Security Advisory

CVE-2020-12040: Baxter Spectrum Infusion Cleartext Protocol

Last updated:
  • CVE-2020-12040

CVE-2020-12040 is a cleartext, unauthenticated protocol between Baxter Spectrum infusion pumps with a Wireless Battery Module and the Baxter Gateway. Someone on the pump network can read status data or tamper with it in transit. NVD scores it 9.8, CISA 7.3. Baxter lists no software fix; it recommends a separate VLAN and strong Wi-Fi security.

Vulnerability summary

CVSS base score

9.8

Critical severity

EPSS score (FIRST)

1.0%

Probability of exploitation in the wild in the next 30 days

CISA KEV catalog

No

Not listed

Published

Vulnerability ID: CVE-2020-12040 · Source: NVD

CVSS v3.1 severity rating scale: None 0.0, Low 0.1 to 3.9, Medium 4.0 to 6.9, High 7.0 to 8.9, Critical 9.0 to 10.0
How a CVSS v3.1 base score maps to a severity rating. Source: FIRST CVSS v3.1 specification, section 5.

Affected components

ComponentCPEVersion range
Sigma Spectrum Large Volume Pump v6.x (model 35700BAX)cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:*:*:*:*:*:*:*:*v6.x (CVE record, Baxter bulletin, CISA); NVD CPE 6.0 through 6.05; no software fix listed, Baxter mitigations apply
Baxter Spectrum Large Volume Pump v8.x (model 35700BAX2)cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:8.0:*:*:*:*:*:*:*v8.x (CVE record, Baxter bulletin, CISA); NVD CPE 8.0; no software fix listed, Baxter mitigations apply
Baxter Spectrum v9.x (model 35700BAX3)—Listed for this CVE in CISA's updated advisory with Wireless Battery Module v22D19 to v22D28; Baxter's bulletin lists only v6.x and v8.x for CVE-2020-12040

What does this mean for affected devices?

The CVE record from CISA says that the Sigma Spectrum Infusion System v6.x (model 35700BAX) and the Baxter Spectrum Infusion System v8.x (model 35700BAX2) "at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data", which "could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack." NVD's Primary score is 9.8 (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, CVSS 2.0 5.0). CISA scores it much lower, 7.3 (CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L); NVD and CISA agree on CWE-319, cleartext transmission of sensitive information (NVD, CVE record).

Baxter's own product security bulletin for the advisory, first published 18 June 2020 and last updated 4 August 2022, says the Spectrum LVP with Wireless Battery Module (WBM) uses this protocol to talk to the Baxter Gateway, that the data is infusion delivery status and operational data, and that it "does not contain sensitive data (passwords) or PHI."

Which Spectrum pumps are affected?

Baxter's bulletin lists the Sigma Spectrum LVP v6.x and the Baxter Spectrum LVP v8.x for this CVE. CISA's medical advisory ICSMA-20-170-04, released on 2020-06-18 and later updated, also lists Spectrum v9.x (model 35700BAX3) for it, in the Healthcare and Public Health sector, deployed in the United States. NVD's CPE covers firmware 6.0 through 6.05 and 8.0. The advisory's siblings concern the WBM: Telnet with hard-coded credentials (CVE-2020-12045, NVD 9.8; CVE-2020-12041, NVD 9.4), FTP with hard-coded credentials (CVE-2020-12047 and CVE-2020-12043, NVD 9.8 each) and a hard-coded Biomed menu PIN (CVE-2020-12039, NVD 2.4).

How do you mitigate CVE-2020-12040?

Baxter's bulletin names software updates for the FTP and Telnet siblings only; for this protocol it lists mitigations:

  • physical controls against unauthorised access to devices;
  • isolate the Spectrum Infusion Systems on their own VLAN, which reduces the chance of a man-in-the-middle attack on the clear-text traffic;
  • use the strongest available wireless security (WPA2, EAP-TLS) for data to and from the pumps;
  • reboot the WBM after network configuration by removing it from the pump for 10 to 15 seconds;
  • monitor for or block unexpected traffic at the boundary of the Spectrum VLAN;
  • as a last resort, disable wireless operation; the pump was designed to work without network access, though drug library updates then take longer.

Other infusion-system advisories on this site are the B. Braun Battery Pack SP with Wi-Fi (CVE-2023-0888) and Fresenius Kabi Agilia Link+ (CVE-2021-23233). For manufacturers, the weak cryptography fix guide covers authenticated encryption for device protocols, and the FDA postmarket guide covers handling flaws in fielded devices. Baxter lists all its bulletins on its product security page.

Is CVE-2020-12040 exploited?

CVE-2020-12040 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.00991, the 61.39th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS). CISA reported no known public exploits specifically targeting these vulnerabilities.

Three signals on every advisory: CVSS says how severe the flaw is, EPSS gives the probability of exploitation in the wild in the next 30 days, and a CISA KEV listing means the flaw has been exploited in the wild
CVSS rates severity, EPSS estimates the likelihood of exploitation, and CISA KEV records exploitation seen in the wild. Source: FIRST EPSS; CISA KEV catalog.
How to fix this class of issueNo dedicated SoftWSP guide covers this weakness class; the fix is usually the vendor patch or firmware update referenced above. Our related guide explains how to plan and verify it: NIST SP 800-40 Rev. 4: enterprise patch management planning, explained.
From advisory to verified fix in four steps: read the advisory, find the affected component in the firmware SBOM, apply the fixed version or vendor mitigation, then re-scan the image to confirm the vulnerable version is gone
Read the advisory, find the component in your firmware SBOM, apply the fix, then re-scan to confirm it.

Evidence

  • NVD (read 2026-11-27): status Modified; published 2020-06-29; [email protected] Primary CVSS 3.1 9.8 AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, CVSS 2.0 5.0; CWE-319 (NVD and ICS-CERT); CPE baxter sigma_spectrum_infusion_system_firmware 6.0 through 6.05 and 8.0. — NVD
  • Baxter Product Security Bulletin 'Spectrum V6, V8, V9 Advisory ICSMA-20-170-04' (initial 2020-06-18, updated 2022-08-04): CVE-2020-12040 Spectrum LVP with WBM uses unencrypted, unauthenticated application protocol to the Baxter Gateway; infusion delivery status and operational data, no passwords or PHI; affected Sigma Spectrum LVP v6.x, Baxter Spectrum LVP v8.x; software updates for FTP/Telnet CVEs only; mitigations: physical controls, own VLAN, WPA2/EAP-TLS, WBM reboot after configuration, monitor/block unexpected traffic, disable wireless as last resort. — Baxter
  • CISA ICSMA-20-170-04 (CSAF, Update B, initial 2020-06-18): CVE-2020-12040 CWE-319 CVSS 3.0 7.3 AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L; known affected Spectrum v6.x 35700BAX, v8.x 35700BAX2, v9.x 35700BAX3; sector Healthcare and Public Health; deployed United States; no known public exploits. — CISA CSAF
  • Baxter product security page (read 2026-11-27) lists 'Spectrum V6, V8, V9 - ICS Advisory (ICSMA-20-170-04)' with an Access Bulletin link. NVD Primary of siblings: CVE-2020-12043/-12045/-12047 9.8, -12041 9.4, -12039 2.4. — Baxter
  • CISA KEV catalog version 2026.11.25, 1,805 entries: CVE-2020-12040 not listed; checked 2026-11-27. — CISA KEV
  • EPSS 0.00991, percentile 0.61393, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories

Sources

  1. NVD
  2. Baxter Product Security Bulletin: Spectrum V6, V8, V9 (ICSMA-20-170-04)
  3. Baxter product security
  4. CISA ICSMA-20-170-04: Baxter Sigma Spectrum Infusion Pumps (CSAF)
  5. CISA ICSMA-20-170-04 advisory page
  6. CVE.org record CVE-2020-12040
  7. CISA KEV
  8. FIRST EPSS

Frequently asked questions

Which Baxter pumps are affected by CVE-2020-12040?

The Sigma Spectrum LVP v6.x (model 35700BAX) and Baxter Spectrum LVP v8.x (model 35700BAX2) with a Wireless Battery Module, per the CVE record and Baxter's bulletin. CISA's updated advisory also lists Spectrum v9.x (model 35700BAX3).

Is there a software update for CVE-2020-12040?

Baxter's bulletin names software updates only for the FTP and Telnet siblings. For this protocol it gives mitigations: a dedicated VLAN, WPA2 or EAP-TLS Wi-Fi security, physical controls, traffic monitoring and, as a last resort, disabling wireless operation.

Does CVE-2020-12040 expose patient data?

Baxter says the clear-text data is infusion delivery status and operational data and contains no passwords or PHI; the CVE record calls it sensitive non-private data.

Is CVE-2020-12040 in the CISA KEV catalog?

No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.00991, the 61.39th percentile, in the 2026-11-26 dataset.