CVE-2021-27489 is an unrestricted file upload flaw (CWE-434) in ZOLL Defibrillator Dashboard before version 2.2, a web application that shows data from ZOLL defibrillators. A logged-in user without administrator rights can upload a malicious file and run commands on the server. NVD scores it 8.8 and CISA 9.9. ZOLL fixed it in version 2.2.
Vulnerability summary
CVSS base score
8.8
High severity
EPSS score (FIRST)
1.3%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2021-27489 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| ZOLL Defibrillator Dashboard | cpe:2.3:a:zoll:defibrillator_dashboard:*:*:*:*:*:*:*:* | before 2.2 (NVD versionEndExcluding 2.2, CISA ICSMA-21-161-01); fix: Defibrillator Dashboard 2.2 or later |
What does this mean for affected devices?
The CVE record from CISA says: "ZOLL Defibrillator Dashboard, v prior to 2.2, The web application allows a non-administrative user to upload a malicious file. This file could allow an attacker to remotely execute arbitrary commands." NVD's Primary score is CVSS v3.1 8.8 (CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H). CISA's advisory scores it 9.9 (CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H); the only difference is that CISA treats the scope as changed. Both classify it as CWE-434, unrestricted upload of file with dangerous type. NVD published the record on 2021-06-16 and covers versions before 2.2 (NVD, CVE record).
What else is in ZOLL's advisory?
CISA's medical advisory ICSMA-21-161-01, released on 2021-06-10, covers six flaws in Defibrillator Dashboard before 2.2, used in the Healthcare and Public Health sector worldwide. Besides this upload flaw, they are a hard-coded encryption key (CVE-2021-27481), credentials stored in cleartext (CVE-2021-27487), passwords stored in a recoverable format (CVE-2021-27485), incorrect privilege assignment (CVE-2021-27483) and cross-site scripting (CVE-2021-27479). CISA says exploitation could allow remote code execution, credential theft, or impact on the confidentiality, integrity and availability of the application. An anonymous party reported the flaws to CISA.
How do you fix CVE-2021-27489?
ZOLL recommends upgrading to Defibrillator Dashboard 2.2 or later; ZOLL Data Products Technical Support helps with the upgrade (contact details are in the CISA advisory). ZOLL also recommends:
- treating the data on the defibrillator itself as the accurate source if it disagrees with the Dashboard, and performing frequent local readiness checks of the devices as the manuals describe;
- disabling password autocomplete in browsers used to reach the Dashboard.
Because the upload needs only a low-privilege login, also review who holds Dashboard accounts.
Other hospital software flaws on this site are CVE-2023-40150 in MedDream PACS and CVE-2023-30560 in BD Alaris. The FDA postmarket cybersecurity guide explains how manufacturers handle fixes like this one, and the patch management guide covers rolling out vendor updates.
Is CVE-2021-27489 exploited?
CVE-2021-27489 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.01291, the 70.31th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).


Evidence
- NVD (read 2026-11-27): status Modified; published 2021-06-16; NVD Primary CVSS v3.1 8.8 CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H; NVD Primary CVSS v2 6.5; CWE-434 from ICS-CERT; CPE zoll defibrillator_dashboard versionEndExcluding 2.2. — NVD
- CISA ICSMA-21-161-01 CSAF (released 2021-06-10, read 2026-11-27): ZOLL Defibrillator Dashboard < 2.2; CVE-2021-27489 CWE-434 CVSS v3.0 9.9 AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H; siblings CVE-2021-27481 (CWE-321), 27487 (CWE-312), 27485 (CWE-257), 27483 (CWE-269), 27479 (CWE-79); ZOLL recommends upgrading to 2.2 or later, contact ZOLL Data Products Technical Support; device data is the accurate source, frequent local readiness checks; disable password autocomplete; reported by an anonymous party; Healthcare and Public Health; worldwide; HQ United States. — CISA ICSMA-21-161-01
- CVE record (checked 2026-11-27): state PUBLISHED, CNA ICS-CERT, published 2021-06-16; the record carries no CNA CVSS metric; affected product ZOLL Defibrillator Dashboard; reference is the CISA advisory. — CVE.org
- Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
- EPSS 0.01291, percentile 0.70312, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which version fixes CVE-2021-27489?
ZOLL Defibrillator Dashboard 2.2 or later. Versions before 2.2 are affected, per NVD and CISA ICSMA-21-161-01.
Does CVE-2021-27489 affect the defibrillators themselves?
The flaw is in the Dashboard web application. ZOLL advises treating the data on the defibrillator as the accurate source if the Dashboard disagrees, and doing frequent local readiness checks per the manuals.
Why does CISA score it 9.9 while NVD gives 8.8?
The vectors match except for scope: CISA rates the scope as changed, meaning a compromise can reach beyond the Dashboard application; NVD rates it unchanged.
Is CVE-2021-27489 in the CISA KEV catalog?
No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.01291 (70.31th percentile) in the 2026-11-26 dataset.
Questions about this page?