Security Advisory

CVE-2018-10597: Philips IntelliVue Monitor Memory Write

Last updated:
  • CVE-2018-10597

CVE-2018-10597 is a memory write flaw in Philips IntelliVue MP and MX patient monitors and Avalon FM20 to FM50 fetal monitors. An unauthenticated attacker on the same subnet can write memory at a device address of their choosing, which may change monitor memory or force a restart. NVD and CISA both score it 8.3. Philips points IntelliVue users to Revision K.2 or newer.

Vulnerability summary

CVSS base score

8.3

High severity

EPSS score (FIRST)

0.4%

Probability of exploitation in the wild in the next 30 days

CISA KEV catalog

No

Not listed

Published

Vulnerability ID: CVE-2018-10597 · Source: NVD

CVSS v3.1 severity rating scale: None 0.0, Low 0.1 to 3.9, Medium 4.0 to 6.9, High 7.0 to 8.9, Critical 9.0 to 10.0
How a CVSS v3.1 base score maps to a severity rating. Source: FIRST CVSS v3.1 specification, section 5.

Affected components

ComponentCPEVersion range
Philips IntelliVue MP series patient monitors (MP2, X2, MP30, MP50, MP70, NP90, MX700, MX800)cpe:2.3:o:philips:intellivue_mp70_firmware:-:*:*:*:*:*:*:*Rev B-M (CISA ICSMA-18-156-01, CVE record); Philips recommends Revision K.2 or newer, patch for J-M, upgrade path for earlier revisions
Philips IntelliVue MX patient monitors (MX400-MX550; X3 and MX100 at Rev M)cpe:2.3:o:philips:intellivue_mx450_firmware:-:*:*:*:*:*:*:*MX400-550 Rev J-M; X3/MX100 Rev M only (CISA, CVE record); same Philips remediation
Philips Avalon Fetal/Maternal Monitors FM20, FM30, FM40, FM50cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm30_firmware:-:*:*:*:*:*:*:*Software revisions F.0, G.0 and J.3 (CISA, CVE record); patch for G.0 and J.3, network security requirements in the service documentation for F.0

What does this mean for affected devices?

The CVE record from CISA says that the affected IntelliVue and Avalon monitors "have a vulnerability that allows an unauthenticated attacker to access memory ('write-what-where') from an attacker-chosen device address within the same subnet." NVD maps it to CWE-787, out-of-bounds write; CISA classes it as CWE-287, improper authentication (NVD, CVE record).

How severe is CVE-2018-10597?

NVD's Primary score is CVSS v3.1 8.3 with CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H; CISA published the same metrics under CVSS v3.0, and NVD's CVSS v2 score is 5.4. The adjacent-network vector is the limiting factor: the attacker has to be on the monitor's own subnet, and CISA notes a high skill level is needed. The impact is what keeps the score high. CISA's risk note says an attacker may read or write memory or restart the monitor, which could delay diagnosis and treatment of patients.

Which monitors are affected and what should hospitals do?

CISA's ICSMA-18-156-01, released on 2018-06-05, lists IntelliVue MP series monitors (MP2, X2, MP30, MP50, MP70, NP90, MX700, MX800) at Rev B-M, IntelliVue MX400-MX550 at Rev J-M plus X3 and MX100 at Rev M, and Avalon FM20, FM30, FM40 and FM50 fetal/maternal monitors at software F.0, G.0 and J.3. Philips announced a patch for IntelliVue Revisions J-M and Avalon G.0 and J.3, and an upgrade path to a supported revision for older IntelliVue software. For IntelliVue it recommends Revision K.2 or newer and the physical and logical controls in its Security for Clinical Networks Guide; for Avalon it points to the data privacy and network security requirements in the service documentation. Field change and service bulletins are in Philips' InCenter portal. The researcher credit goes to Oran Avraham of Medigate, together with Philips. Two siblings sit in the same advisory: CVE-2018-10599 (6.4), an unauthenticated memory read, and CVE-2018-10601 (8.2), a stack overflow in an exposed echo service. Because the attack needs the same subnet, segmentation of the monitoring network is the main control; the IEC 80001-1 guide covers that, the FDA postmarket guidance the patch cycle, and the memory corruption fix guide the code-level class.

Other patient-monitoring pages: CVE-2022-38100 (Contec CMS8000) and CVE-2021-43935 (Welch Allyn cardiology software).

Is CVE-2018-10597 exploited?

CVE-2018-10597 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries), and CISA's advisory noted no known public exploit. FIRST EPSS gives it 0.00395, the 25.28th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).

Three signals on every advisory: CVSS says how severe the flaw is, EPSS gives the probability of exploitation in the wild in the next 30 days, and a CISA KEV listing means the flaw has been exploited in the wild
CVSS rates severity, EPSS estimates the likelihood of exploitation, and CISA KEV records exploitation seen in the wild. Source: FIRST EPSS; CISA KEV catalog.
How to fix this class of issueThis advisory's weakness class is Memory corruption. Our practical guide covers how to detect, fix, verify and disclose it: How to fix memory corruption in embedded firmware.
From advisory to verified fix in four steps: read the advisory, find the affected component in the firmware SBOM, apply the fixed version or vendor mitigation, then re-scan the image to confirm the vulnerable version is gone
Read the advisory, find the component in your firmware SBOM, apply the fix, then re-scan to confirm it.

Evidence

  • NVD (read 2026-11-27): status Modified; published 2018-06-05; NVD Primary CVSS v3.1 8.3 CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H; NVD Primary CVSS v2 5.4 AV:A/AC:M/Au:N/C:P/I:P/A:P; CWE-787 (NVD Primary), CWE-287 (ICS-CERT); CPEs for IntelliVue MP2, X2, MP30, MP50, MP70, NP90, MX700, MX800, MX400, MX450, MX500, MX550, X3, MX100 and Avalon FM20/FM30/FM40/FM50 firmware without version ranges. — NVD
  • CISA ICSMA-18-156-01 CSAF (released 2018-06-05, read 2026-11-27): IntelliVue MP series Rev B-M, MX400-550 Rev J-M and X3/MX100 Rev M, Avalon FM20-FM50 F.0, G.0, J.3; CVE-2018-10597 CWE-287 8.3 CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H; CVE-2018-10599 CWE-200 6.4; CVE-2018-10601 CWE-121 8.2; read/write memory or restart, potential delay in diagnosis and treatment; same-subnet only, high skill; Philips patch for supported versions and upgrade path, bulletins via InCenter; reported by Oran Avraham (Medigate) and Philips; no known public exploits. — CISA ICSMA-18-156-01
  • CISA advisory page (read 2026-11-27 via a page fetch): Philips remediation patch in 2018 for IntelliVue Revisions J-M and Avalon G.0 and J.3, upgrade path for earlier IntelliVue revisions; IntelliVue users to follow the Security for Clinical Networks Guide and update to Revision K.2 or newer; Avalon G.0/J.3 and F.0 users to follow the data privacy and network security requirements in the service documentation. — CISA
  • CVE record (read 2026-11-27): state PUBLISHED, CNA ICS-CERT, published 2018-06-05; affected IntelliVue MP series Rev B-M, MX400-550 Rev J-M, X3/MX100 Rev M, Avalon FM20-FM50 F.0, G.0, J.3; no CNA CVSS in the record. — CVE.org
  • Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
  • EPSS 0.00395, percentile 0.25275, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories

Sources

  1. NVD
  2. CISA ICSMA-18-156-01: Philips' IntelliVue Patient and Avalon Fetal Monitors (CSAF)
  3. CISA ICSMA-18-156-01 advisory page
  4. CVE.org record CVE-2018-10597
  5. CISA KEV
  6. FIRST EPSS

Frequently asked questions

What should IntelliVue users install for CVE-2018-10597?

Philips recommends Revision K.2 or newer software and announced a patch for Revisions J-M, with an upgrade path for older revisions. Field change and service bulletins are in Philips' InCenter portal (CISA ICSMA-18-156-01).

Can CVE-2018-10597 be exploited from the internet?

Not directly. The attacker has to be on the same subnet as the monitor, and CISA notes a high skill level is needed. That is why segmentation of the patient-monitoring network is the main compensating control.

What is the clinical impact?

CISA's risk evaluation says an attacker may read or write monitor memory or force a restart, which could delay diagnosis and treatment of patients.

Is CVE-2018-10597 in the CISA KEV catalog?

No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.00395 (25.28th percentile) in the 2026-11-26 dataset.