CVE-2019-13539 is weak password hashing in Medtronic Valleylab FT10 and FX8 energy platforms and the Valleylab Exchange Client: operating system passwords are hashed with the old descrypt algorithm. An attacker who first gets a local shell through sibling flaws can read and crack them. NVD scores it 7.8, CISA 7.0. Apply Medtronic's patches and keep the devices off IP networks.
Vulnerability summary
CVSS base score
7.8
High severity
EPSS score (FIRST)
0.3%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2019-13539 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) | cpe:2.3:o:medtronic:valleylab_ft10_energy_platform_firmware:*:*:*:*:*:*:*:* | Software 4.0.0 and earlier (NVD, CVE record, CISA ICSMA-19-311-02); Medtronic software patch available (CISA) |
| Medtronic Valleylab FX8 Energy Platform (VLFX8GEN) | cpe:2.3:o:medtronic:valleylab_fx8_energy_platform_firmware:*:*:*:*:*:*:*:* | Software 1.1.0 and earlier (NVD, CVE record; CISA's product list says 1.0.0 and earlier); Medtronic named early 2020 for the FX8 patch (CISA, 2019-11-07) |
| Medtronic Valleylab Exchange Client | cpe:2.3:a:medtronic:valleylab_exchange_client:*:*:*:*:*:*:*:* | 3.4 and earlier (NVD, CVE record, CISA) |
What does this mean for affected devices?
CVE-2019-13539 is a weak password hash in Medtronic's Valleylab energy platforms. The CVE record, published by CISA on 2019-11-08, says Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below "use the descrypt algorithm for OS password hashing", and that while interactive network logons are disabled, attackers "can use the other vulnerabilities within this report to obtain local shell access and access these hashes" (NVD, CVE record). NVD's Primary score is 7.8, CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H, with CWE-326, inadequate encryption strength. CISA, the CNA, scores 7.0, CVSS:3.1/AV:L/AC:H/PR:L/UI:N/S:U/C:H/I:H/A:H, and uses CWE-328, use of weak hash. The only difference is attack complexity: CISA rates it high because the hashes are reachable only after another flaw gives a shell.
Which Medtronic devices are affected?
CISA's medical advisory ICSMA-19-311-02, released on 2019-11-07, covers the FT10 and FX8 platforms and the Exchange Client in the Healthcare and Public Health sector, worldwide. Its product list gives FX8 software 1.0.0 and below, while the CVE record and NVD say 1.1.0 and below. Two sibling CVEs are in the same advisory: CVE-2019-13543, multiple sets of hard-coded credentials that can be used to read files on the device (CISA 5.8), and CVE-2019-3464 with CVE-2019-3463, a vulnerable rssh utility used for file uploads that could give administrative file access or code execution (CISA 9.8). CISA's risk summary is that together they may allow a remote, non-root shell. CISA notes that network connections on these devices are disabled by default and the Ethernet port is disabled on reboot, but that network connectivity is often enabled. Medtronic reported the issues to CISA. A separate advisory, ICSMA-19-311-01, covers RFID issues in the FT10.
How do you fix CVE-2019-13539?
CISA reported in November 2019 that Medtronic software patches were available for the FT10 and would be available in early 2020 for the FX8, downloadable from Medtronic's Covidien software support site. Until a device is patched, Medtronic recommends either disconnecting it from IP networks or segregating those networks so the device cannot be reached from an untrusted network such as the internet. Hospitals should confirm the installed software version with Medtronic service and keep the Ethernet port disabled when no data transfer is needed. The weak cryptography fix guide covers modern password hashing for device makers, and the FDA postmarket cybersecurity guide covers manufacturer obligations. Other connected medical devices with CISA advisories include the Medtronic MyCareLink Smart reader (CVE-2020-25187) and the BD Alaris PC Unit (CVE-2020-25165).
Is CVE-2019-13539 exploited?
CVE-2019-13539 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). CISA reported no known public exploits specifically targeting these vulnerabilities at release. FIRST EPSS gives it 0.00268, the 11.85th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).


Evidence
- NVD (read 2026-11-27): status Modified, published 2019-11-08; [email protected] Primary CVSS 3.1 7.8 AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H, CVSS 2.0 7.2; ICS-CERT Secondary 7.0 AV:L/AC:H/PR:L/UI:N/S:U/C:H/I:H/A:H; CWE-326 (NVD), CWE-328 (ICS-CERT); CPEs valleylab_exchange_client <= 3.4, valleylab_ft10_energy_platform_firmware <= 4.0.0, valleylab_fx8_energy_platform_firmware <= 1.1.0. — NVD
- CISA ICSMA-19-311-02 (2019-11-07): FX8 software <= 1.0.0, FT10 <= 4.0.0, Exchange Client <= 3.4; CVE-2019-13543 CWE-798 5.8, CVE-2019-13539 CWE-328 7.0, CVE-2019-3464/3463 rssh CWE-20 9.8; network connections disabled by default, Ethernet port disabled on reboot, connectivity often enabled; FT10 patches available, FX8 in early 2020; disconnect or segregate networks; reported by Medtronic. — CISA CSAF
- CVE record (CNA ICS-CERT, updated 2025-05-22): CVSS 3.1 7.0 AV:L/AC:H/PR:L/UI:N/S:U/C:H/I:H/A:H; CWE-328; Exchange Client <= 3.4, FT10 <= 4.0.0, FX8 <= 1.1.0. — CVE.org
- CISA KEV catalog 2026.11.25 (1,805 entries): not listed, checked 2026-11-27. — CISA KEV
- EPSS 0.00268, percentile 0.11850, dataset 2026-11-26. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which Medtronic products are affected by CVE-2019-13539?
Valleylab FT10 Energy Platform software 4.0.0 and earlier, Valleylab FX8 Energy Platform software 1.1.0 and earlier (CISA's product list says 1.0.0) and Valleylab Exchange Client 3.4 and earlier (CVE record, NVD, CISA ICSMA-19-311-02).
Can CVE-2019-13539 be exploited over the network on its own?
No. Interactive network logons are disabled, so an attacker first needs local shell access, which CISA says the sibling flaws in the same advisory can provide. That is why both NVD and CISA use a local attack vector.
How do hospitals mitigate CVE-2019-13539?
Install Medtronic's software patch (released for the FT10 in 2019; Medtronic named early 2020 for the FX8) and, until then, disconnect the platforms from IP networks or segregate those networks from untrusted ones, per CISA's advisory.
Is CVE-2019-13539 exploited in the wild?
It is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.00268, the 11.85th percentile, in the 2026-11-26 dataset.