**CVE-2012-6693** is a default-credential flaw in GE Healthcare's Centricity PACS 4.0 Server, the archive that stores and serves a hospital's medical images: two built-in accounts, nasro (read-only) and nasrw (read/write), ship with known passwords. NVD scores it 10.0 under CVSS v2, and CISA's medical advisory ICSMA-18-037-02 lists it at CVSS v3 9.8.
Vulnerability summary
CVSS base score
10
Critical severity
EPSS score (FIRST)
1.7%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2012-6693 · Source: NVD
Affected components
| Component | CPE | Version range |
|---|---|---|
| GE Healthcare Centricity PACS Server | cpe:2.3:a:gehealthcare:centricity_pacs_server:4.0:*:*:*:*:*:*:* | 4.0 (NVD CPE); CISA ICSMA-18-037-02 lists Centricity PACS Server as known affected; GE product updates replacing the default credentials are available on request |
What does this mean for affected devices?
NVD's description reads: "GE Healthcare Centricity PACS 4.0 Server has a default password of (1) nasro for the nasro (ReadOnly) user and (2) nasrw for the nasrw (Read/Write) user, which has unspecified impact and attack vectors" (NVD). NVD published the record on 2015-08-04 and scored it under CVSS v2 only, at 10.0 with vector AV:N/AC:L/Au:N/C:C/I:C/A:C: network access, no authentication, and complete impact on confidentiality, integrity and availability. NVD records CWE-255 (credentials management) and a single CPE, Centricity PACS Server 4.0. Its references include GE's installation and service manuals for Centricity PACS 4.x and a 2015 Forbes report on hard-coded passwords in medical devices.
A PACS (picture archiving and communication system) server is the central archive that imaging modalities send studies to and diagnostic workstations read from. An account that can read the archive exposes patient images and demographics; a read/write account can also change or delete them. Because the passwords are published in the CVE record itself, anyone who can reach the server's login services can try them.
CISA covered this CVE in its ICS medical advisory ICSMA-18-037-02, "GE Medical Devices Vulnerability", first posted on 2018-02-06 and updated (Update A) on 2018-03-13 (CISA ICSMA-18-037-02). CISA's machine-readable copy (CSAF) lists 23 CVEs across GE imaging products, all under CWE-287 (improper authentication) at CVSS v3 9.8 (CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H), and names Centricity PACS Server as the known-affected product for CVE-2012-6693. The same advisory lists further Centricity PACS credentials: CVE-2012-6694 (a geservice account on Workstation 4.0 and 4.0.1 and Server 4.0), CVE-2012-6695 and CVE-2013-7442 (Workstation accounts), CVE-2017-14008 (Centricity PACS RA1000) and CVE-2011-5322 (listed under Centricity PACS-IW). CISA gives the sector as Healthcare and Public Health, deployed worldwide. Its risk evaluation says exploitation "may allow a remote attacker to bypass authentication and gain access to the affected devices". The CSAF copy notes it was extracted from unstructured data, so the per-CVE wording on this page follows NVD.
The fix comes from GE on request. The advisory states that "GE has produced product updates that are available upon request, which replace default or hard-coded credentials with custom credentials for all but three of the affected products"; the three without updates are the Optima 680, Revolution XQ/i and THUNIS-800+, not Centricity PACS. CISA also recommends keeping such systems off the internet, behind firewalls and reachable remotely only through a maintained VPN. At the time of the advisory, CISA stated that no known public exploitation specifically targeting these vulnerabilities had been reported to it.
CVE-2012-6693 is not in the CISA KEV catalog as of 2026-11-16 (catalog version 2026.11.13). FIRST EPSS gives it 0.0168, percentile 77.3, in the 2026-11-16 dataset (CISA KEV, FIRST EPSS).
Sites still running Centricity PACS 4.x should request GE's credential update, change the nasro and nasrw passwords, and limit which hosts can reach the server. Our DICOM and PACS cybersecurity guide places this advisory among the imaging findings CISA has published since, and removing hard-coded credentials covers how device makers retire factory accounts. For imaging systems kept in service past vendor support, see IMDRF guidance on legacy medical devices. A newer PACS credential finding is CVE-2025-54156, where Sante PACS Server's web portal sent passwords without encryption.
Evidence
- Centricity PACS 4.0 Server default passwords for the nasro (ReadOnly) and nasrw (Read/Write) users; NVD Primary CVSS v2 10.0 (AV:N/AC:L/Au:N/C:C/I:C/A:C); CWE-255; CPE centricity_pacs_server 4.0; published 2015-08-04. — NVD
- ICSMA-18-037-02 initial 2018-02-06, Update A 2018-03-13; 23 CVEs, CWE-287, CVSS v3 9.8; Centricity PACS Server known affected; GE updates on request replace default credentials for all but Optima 680, Revolution XQ/i and THUNIS-800+; sector Healthcare and Public Health; no known public exploitation reported to CISA at the time. — CISA ICSMA-18-037-02: GE Medical Devices Vulnerability (Update A, CSAF)
- Not in the CISA KEV catalog as of 2026-11-16 (catalog version 2026.11.13). — CISA KEV
- EPSS 0.01679 (percentile 77.3), dataset 2026-11-16, read 2026-11-16. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which GE product does CVE-2012-6693 affect?
GE Healthcare Centricity PACS 4.0 Server, the image archive in GE's Centricity PACS line. NVD's CPE lists version 4.0, and CISA ICSMA-18-037-02 names Centricity PACS Server as the known-affected product (per NVD and CISA ICSMA-18-037-02).
Why does NVD score CVE-2012-6693 at 10.0 and CISA at 9.8?
NVD scored it in 2015 under CVSS v2 only (AV:N/AC:L/Au:N/C:C/I:C/A:C, 10.0). CISA's 2018 advisory used CVSS v3 (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, 9.8). Both describe a network attack with no authentication and full impact (per NVD and CISA ICSMA-18-037-02).
Is there a fix for CVE-2012-6693?
Yes. GE offers product updates on request that replace default or hard-coded credentials with custom credentials; the three products without updates in the advisory are the Optima 680, Revolution XQ/i and THUNIS-800+, not Centricity PACS (per CISA ICSMA-18-037-02).
Are there other Centricity PACS credential CVEs?
Yes. The same advisory lists CVE-2012-6694, CVE-2012-6695 and CVE-2013-7442 for Centricity PACS server and workstation accounts, CVE-2017-14008 for Centricity PACS RA1000 and CVE-2011-5322 under Centricity PACS-IW (per CISA ICSMA-18-037-02).
Is CVE-2012-6693 in the CISA KEV catalog?
No, as of 2026-11-16 (catalog version 2026.11.13). FIRST EPSS gives it 0.0168, percentile 77.3, in the 2026-11-16 dataset (per CISA KEV and FIRST EPSS).