS score. The weakness is CWE-287, improper authentication, as mapped by CISA (NVD, CVE record).
Which BD Alaris products are affected?
CISA's medical advisory ICSMA-20-317-01, released on 2020-11-12, lists the BD Alaris PC Unit Model 8015 at 9.33.1 and earlier and the BD Alaris Systems Manager at 4.33 and earlier, in the Healthcare and Public Health sector, deployed worldwide. Medigate discovered the flaw and reported it to BD. To exploit it, CISA says, an attacker would need access to the network of the devices, redirect the PC Unit's authentication requests with custom code and complete a handshake from the extracted information. The PC Unit "will continue to function as programmed", but network services stop: pre-populating infusion parameters through EMR interoperability and wireless updates of the Alaris System Guardrails (DERS) are unavailable.
How do you fix CVE-2020-25165?
BD's remedy, as recorded by CISA, has two parts. On the server side, BD Alaris Systems Manager versions 12.0.1, 12.0.2, 12.1.0 and 12.1.2 address the flaw, and many installations were already updated during BD's normal server upgrades. On the device side, BD said a later version of the PC Unit software would address it. BD's compensating controls:
- enable the firewall on the Systems Manager server image with rules restricting ports and services, inbound and outbound;
- filter traffic between the wireless segment and the server segment with access control lists limited to authorised device MAC addresses;
- treat Systems Manager as critical infrastructure: isolate it behind firewalls, patch it and run malware protection;
- disable unnecessary accounts, protocols and services.
A later Alaris flaw, CVE-2023-30560, lets someone with physical access change the PC Unit configuration without authentication. Another infusion-pump component with a CISA advisory is the B. Braun Battery Pack SP with Wi-Fi (CVE-2023-0888). For manufacturers, the authentication bypass fix guide covers mutual authentication between devices and servers, and the FDA postmarket guide covers how such flaws are handled after release.
Is CVE-2020-25165 exploited?
CVE-2020-25165 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.01715, the 77.85th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS). CISA reported no known public exploits specifically targeting this vulnerability.


Evidence
- NVD (read 2026-11-27): status Modified; published 2020-11-13; [email protected] Primary CVSS 3.1 7.5 AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H, CVSS 2.0 5.0; CWE-287 from ICS-CERT (Secondary); CPE bd alaris_8015_pcu_firmware up to and including 9.33.1, alaris_systems_manager up to and including 4.33. — NVD
- CISA ICSMA-20-317-01 (2020-11-12): PC Unit 8015 <= 9.33.1, Systems Manager <= 4.33; CWE-287 CVSS 3.0 6.5 AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:L; Systems Manager 12.0.1, 12.0.2, 12.1.0, 12.1.2 address it; many installations already updated; upcoming PC Unit software version; reported by Medigate; PC Unit continues to function, EMR interoperability and wireless Guardrails updates unavailable; no known public exploits. — CISA CSAF
- CISA advisory page (read 2026-11-27): BD mitigations: Systems Manager server firewall with port/service rules, ACL filtering between wireless and server segments limited to authorised MAC addresses, isolate/patch/malware-protect Systems Manager, disable unnecessary accounts, protocols and services. — CISA
- CISA KEV catalog version 2026.11.25, 1,805 entries: CVE-2020-25165 not listed; checked 2026-11-27. — CISA KEV
- EPSS 0.01715, percentile 0.77849, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which BD Alaris versions are affected by CVE-2020-25165?
BD Alaris PC Unit Model 8015 versions 9.33.1 and earlier and BD Alaris Systems Manager versions 4.33 and earlier, per CISA ICSMA-20-317-01 and NVD.
Does CVE-2020-25165 stop infusions?
No. CISA says the PC Unit continues to function as programmed; the attack drops its wireless link, so EMR interoperability and wireless Guardrails (DERS) updates are unavailable and the unit runs in manual operation.
How is CVE-2020-25165 fixed?
BD Alaris Systems Manager 12.0.1, 12.0.2, 12.1.0 and 12.1.2 address it, and BD stated a later PC Unit software version would too. Meanwhile BD recommends server firewalling, MAC-based ACLs between the wireless and server segments, and disabling unneeded services.
Is CVE-2020-25165 in the CISA KEV catalog?
No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.01715, the 77.85th percentile, in the 2026-11-26 dataset.