CVE-2022-38100 lets anyone on the network crash a Contec Health CMS8000 patient monitor with one malformed UDP request; the monitor stays down until someone reboots it by hand, and a broadcast can hit every CMS8000 on the segment. NVD and CISA score it 7.5. Contec released no fix, and CISA recommends taking the monitors off the network.
Vulnerability summary
CVSS base score
7.5
High severity
EPSS score (FIRST)
1.0%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2022-38100 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| Contec Health CMS8000 ICU/CCU vital signs patient monitor firmware | cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:* | smart3250-2.6.27-wlan2.1.7.cramfs and earlier, CMS7.820.075.08/0.740.75 and earlier, CMS7.820.120.01/0.930.95 and earlier (CISA Update A); no vendor fix; CISA: remove from networks |
What does this mean for affected devices?
The CVE record from CISA says the monitor "fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack" on all CMS8000 devices on the same network. NVD's Primary score and CISA's score are both 7.5 (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H). CISA's current advisory classifies it as CWE-770, allocation of resources without limits; NVD's own weakness entry is NVD-CWE-Other. NVD's configuration names CMS8000 firmware without a version (NVD, CVE record).
What is affected?
CISA's ICSMA-22-244-01 was released on 2022-09-01 and revised as Update A on 2025-01-30, which reworked the risk evaluation, CWEs and mitigations. It covers the CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor with firmware smart3250-2.6.27-wlan2.1.7.cramfs, CMS7.820.075.08/0.740.75 and CMS7.820.120.01/0.930.95 and earlier. CISA lists the Healthcare and Public Health sector, worldwide deployment and a headquarters in China. Level Nine reported the flaws.
The advisory has five more CVEs in the same monitor, most needing physical access: firmware modification through a USB drive (CVE-2022-36385), hard-coded credentials (CVE-2022-38069), debug symbols left in binaries (CVE-2022-38453) and an unsanitized Wi-Fi SSID name (CVE-2022-3027). A separate 2025 CISA advisory covers a backdoor and an out-of-bounds write in the same product, described on the CVE-2024-12248 page.
How do you fix CVE-2022-38100?
There is no vendor fix: Contec Health did not respond to CISA's requests. In Update A, CISA recommends, per FDA recommendation, removing any Contec CMS8000 devices from networks. The original mitigations from CISA's CVE record (disable UART at the CPU level, require authentication before terminal or bootloader access, secure boot where possible, tamper stickers) address the physical-access flaws, not this network crash. Where a monitor cannot be replaced at once, run it standalone without a network connection and plan its replacement.
The FDA postmarket cybersecurity guide explains what manufacturers owe when a device flaw has no fix, and the input handling fix guide covers parsing malformed network data without crashing.
Is CVE-2022-38100 exploited?
CVE-2022-38100 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.00962, percentile 60.28, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).


Evidence
- NVD (read 2026-11-27): status Modified; published 2022-09-13; NVD Primary CVSS v3.1 7.5 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H; ICS-CERT Secondary 7.5; CWE NVD-CWE-Other (NVD), CWE-400 (ICS-CERT); CPE contechealth cms8000_firmware (no version). — NVD
- CISA ICSMA-22-244-01 CSAF (released 2022-09-01, Update A 2025-01-30, read 2026-11-27): CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor <= smart3250-2.6.27-wlan2.1.7.cramfs, <= CMS7.820.075.08/0.740.75, <= CMS7.820.120.01/0.930.95; CVE-2022-38100 CWE-770 7.5; siblings 36385, 38069, 38453, 3027; mitigation: per FDA recommendation remove CMS8000 devices from networks; Healthcare and Public Health, worldwide, China; reported by Level Nine. — CISA ICSMA-22-244-01
- CVE record (checked 2026-11-27): PUBLISHED, CNA ICS-CERT, CNA CVSS v3.1 7.5; Contec Health did not respond; mitigations: disable UART, unique authentication before terminal/bootloader, secure boot, tamper stickers. — CVE.org
- Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
- EPSS 0.00962, percentile 0.60277, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Is there a patch for CVE-2022-38100?
No. Contec Health did not respond to CISA. In ICSMA-22-244-01 Update A (2025-01-30) CISA recommends, per FDA recommendation, removing any Contec CMS8000 devices from networks.
What happens when CVE-2022-38100 is triggered?
The monitor crashes and needs a physical reboot. A UDP broadcast can crash every CMS8000 on the same network at once, which interrupts patient monitoring.
Which firmware versions are affected?
CISA lists smart3250-2.6.27-wlan2.1.7.cramfs, CMS7.820.075.08/0.740.75 and CMS7.820.120.01/0.930.95 and earlier. NVD's configuration names the CMS8000 firmware without a version.
Is CVE-2022-38100 in the CISA KEV catalog?
No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.00962 (percentile 60.28) in the 2026-11-26 dataset.
Questions about this page?