CVE-2019-10959 is a firmware update flaw in the BD Alaris Gateway Workstation: the update function does not restrict the upload of malicious files, so an attacker on the network could install unauthorised firmware affecting the docked Alaris GS, GH, CC and TIVA syringe pumps. NVD and CISA both score it 10.0. BD released a fix in October 2019.
Vulnerability summary
CVSS base score
10
Critical severity
EPSS score (FIRST)
2.5%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2019-10959 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| BD Alaris Gateway Workstation | cpe:2.3:o:bd:alaris_gateway_workstation_firmware:1.3.1:13:*:*:*:*:*:* | Firmware 1.1.3 Build 10, 1.1.3 MR Build 11, 1.2 Build 15, 1.3.0 Build 14 and 1.3.1 Build 13 (NVD, CVE record); 1.3.2 and 1.6.1 not impacted (CVE record); BD fix released October 2019 (MHRA MDA/2019/040) |
| BD Alaris GS syringe pump | cpe:2.3:o:bd:alaris_gs_syringe_pump_firmware:*:*:*:*:*:*:*:* | Software 2.3.6 and below (NVD, CISA ICSMA-19-164-01) |
| BD Alaris GH syringe pump | cpe:2.3:o:bd:alaris_gh_syringe_pump_firmware:*:*:*:*:*:*:*:* | Software 2.3.6 and below (NVD, CISA ICSMA-19-164-01) |
| BD Alaris CC syringe pump | cpe:2.3:o:bd:alaris_cc_syringe_pump_firmware:*:*:*:*:*:*:*:* | Software 2.3.6 and below (NVD, CISA ICSMA-19-164-01) |
| BD Alaris TIVA syringe pump | cpe:2.3:o:bd:alaris_tiva_syringe_pump_firmware:*:*:*:*:*:*:*:* | Software 2.3.6 and below (NVD, CISA ICSMA-19-164-01) |
What does this mean for affected devices?
CVE-2019-10959 is an unrestricted file upload weakness (CWE-434) in the firmware update function of the BD Alaris Gateway Workstation, the docking and communications unit for BD's Alaris syringe pumps. The CVE record from ICS-CERT says simply that "the application does not restrict the upload of malicious files during a firmware update." NVD's Primary score is 10.0, CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H, with a CVSS 2.0 score of 7.5, and NVD maps the weakness to CWE-434, the same CWE as ICS-CERT (NVD, CVE record).
Which BD Alaris products and versions are affected?
The CVE record and NVD's CPE configuration list Alaris Gateway Workstation firmware 1.1.3 Build 10, 1.1.3 MR Build 11, 1.2 Build 15, 1.3.0 Build 14 and 1.3.1 Build 13, plus the Alaris GS, Alaris GH, Alaris CC and Alaris TIVA syringe pumps running software 2.3.6 and below. The record states that the firmware versions 1.3.2 and 1.6.1 are not impacted. CISA published the medical advisory ICSMA-19-164-01 on 2019-06-13 (the CSAF copy has a single revision with that date). It places the products in the Healthcare and Public Health sector, deployed in Europe and Asia, and records the vendor's statement that the affected products are not sold in the United States. Elad Luz of CyberMDX reported the flaws.
What can an attacker do with CVE-2019-10959?
CISA's risk evaluation says exploitation of the advisory's vulnerabilities "could allow unauthorized arbitrary code execution", letting an attacker view and edit device status and configuration details and make devices unavailable. The UK regulator MHRA, in Medical Device Alert MDA/2019/040, summarises it as a flaw that may let an attacker "remotely install unauthorized firmware" on the workstation. Canada's Cyber Centre (AV19-116) adds that by installing modified firmware an actor could adjust specific pump commands, including the infusion rate, on certain versions, or disable the pump completely.
How do you fix CVE-2019-10959?
CISA's advisory records BD's mitigations and compensating controls and notes that BD was assessing additional remediation, including an adjustment to restrict the SMB protocol. MHRA's alert, issued 2019-11-27, says BD developed a fix for the vulnerability, released in October 2019 and available from BD's password-protected support site, and asks hospitals to identify affected devices and follow BD's customer letter. Until the fix is applied, keep Gateway Workstations on a segregated clinical network that only authorised systems can reach. For device makers, the firmware OTA update security guide covers the signed-update checks that stop an unauthorised image from being accepted, and the FDA postmarket guide covers handling such flaws after release.
What else is in ICSMA-19-164-01?
The same advisory covers CVE-2019-10962, improper access control (CWE-284) in the workstation's web browser interface: anyone who knows the terminal's IP address can read its status and configuration information. CISA scores it 7.3, and its CVE record lists Gateway versions 1.0.13, 1.1.3 Build 10, 1.1.3 MR Build 11, 1.1.5 and 1.1.6. Other BD Alaris issues on this site are the PC Unit session authentication flaw CVE-2020-25165 and the PC Unit configuration change without authentication CVE-2023-30560; a comparable medical-device file upload flaw is CVE-2021-27489 in the ZOLL Defibrillator Dashboard.
Do NVD and CISA score CVE-2019-10959 the same?
Both give 10.0, but the vectors differ. CISA's advisory (CVSS 3.0) uses AV:N/AC:L/PR:N/UI:N/S:C/C:L/I:H/A:H, a low confidentiality impact, while NVD rates confidentiality high. The CVE record itself carries no CVSS score.
Is CVE-2019-10959 exploited?
It is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25). FIRST EPSS gives it 0.02528, the 85.62nd percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS). At publication CISA reported no known public exploits specifically targeting these vulnerabilities. For how to weigh those two signals, see the KEV and EPSS prioritization guide.


Evidence
- NVD (read 2026-11-27): status Modified; published 2019-06-13; [email protected] Primary CVSS 3.0 10.0 AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H, CVSS 2.0 7.5; CWE-434 (NVD Primary and ICS-CERT); CPEs Gateway Workstation firmware 1.1.3 build 10/11, 1.2 build 15, 1.3.0 build 14, 1.3.1 build 13; GS/GH/CC/TIVA syringe pump firmware up to and including 2.3.6. — NVD
- CVE record (CNA ICS-CERT, published 2019-06-13): application does not restrict the upload of malicious files during a firmware update; affected Gateway versions listed; does not impact latest firmware 1.3.2 and 1.6.1; Alaris GS, GH, CC, TIVA on software 2.3.6 and below; CWE-434; no CVSS in the record. — CVE.org
- CISA ICSMA-19-164-01 (CSAF, initial release 2019-06-13, one revision): CVE-2019-10959 CWE-434 CVSS 3.0 10.0 AV:N/AC:L/PR:N/UI:N/S:C/C:L/I:H/A:H; CVE-2019-10962 CWE-284 7.3; risk: arbitrary code execution, view/edit status and configuration, devices unavailable; not sold in the US; deployed Europe, Asia; Healthcare and Public Health; reported by Elad Luz (CyberMDX); BD assessing additional remediation incl. restricting SMB; no known public exploits. — CISA CSAF
- MHRA Medical Device Alert MDA/2019/040 (issued 2019-11-27): vulnerability may allow an attacker to remotely install unauthorized firmware; BD fix released October 2019, available from BD's password-protected website; identify affected devices and follow BD's customer letter. — MHRA
- Canadian Centre for Cyber Security AV19-116 (2019-06-14): actor could adjust specific pump commands including infusion rate by installing modified firmware; pump could be disabled completely. — Canadian Centre for Cyber Security
- CISA KEV catalog version 2026.11.25, 1,805 entries: CVE-2019-10959 not listed; checked 2026-11-27. — CISA KEV
- EPSS 0.02528, percentile 0.85617, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
- NVD
- CISA ICSMA-19-164-01: BD Alaris Gateway Workstation (CSAF)
- CISA ICSMA-19-164-01 advisory page
- CVE.org record CVE-2019-10959 (CNA: ICS-CERT)
- MHRA Medical Device Alert MDA/2019/040: Alaris Gateway Workstation
- Canadian Centre for Cyber Security AV19-116: Becton Dickinson (BD) Security Advisories
- CISA KEV
- FIRST EPSS
Frequently asked questions
Which BD Alaris versions are affected by CVE-2019-10959?
Alaris Gateway Workstation firmware 1.1.3 Build 10, 1.1.3 MR Build 11, 1.2 Build 15, 1.3.0 Build 14 and 1.3.1 Build 13, and Alaris GS, GH, CC and TIVA syringe pumps on software 2.3.6 and below. Firmware 1.3.2 and 1.6.1 are not impacted (per the CVE record and NVD).
Could CVE-2019-10959 change how an infusion pump behaves?
Canada's Cyber Centre warned that, by installing modified firmware, an actor could adjust specific pump commands including the infusion rate on certain versions, or disable the pump completely; CISA describes possible arbitrary code execution and device unavailability (per Cyber Centre AV19-116 and CISA ICSMA-19-164-01).
Is there a fix for CVE-2019-10959?
Yes. BD developed a fix released in October 2019 and distributed it through its protected support site; hospitals were asked to identify affected devices and follow BD's customer letter. Firmware 1.3.2 and 1.6.1 are not impacted (per MHRA MDA/2019/040 and the CVE record).
Is CVE-2019-10959 in the CISA KEV catalog?
No, not as of 2026-11-27 (catalog version 2026.11.25). FIRST EPSS rates it 0.02528, the 85.62nd percentile, in the 2026-11-26 dataset (per CISA KEV and FIRST EPSS).