CVE-2016-8355 is a permissions flaw in Smiths Medical CADD-Solis Medication Safety Software 1.0 to 3.1, the server application that sets dosage limits for CADD infusion pumps: any authenticated user gets elevated rights on its SQL database and can change drug libraries, add or delete users and change permissions. NVD and CISA score it 9.9. Versions 3.2 and 4.1 fix it.
Vulnerability summary
CVSS base score
9.9
Critical severity
EPSS score (FIRST)
1.5%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2016-8355 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| Smiths Medical CADD-Solis Medication Safety Software | cpe:2.3:a:smiths-medical:cadd-solis_medication_safety_software:3.1:*:*:*:*:*:*:* | 1.0, 2.0, 3.0 and 3.1; fixed in 3.2 (continental US) and 4.1 (outside the continental US) (CISA ICSMA-16-306-01) |
What does this mean for affected devices?
What is CVE-2016-8355?
NVD describes it as follows: "CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates." Affected are versions 1.0, 2.0, 3.0 and 3.1. NVD's Primary score is CVSS v3.0 9.9 (critical), vector CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H, with CVSS v2 9.0 and CWE-306 as NVD's weakness (NVD). The CVE record from CISA, the CNA, describes it as incorrect permissions, and CISA's advisory uses CWE-732, Incorrect Permission Assignment for Critical Resource. NVD's configuration lists the four versions exactly.
What does CADD-Solis Medication Safety Software do?
CISA released ICSMA-16-306-01 on 2016-11-01. It describes the product as "an application that is used to set dosage limits for infusion pumps", working with the CADD-Prizm system and CADD-Solis ambulatory infusion pumps in the Healthcare and Public Health sector. The drug library the database holds defines which doses a pump will accept, so unauthorised changes to it are a patient-safety issue. Smiths Medical's own mitigating factor, quoted by NVD and CISA, is that installing a drug library update on a pump requires physical access to the pump.
The CISA CSAF record scores CVE-2016-8355 at 9.9 with the same vector as NVD, and lists a sibling, CVE-2016-8358 (CWE-300, 8.5), in the same software. CISA credits Andrew Gothard of Newcastle upon Tyne Hospitals NHS Foundation Trust with the report.
How do I fix CVE-2016-8355?
CISA states that Smiths Medical released new versions that mitigate the vulnerabilities: "Version 3.2, which is distributed within the continental US, and Version 4.1, which is released outside the continental US." Smiths Medical recommends installing the applicable version and adds hardening steps:
- strict password rules (upper and lower case, special characters, at least 8 characters)
- a managed Active Directory and a SQL Express database account with managed permissions for the application
- managed service accounts for all servers and infusion pumps
- VLAN tagging, and server hardening for both SQL and Windows servers
Least-privilege database access for application users is the class fix covered in our authentication bypass fix guide. For the regulatory side of patching medication management systems, see our FDA postmarket cybersecurity guide. Other infusion-system advisories on this site: Baxter Spectrum cleartext protocol (CVE-2020-12040), BD Alaris PC Unit session authentication (CVE-2020-25165) and Fresenius Kabi Agilia Link+ missing authentication (CVE-2021-23233).
Is CVE-2016-8355 exploited?
CVE-2016-8355 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25). FIRST EPSS gives it 0.01491, the 74th percentile, in the 2026-11-26 dataset, as of 2026-11-27 (CISA KEV, FIRST EPSS).


Evidence
- Smiths-Medical CADD-Solis Medication Safety Software 1.0, 2.0, 3.0, 3.1 grants an authenticated user elevated SQL privileges (modify drug libraries, add/delete users, change permissions); physical access to the pump required to install drug library updates; NVD Primary CVSS v3.0 9.9 AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H; v2 9.0; NVD CWE-306; CPEs for the four versions. — NVD
- ICSMA-16-306-01 released 2016-11-01: application that sets dosage limits for infusion pumps, with CADD-Prizm and CADD-Solis pumps; CVE-2016-8355 9.9 CWE-732; sibling CVE-2016-8358 (CWE-300, 8.5); fixed in 3.2 (continental US) and 4.1 (outside); hardening recommendations; Healthcare and Public Health; credited to Andrew Gothard of Newcastle upon Tyne Hospitals NHS Foundation Trust. — CISA ICSMA-16-306-01
- Not in the KEV catalog, version 2026.11.25, checked 2026-11-27. — CISA KEV
- EPSS 0.01491, 74th percentile, 2026-11-26 dataset. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which version fixes CVE-2016-8355?
CADD-Solis Medication Safety Software 3.2 within the continental US and 4.1 outside it. Versions 1.0, 2.0, 3.0 and 3.1 are affected (per CISA ICSMA-16-306-01).
Can an attacker change pump doses remotely with CVE-2016-8355?
An authenticated user can modify drug libraries in the database, but Smiths Medical states that installing a drug library update on a pump requires physical access to the pump (per NVD and CISA ICSMA-16-306-01).
What hardening does Smiths Medical recommend?
Strict password rules, a managed Active Directory, a SQL Express account with managed permissions for the application, managed service accounts for servers and pumps, VLAN tagging and SQL and Windows server hardening (per CISA ICSMA-16-306-01).
Is CVE-2016-8355 in the CISA KEV catalog?
No. It is not listed in catalog version 2026.11.25, checked 2026-11-27, and FIRST EPSS gives it 0.01491, the 74th percentile, in the 2026-11-26 dataset (per CISA KEV and FIRST EPSS).