Security Advisory

CVE-2021-43935: Welch Allyn Cardio SSO Auth Bypass

Last updated:
  • CVE-2021-43935

CVE-2021-43935 is an authentication bypass in Hillrom Welch Allyn cardiology systems, including Q-Stress, X-Scribe, R-Scribe and H-Scribe, when they use single sign-on. A user can type in any provisioned Active Directory account and get in without its password. NVD scores it 9.8, CISA 8.1. Hillrom's mitigation is to disable SSO in Modality Manager.

Vulnerability summary

CVSS base score

9.8

Critical severity

EPSS score (FIRST)

1.1%

Probability of exploitation in the wild in the next 30 days

CISA KEV catalog

No

Not listed

Published

Vulnerability ID: CVE-2021-43935 · Source: NVD

CVSS v3.1 severity rating scale: None 0.0, Low 0.1 to 3.9, Medium 4.0 to 6.9, High 7.0 to 8.9, Critical 9.0 to 10.0
How a CVSS v3.1 base score maps to a severity rating. Source: FIRST CVSS v3.1 specification, section 5.

Affected components

ComponentCPEVersion range
Welch Allyn Q-Stress Cardiac Stress Testing Systemcpe:2.3:o:baxter:welch_allyn_q-stress_cardiac_stress_testing_system_firmware:*:*:*:*:*:*:*:*6.0.0 through 6.3.1 when configured for SSO (CISA ICSMA-21-343-01, NVD)
Welch Allyn X-Scribe Cardiac Stress Testing Systemcpe:2.3:o:baxter:welch_allyn_xscribe_cardiac_stress_testing_system_firmware:*:*:*:*:*:*:*:*5.01 through 6.3.1 when configured for SSO
Welch Allyn R-Scribe Resting ECG Systemcpe:2.3:a:baxter:welch_allyn_rscribe_resting_ecg_system:*:*:*:*:*:*:*:*5.01 through 7.0.0 when configured for SSO
Welch Allyn H-Scribe Holter Analysis Systemcpe:2.3:o:baxter:welch_allyn_hscribe_holter_analysis_system_firmware:*:*:*:*:*:*:*:*5.01 through 6.4.0 when configured for SSO
Welch Allyn Vision Expresscpe:2.3:a:baxter:welch_allyn_vision_express_holter_analysis_system:*:*:*:*:*:*:*:*6.1.0 through 6.4.0 when configured for SSO
Welch Allyn Diagnostic Cardiology Suitecpe:2.3:a:baxter:welch_allyn_diagnostic_cardiology_suite:2.1.0:*:*:*:*:*:*:*2.1.0 when configured for SSO
Welch Allyn Connex Cardiocpe:2.3:a:baxter:welch_allyn_connex_cardio:*:*:*:*:*:*:*:*1.0.0 through 1.1.1 when configured for SSO; for all products CISA lists no fixed version, mitigation: disable SSO in Modality Manager

What does this mean for affected devices?

The CVE record from CISA says that the impacted products, "when configured to use SSO," accept "manual entry of any active directory (AD) account provisioned in the application without supplying a password, resulting in access to the application as the supplied AD account, with all associated privileges." NVD classifies it as CWE-287 (improper authentication) and scores it 9.8, CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H. CISA, the CNA, files it as CWE-288 (authentication bypass using an alternate path) and scores it 8.1, CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H; the difference is attack complexity, high for CISA and low for NVD. NVD published the record on 2021-12-15 (NVD, CVE record).

Which Welch Allyn products are affected?

CISA's medical advisory ICSMA-21-343-01, released on 2021-12-09, lists these products, vulnerable only when configured for SSO:

  • Welch Allyn Q-Stress Cardiac Stress Testing System 6.0.0 to 6.3.1
  • Welch Allyn X-Scribe Cardiac Stress Testing System 5.01 to 6.3.1
  • Welch Allyn R-Scribe Resting ECG System 5.01 to 7.0.0
  • Welch Allyn H-Scribe Holter Analysis System 5.01 to 6.4.0
  • Welch Allyn Vision Express 6.1.0 to 6.4.0
  • Welch Allyn Diagnostic Cardiology Suite 2.1.0
  • Welch Allyn Connex Cardio 1.0.0 to 1.1.1

CISA lists the Healthcare and Public Health sector and worldwide deployment. Hillrom, headquartered in the United States, reported the vulnerability to CISA itself. NVD's configuration lists the products under the vendor name Baxter, and the Hillrom disclosure page that CISA links now redirects to Baxter's product security page, which as of 2026-11-27 lists no separate bulletin for these cardiology products.

How do you fix CVE-2021-43935?

At release, the advisory said Hillrom would address the flaw in its next software release and asked users to move to the latest product versions once updated products were available; CISA has not revised the advisory with fixed version numbers. Hillrom's interim measures (CISA and the CVE record) are:

  • disable the SSO feature in the respective Modality Manager configuration settings, following the instructions for use or service manual;
  • apply proper network and physical security controls;
  • apply authentication for server access.

Ask Baxter for the current software release of each cardiology system and confirm whether it still offers SSO through Modality Manager before you turn it back on. Other medical device flaws on this site include CVE-2023-30560 in BD Alaris infusion systems and CVE-2023-40704 in Philips Vue PACS. Our authentication bypass fix guide explains the design error, and the FDA postmarket cybersecurity guide covers what manufacturers owe hospitals after a disclosure like this.

Is CVE-2021-43935 exploited?

CVE-2021-43935 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.01145, the 66.51th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).

Three signals on every advisory: CVSS says how severe the flaw is, EPSS gives the probability of exploitation in the wild in the next 30 days, and a CISA KEV listing means the flaw has been exploited in the wild
CVSS rates severity, EPSS estimates the likelihood of exploitation, and CISA KEV records exploitation seen in the wild. Source: FIRST EPSS; CISA KEV catalog.
How to fix this class of issueThis advisory's weakness class is Authentication bypass. Our practical guide covers how to detect, fix, verify and disclose it: How to fix authentication bypass in embedded firmware.
From advisory to verified fix in four steps: read the advisory, find the affected component in the firmware SBOM, apply the fixed version or vendor mitigation, then re-scan the image to confirm the vulnerable version is gone
Read the advisory, find the component in your firmware SBOM, apply the fix, then re-scan to confirm it.

Evidence

  • NVD (read 2026-11-27): status Modified; published 2021-12-15; NVD Primary CVSS v3.1 9.8 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, CVSS v2 6.8; ICS-CERT Secondary 8.1 CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H; NVD CWE-287, ICS-CERT CWE-288; CPEs under vendor baxter for Connex Cardio 1.0.0-1.1.1, Diagnostic Cardiology Suite 2.1.0, R-Scribe 5.01-7.0.0, Vision Express 6.1.0-6.4.0, H-Scribe 5.01-6.4.0, Q-Stress 6.0.0-6.3.1, X-Scribe 5.01-6.3.1. — NVD
  • CISA ICSMA-21-343-01 CSAF (released 2021-12-09, read 2026-11-27): Hillrom Welch Allyn Cardio Products, same seven products and version ranges; CVE-2021-43935 CWE-288 8.1; Hillrom to release software updates in next software release, upgrade to latest versions when available, Hillrom disclosure page; Healthcare and Public Health, worldwide, HQ United States. — CISA ICSMA-21-343-01
  • CVE record (checked 2026-11-27): PUBLISHED, CNA ICS-CERT, published 2021-12-15, CNA CVSS v3.1 8.1; workarounds: disable SSO in Modality Manager configuration settings per IFU/service manual, apply network and physical security controls, apply authentication for server access; credit: Hillrom reported this vulnerability to CISA. — CVE.org
  • CISA advisory page (read 2026-11-27): products vulnerable when configured for SSO; CVSS 8.1; mitigations disable SSO in Modality Manager, network and physical security controls, authentication for server access; no fixed versions listed. — CISA ICSMA-21-343-01 page
  • Baxter product security page (read 2026-11-27; hillrom.com/en/responsible-disclosures/ redirects here): lists bulletins for other Hillrom/Baxter products, none for Welch Allyn cardiology SSO. — Baxter
  • Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
  • EPSS 0.01145, percentile 0.66506, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories

Sources

  1. NVD
  2. CISA ICSMA-21-343-01: Hillrom Welch Allyn Cardio Products (CSAF)
  3. CISA ICSMA-21-343-01 advisory page
  4. CVE.org record CVE-2021-43935
  5. Baxter product security (former Hillrom responsible disclosures)
  6. CISA KEV
  7. FIRST EPSS

Frequently asked questions

Which products are affected by CVE-2021-43935?

Welch Allyn Q-Stress 6.0.0-6.3.1, X-Scribe 5.01-6.3.1, R-Scribe 5.01-7.0.0, H-Scribe 5.01-6.4.0, Vision Express 6.1.0-6.4.0, Diagnostic Cardiology Suite 2.1.0 and Connex Cardio 1.0.0-1.1.1, only when configured for SSO (CISA ICSMA-21-343-01).

How do I mitigate CVE-2021-43935?

Disable SSO in the Modality Manager configuration settings, following the instructions for use or service manual, and apply network, physical and server authentication controls. CISA's advisory lists no fixed version numbers.

Why does NVD score it 9.8 while CISA scores it 8.1?

CISA rates attack complexity high and NVD rates it low; CISA's advisory gives no reason. Both agree on network access, no privileges and high impact.

Is CVE-2021-43935 in the CISA KEV catalog?

No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.01145 (66.51th percentile) in the 2026-11-26 dataset.