Security Advisory

CVE-2020-25187: MyCareLink Smart Reader Heap Overflow

Last updated:
  • CVE-2020-25187

CVE-2020-25187 is a heap overflow in the Medtronic MyCareLink Smart Model 25000 Patient Reader, which relays data from an implanted cardiac device. An authenticated attacker can send a debug command that overflows the heap and may run code on the reader. NVD scores it 9.8, CISA 8.8. Updating the MyCareLink Smart app to 5.2.0 or later updates the reader.

Vulnerability summary

CVSS base score

9.8

Critical severity

EPSS score (FIRST)

3.9%

Probability of exploitation in the wild in the next 30 days

CISA KEV catalog

No

Not listed

Published

Vulnerability ID: CVE-2020-25187 · Source: NVD

CVSS v3.1 severity rating scale: None 0.0, Low 0.1 to 3.9, Medium 4.0 to 6.9, High 7.0 to 8.9, Critical 9.0 to 10.0
How a CVSS v3.1 base score maps to a severity rating. Source: FIRST CVSS v3.1 specification, section 5.

Affected components

ComponentCPEVersion range
Medtronic MyCareLink Smart Model 25000 Patient Readercpe:2.3:o:medtronic:mycarelink_smart_model_25000_firmware:*:*:*:*:*:*:*:*All versions before the December 2020 update (CISA ICSMA-20-345-01: all versions); fixed by updating the MyCareLink Smart app to 5.2.0 or later, which updates the reader on next use (Medtronic bulletin)

What does this mean for affected devices?

The CVE record from CISA says the MyCareLink Smart 25000 is "vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack", and that the overflow "could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device". NVD's Primary score is 9.8 (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H) with CWE-787, out-of-bounds write. CISA, the CNA, scores it 8.8 (CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H) with CWE-122, heap-based buffer overflow. The difference is the attack vector: CISA's advisory says exploitation "must be initiated within Bluetooth signal proximity" and that the vulnerabilities are not exploitable remotely, which matches CISA's adjacent-network vector rather than NVD's network vector (NVD, CVE record).

Which Medtronic devices are affected?

CISA's medical advisory ICSMA-20-345-01, released on 2020-12-08, lists all versions of the Smart Model 25000 Patient Reader used with the MyCareLink Smart app, in the Healthcare and Public Health sector, deployed worldwide. Sternum discovered the flaws and reported them to Medtronic. Two sibling CVEs affect the same reader:

  • CVE-2020-25183, an authentication protocol flaw that lets another phone or a malicious app pass as the patient's MyCareLink Smart app (CISA 8.0);
  • CVE-2020-27252, a race condition in the reader's software update system that allows unsigned firmware to run (CISA 8.8).

CISA's risk summary is that, chained together, the flaws could let an attacker modify or fabricate data uploaded from the implanted cardiac device to the CareLink Network and run code on the reader, which could allow control of a paired cardiac device.

How do you fix CVE-2020-25187?

Medtronic's bulletin MyCareLink Smart security vulnerability patch, published on 2020-12-10, tells patients to update the MyCareLink Smart app to version 5.2.0 or higher from the Apple App Store or Google Play; CISA adds that the updated app updates the Patient Reader on its next use and that the phone needs iOS 10 or Android 6.0 or later. Medtronic also recommends turning on automatic app updates, keeping physical control of the reader, obtaining devices only from authorised providers and keeping the phone's operating system current. CISA records that Medtronic also applied additional controls for monitoring and responding to improper use of the reader, and Medtronic says no cyberattack, unauthorised access to patient data or patient harm had been observed. For manufacturers, the memory corruption fix guide covers this bug class and the Bluetooth LE security guide covers pairing and authentication between a reader and a phone app. Other connected medical devices with CISA advisories include the BD Alaris PC Unit (CVE-2023-30560) and the B. Braun Battery Pack SP with Wi-Fi (CVE-2023-0888).

Is CVE-2020-25187 exploited?

CVE-2020-25187 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.03931, the 91.14th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).

Three signals on every advisory: CVSS says how severe the flaw is, EPSS gives the probability of exploitation in the wild in the next 30 days, and a CISA KEV listing means the flaw has been exploited in the wild
CVSS rates severity, EPSS estimates the likelihood of exploitation, and CISA KEV records exploitation seen in the wild. Source: FIRST EPSS; CISA KEV catalog.
How to fix this class of issueThis advisory's weakness class is Memory corruption. Our practical guide covers how to detect, fix, verify and disclose it: How to fix memory corruption in embedded firmware.
From advisory to verified fix in four steps: read the advisory, find the affected component in the firmware SBOM, apply the fixed version or vendor mitigation, then re-scan the image to confirm the vulnerable version is gone
Read the advisory, find the component in your firmware SBOM, apply the fix, then re-scan to confirm it.

Evidence

  • NVD (read 2026-11-27): status Modified; published 2020-12-14; [email protected] Primary CVSS 3.1 9.8 AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, CVSS 2.0 10.0; ICS-CERT Secondary 8.8 AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H; CWE-787 (NVD), CWE-122 (ICS-CERT); CPE medtronic mycarelink_smart_model_25000_firmware (no version). — NVD
  • CISA ICSMA-20-345-01 (2020-12-08): Smart Model 25000 Patient Reader all versions; CVE-2020-25183 (CWE-287, 8.0), CVE-2020-25187 (CWE-122, 8.8), CVE-2020-27252 (CWE-367, 8.8); exploitation must be initiated within Bluetooth signal proximity, not exploitable remotely; fix via app v5.2, reader updated on next use, iOS 10+/Android 6.0+; reported by Sternum. — CISA CSAF
  • Medtronic bulletin (2020-12-10): update MyCareLink Smart app to version 5.2.0 or higher via App Store/Google Play; enable auto-update, keep physical control, authorised providers only, keep OS current; no cyberattack, no unauthorized access to patient data and no patient harm observed. — Medtronic
  • CVE.org record (CNA icscert, updated 2025-05-22): CVSS 3.1 8.8 AV:A; CWE-122; Smart Model 25000 Patient Reader all versions; credit Sternum. — CVE.org
  • CISA KEV catalog version 2026.11.25, 1,805 entries: CVE-2020-25187 not listed; checked 2026-11-27. — CISA KEV
  • EPSS 0.03931, percentile 0.91143, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories

Sources

  1. NVD
  2. CISA ICSMA-20-345-01: Medtronic MyCareLink Smart (CSAF)
  3. CISA ICSMA-20-345-01 advisory page
  4. Medtronic: MyCareLink Smart security vulnerability patch
  5. CVE.org record CVE-2020-25187
  6. CISA KEV
  7. FIRST EPSS

Frequently asked questions

Which Medtronic products are affected by CVE-2020-25187?

The MyCareLink Smart Model 25000 Patient Reader, all versions before the December 2020 update, per CISA ICSMA-20-345-01 and Medtronic's bulletin.

How is CVE-2020-25187 fixed?

Update the MyCareLink Smart app to version 5.2.0 or higher. The updated app updates the Patient Reader the next time it is used; the phone needs iOS 10 or Android 6.0 or later (Medtronic bulletin, CISA ICSMA-20-345-01).

Why does NVD score CVE-2020-25187 9.8 while CISA gives 8.8?

NVD uses a network attack vector with no user interaction. CISA uses an adjacent-network vector with user interaction and changed scope, because exploitation has to start within Bluetooth range of the reader.

Is CVE-2020-25187 exploited in the wild?

It is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries), and Medtronic reported no cyberattack or patient harm. FIRST EPSS rates it 0.03931, the 91.14th percentile, in the 2026-11-26 dataset.