Security Advisory

CVE-2017-14018: Ethicon Generator Gen11 Auth Bypass

Last updated:
  • CVE-2017-14018

CVE-2017-14018 is an improper authentication flaw (CWE-287) in the Johnson & Johnson Ethicon Endo-Surgery Generator Gen11: the check between the generator and its single-patient use products can be bypassed, so unauthorized devices can be connected. NVD scores it 4.8 (physical access, high complexity). CISA's ICSMA-17-332-01 covers versions released before November 29, 2017, when Ethicon made a field update available.

Vulnerability summary

CVSS base score

4.8

Medium severity

EPSS score (FIRST)

0.4%

Probability of exploitation in the wild in the next 30 days

CISA KEV catalog

No

Not listed

Published

Vulnerability ID: CVE-2017-14018 · Source: NVD

CVSS v3.1 severity rating scale: None 0.0, Low 0.1 to 3.9, Medium 4.0 to 6.9, High 7.0 to 8.9, Critical 9.0 to 10.0
How a CVSS v3.1 base score maps to a severity rating. Source: FIRST CVSS v3.1 specification, section 5.

Affected components

ComponentCPEVersion range
Johnson & Johnson Ethicon Endo-Surgery Generator Gen11cpe:2.3:o:ethicon:endo-surgery_generator_gen11_firmware:-:*:*:*:*:*:*:*all versions released before November 29, 2017 (CISA ICSMA-17-332-01); fixed by the Ethicon field cybersecurity update made available on November 29, 2017

What does this mean for affected devices?

NVD describes CVE-2017-14018 as "an improper authentication issue" in "Johnson & Johnson Ethicon Endo-Surgery Generator Gen11, all versions released before November 29, 2017." In its words, "the security authentication mechanism used between the Ethicon Endo-Surgery Generator Gen11 and single-patient use products can be bypassed, allowing for unauthorized devices to be connected to the generator, which could result in a loss of integrity or availability." NVD's Primary CVSS v3.0 score is 4.8 (medium), vector CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L, with a legacy CVSS v2 score of 3.3 (AV:L/AC:M/Au:N/C:N/I:P/A:P). Both NVD and CISA map the weakness to CWE-287, improper authentication (NVD, CVE record).

What does the Gen11 flaw let an attacker do?

The authentication at issue is the check the Gen11 generator runs on the single-patient use products connected to it. A bypass means the generator can accept a device it should reject. The vector explains the medium score: the attack is physical (AV:P) with high complexity (AC:H), so it needs hands on the generator rather than network access, and the impact is high integrity and low availability with no confidentiality loss. CISA's 4.8 score in the advisory is the same as NVD's, with the same vector. CISA also notes that no known public exploits specifically target this vulnerability.

Which products are affected?

CISA's medical advisory ICSMA-17-332-01, released on 2017-11-28, covers one product: the Ethicon Endo-Surgery Generator Gen11, all versions released before November 29, 2017. It is the only CVE in that advisory, so there are no sibling CVEs to track alongside it. NVD's configuration lists the firmware CPE cpe:2.3:o:ethicon:endo-surgery_generator_gen11_firmware:-:*:*:*:*:*:*:* running on the Gen11 hardware, without a version range. The disclosure came from the manufacturer itself: CISA's acknowledgments credit Johnson & Johnson, the parent company, and Ethicon Endo-Surgery, LLC with reporting the vulnerability (CISA CSAF advisory).

How do you fix CVE-2017-14018?

The fix is a vendor-run field update rather than a download. According to CISA's remediation text, Ethicon Endo-Surgery "has contacted users and initiated a field cybersecurity update" for the Gen11, made available on November 29, 2017. Users with questions about the vulnerability or the update are directed to their Ethicon Endo-Surgery sales representative or the Ethicon Customer Support Center at 1-877-ETHICON. For a hospital, that means confirming with Ethicon that every Gen11 in the fleet received the November 2017 update, and recording it in the medical device inventory.

Two of our guides cover the wider picture. The FDA postmarket cybersecurity guide explains how FDA frames manufacturer fixes for marketed devices, and the IMDRF legacy medical device guide covers devices that stay in service long after their software is fixed. For engineers, the bug class is the one our authentication bypass fix guide covers. Related device advisories with a CWE-287 or missing-authentication angle are CVE-2020-25165 in the BD Alaris PC Unit and CVE-2023-30560 in the BD Alaris PCU.

Is CVE-2017-14018 exploited or likely to be?

CVE-2017-14018 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25). FIRST EPSS gives it 0.00364, the 22nd percentile, in the 2026-11-26 dataset, as of 2026-11-27 (FIRST EPSS). Combined with the physical attack vector, that places it low on a patching queue, though for a device used in surgery the high integrity impact still makes the 2017 update worth confirming.

Three signals on every advisory: CVSS says how severe the flaw is, EPSS gives the probability of exploitation in the wild in the next 30 days, and a CISA KEV listing means the flaw has been exploited in the wild
CVSS rates severity, EPSS estimates the likelihood of exploitation, and CISA KEV records exploitation seen in the wild. Source: FIRST EPSS; CISA KEV catalog.
How to fix this class of issueThis advisory's weakness class is Authentication bypass. Our practical guide covers how to detect, fix, verify and disclose it: How to fix authentication bypass in embedded firmware.
From advisory to verified fix in four steps: read the advisory, find the affected component in the firmware SBOM, apply the fixed version or vendor mitigation, then re-scan the image to confirm the vulnerable version is gone
Read the advisory, find the component in your firmware SBOM, apply the fix, then re-scan to confirm it.

Evidence

  • Improper authentication in Ethicon Endo-Surgery Generator Gen11, all versions released before November 29, 2017; authentication with single-patient use products can be bypassed; Primary CVSS v3.0 4.8 CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L; CVSS v2 3.3; CWE-287; firmware CPE without version. — NVD
  • ICSMA-17-332-01 released 2017-11-28; single CVE; CVSS v3 4.8 same vector; CWE-287; reported by Johnson & Johnson and Ethicon Endo-Surgery, LLC; field cybersecurity update available November 29, 2017; contact sales representative or 1-877-ETHICON; no known public exploits. — CISA ICSMA-17-332-01 (CSAF)
  • CNA ICS-CERT record with CWE-287 and the same description. — CVE.org
  • Not in the KEV catalog, version 2026.11.25, checked 2026-11-27. — CISA KEV
  • EPSS 0.00364, 22nd percentile, 2026-11-26 dataset. — FIRST EPSS

Related advisories

Sources

  1. NVD
  2. CISA ICSMA-17-332-01: Ethicon Endo-Surgery Generator Gen11 (CSAF)
  3. CISA ICSMA-17-332-01 advisory page
  4. CVE.org record CVE-2017-14018
  5. CISA KEV
  6. FIRST EPSS

Frequently asked questions

Is there a fix for CVE-2017-14018?

Yes. Ethicon Endo-Surgery contacted users and started a field cybersecurity update for the Gen11, made available on November 29, 2017. Questions go to the Ethicon sales representative or the Ethicon Customer Support Center at 1-877-ETHICON (per CISA ICSMA-17-332-01).

Why is the score only 4.8 when integrity impact is high?

The vector requires physical access with high attack complexity and has no confidentiality impact, which keeps the base score at 4.8 even though integrity impact is high. NVD and CISA give the same score and vector (per NVD and CISA ICSMA-17-332-01).

Who reported CVE-2017-14018?

The manufacturer did. CISA credits Johnson & Johnson, the parent company, and Ethicon Endo-Surgery, LLC with reporting the improper authentication vulnerability, and it is the only CVE in ICSMA-17-332-01 (per CISA ICSMA-17-332-01).

Is CVE-2017-14018 in the CISA KEV catalog?

No. It is not listed in catalog version 2026.11.25, checked 2026-11-27, and FIRST EPSS gives it 0.00364, the 22nd percentile, in the 2026-11-26 dataset (per CISA KEV and FIRST EPSS).