Security Advisory

CVE-2015-7909: Hospira Infusion Pump CE Buffer Overflow

Last updated:
  • CVE-2015-7909

CVE-2015-7909 is a stack-based buffer overflow in the Communication Engine (CE) networking module of Hospira LifeCare PCA, Plum A+ and Plum A+3 infusion pumps. Traffic sent to TCP port 5000 on CE 1.0 or earlier can crash the module and may have other impact. NVD and CISA score it 7.3. The fix is CE version 1.2 or later.

Vulnerability summary

CVSS base score

7.3

High severity

EPSS score (FIRST)

1.1%

Probability of exploitation in the wild in the next 30 days

CISA KEV catalog

No

Not listed

Published

Vulnerability ID: CVE-2015-7909 · Source: NVD

CVSS v3.1 severity rating scale: None 0.0, Low 0.1 to 3.9, Medium 4.0 to 6.9, High 7.0 to 8.9, Critical 9.0 to 10.0
How a CVSS v3.1 base score maps to a severity rating. Source: FIRST CVSS v3.1 specification, section 5.

Affected components

ComponentCPEVersion range
Hospira LifeCare PCA Infusion System 5.07 with Communication Engine 1.0 or earliercpe:2.3:h:hospira:lifecare_pca_infusion_system:5.0.7:*:*:*:*:*:*:*CE 1.0 and earlier (CISA ICSA-15-337-02; NVD CPE communication_engine up to and including 1.0); fixed in CE 1.2 or later
Hospira Plum A+ and Plum A+3 Infusion Systems 13.40 with Communication Engine 1.0 or earliercpe:2.3:o:hospira:communication_engine:*:*:*:*:*:*:*:*CE 1.0 and earlier (CISA ICSA-15-337-02); fixed in CE 1.2 or later

What does this mean for affected devices?

The CVE record from CISA says the stack-based buffer overflow in Hospira Communication Engine (CE) before 1.2, in LifeCare PCA Infusion System 5.07 and Plum A+ and Plum A+3 Infusion Systems 13.40, "allows remote attackers to cause a denial of service or possibly have unspecified other impact via traffic on TCP port 5000." NVD maps it to CWE-119 and CISA to CWE-121 (NVD, CVE record).

How severe is CVE-2015-7909?

NVD's Primary score is CVSS v3.0 7.3, CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L, with a CVSS v2 score of 7.5. CISA's score is the same 7.3. The impact ratings are low because the flaw sits in the communication module, not the pumping function, but a hospital network that reaches the pump can still disrupt its connectivity.

Which pumps are affected and what fixes it?

CISA's ICSA-15-337-02, released on 2015-12-03, lists the LifeCare PCA Infusion System 5.07 (released before July 2009) and the Plum A+ and Plum A+3 Infusion Systems 13.40 (released before March 2009), each running CE 1.0 or earlier. The LifeCare PCA is used mainly in the United States and Canada and the Plum A+ pumps worldwide, in the healthcare and public health sector. The researcher was Jeremy Richards of SAINT Corporation. The fix is to update to CE version 1.2 or later.

CISA's advice for pumps that cannot be updated right away:

  • close unused ports, including 20 and 21 (FTP), 23 (Telnet), 5000/TCP and 8443;
  • change default passwords, or verify that those ports are closed;
  • monitor network traffic to the pumps, and keep them isolated from the internet and untrusted systems, behind firewalls;
  • produce file hashes to detect unauthorized changes, and use a VPN when remote access is needed.

The memory corruption fix guide covers the code-level class, and the IMDRF legacy medical device guide covers how to manage pumps of this age. Other infusion pump pages: CVE-2020-12040 for the Baxter Spectrum and CVE-2020-25165 for the BD Alaris.

Is CVE-2015-7909 exploited?

CVE-2015-7909 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries), and CISA's advisory noted no known public exploit. FIRST EPSS gives it 0.01112, the 65.61st percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).

Three signals on every advisory: CVSS says how severe the flaw is, EPSS gives the probability of exploitation in the wild in the next 30 days, and a CISA KEV listing means the flaw has been exploited in the wild
CVSS rates severity, EPSS estimates the likelihood of exploitation, and CISA KEV records exploitation seen in the wild. Source: FIRST EPSS; CISA KEV catalog.
How to fix this class of issueThis advisory's weakness class is Memory corruption. Our practical guide covers how to detect, fix, verify and disclose it: How to fix memory corruption in embedded firmware.
From advisory to verified fix in four steps: read the advisory, find the affected component in the firmware SBOM, apply the fixed version or vendor mitigation, then re-scan the image to confirm the vulnerable version is gone
Read the advisory, find the component in your firmware SBOM, apply the fix, then re-scan to confirm it.

Evidence

  • NVD (read 2026-11-27): status Modified; published 2016-01-22; NVD Primary CVSS v3.0 7.3 CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L; NVD Primary CVSS v2 7.5; CWE-119; CPEs hospira:communication_engine up to and including 1.0, hospira:lifecare_pca_infusion_system 5.0.7. — NVD
  • CISA ICSA-15-337-02 CSAF (read 2026-11-27): LifeCare PCA 5.07, Plum A+ 13.40, Plum A+3 13.40 running CE 1.0 or earlier; CWE-121; CVSS 7.3; update to CE version 1.2 or later. — CISA ICSA-15-337-02
  • CISA advisory page (read 2026-11-27 via a page fetch): released 2015-12-03; LifeCare PCA 5.07 released prior to July 2009, Plum A+/A+3 13.40 prior to March 2009; healthcare and public health; researcher Jeremy Richards of SAINT Corporation; mitigations close ports 20/21/23/5000/8443, change default passwords, monitor traffic, isolate, file hashes, firewalls, VPN; no known public exploits. — CISA
  • CVE record (read 2026-11-27): CNA ICS-CERT; no CNA CVSS in the record. — CVE.org
  • Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
  • EPSS 0.01112, percentile 0.65608, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories

Sources

  1. NVD
  2. CISA ICSA-15-337-02: Hospira Multiple Products Buffer Overflow Vulnerability (CSAF)
  3. CISA ICSA-15-337-02 advisory page
  4. CVE.org record CVE-2015-7909
  5. CISA KEV
  6. FIRST EPSS

Frequently asked questions

Which version fixes CVE-2015-7909?

Communication Engine (CE) version 1.2 or later. Pumps running CE 1.0 or earlier are affected (CISA ICSA-15-337-02).

Which Hospira pumps are affected?

The LifeCare PCA Infusion System 5.07 and the Plum A+ and Plum A+3 Infusion Systems 13.40, each with CE 1.0 or earlier.

What can a hospital do before updating the CE?

CISA advises closing unused ports (including 5000/TCP), changing default passwords, monitoring pump traffic, isolating the pumps from the internet and untrusted systems behind firewalls, and producing file hashes to detect unauthorized changes.

Is CVE-2015-7909 in the CISA KEV catalog?

No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.01112 (65.61st percentile) in the 2026-11-26 dataset.