CVE-2015-7909 is a stack-based buffer overflow in the Communication Engine (CE) networking module of Hospira LifeCare PCA, Plum A+ and Plum A+3 infusion pumps. Traffic sent to TCP port 5000 on CE 1.0 or earlier can crash the module and may have other impact. NVD and CISA score it 7.3. The fix is CE version 1.2 or later.
Vulnerability summary
CVSS base score
7.3
High severity
EPSS score (FIRST)
1.1%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2015-7909 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| Hospira LifeCare PCA Infusion System 5.07 with Communication Engine 1.0 or earlier | cpe:2.3:h:hospira:lifecare_pca_infusion_system:5.0.7:*:*:*:*:*:*:* | CE 1.0 and earlier (CISA ICSA-15-337-02; NVD CPE communication_engine up to and including 1.0); fixed in CE 1.2 or later |
| Hospira Plum A+ and Plum A+3 Infusion Systems 13.40 with Communication Engine 1.0 or earlier | cpe:2.3:o:hospira:communication_engine:*:*:*:*:*:*:*:* | CE 1.0 and earlier (CISA ICSA-15-337-02); fixed in CE 1.2 or later |
What does this mean for affected devices?
The CVE record from CISA says the stack-based buffer overflow in Hospira Communication Engine (CE) before 1.2, in LifeCare PCA Infusion System 5.07 and Plum A+ and Plum A+3 Infusion Systems 13.40, "allows remote attackers to cause a denial of service or possibly have unspecified other impact via traffic on TCP port 5000." NVD maps it to CWE-119 and CISA to CWE-121 (NVD, CVE record).
How severe is CVE-2015-7909?
NVD's Primary score is CVSS v3.0 7.3, CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L, with a CVSS v2 score of 7.5. CISA's score is the same 7.3. The impact ratings are low because the flaw sits in the communication module, not the pumping function, but a hospital network that reaches the pump can still disrupt its connectivity.
Which pumps are affected and what fixes it?
CISA's ICSA-15-337-02, released on 2015-12-03, lists the LifeCare PCA Infusion System 5.07 (released before July 2009) and the Plum A+ and Plum A+3 Infusion Systems 13.40 (released before March 2009), each running CE 1.0 or earlier. The LifeCare PCA is used mainly in the United States and Canada and the Plum A+ pumps worldwide, in the healthcare and public health sector. The researcher was Jeremy Richards of SAINT Corporation. The fix is to update to CE version 1.2 or later.
CISA's advice for pumps that cannot be updated right away:
- close unused ports, including 20 and 21 (FTP), 23 (Telnet), 5000/TCP and 8443;
- change default passwords, or verify that those ports are closed;
- monitor network traffic to the pumps, and keep them isolated from the internet and untrusted systems, behind firewalls;
- produce file hashes to detect unauthorized changes, and use a VPN when remote access is needed.
The memory corruption fix guide covers the code-level class, and the IMDRF legacy medical device guide covers how to manage pumps of this age. Other infusion pump pages: CVE-2020-12040 for the Baxter Spectrum and CVE-2020-25165 for the BD Alaris.
Is CVE-2015-7909 exploited?
CVE-2015-7909 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries), and CISA's advisory noted no known public exploit. FIRST EPSS gives it 0.01112, the 65.61st percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).


Evidence
- NVD (read 2026-11-27): status Modified; published 2016-01-22; NVD Primary CVSS v3.0 7.3 CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L; NVD Primary CVSS v2 7.5; CWE-119; CPEs hospira:communication_engine up to and including 1.0, hospira:lifecare_pca_infusion_system 5.0.7. — NVD
- CISA ICSA-15-337-02 CSAF (read 2026-11-27): LifeCare PCA 5.07, Plum A+ 13.40, Plum A+3 13.40 running CE 1.0 or earlier; CWE-121; CVSS 7.3; update to CE version 1.2 or later. — CISA ICSA-15-337-02
- CISA advisory page (read 2026-11-27 via a page fetch): released 2015-12-03; LifeCare PCA 5.07 released prior to July 2009, Plum A+/A+3 13.40 prior to March 2009; healthcare and public health; researcher Jeremy Richards of SAINT Corporation; mitigations close ports 20/21/23/5000/8443, change default passwords, monitor traffic, isolate, file hashes, firewalls, VPN; no known public exploits. — CISA
- CVE record (read 2026-11-27): CNA ICS-CERT; no CNA CVSS in the record. — CVE.org
- Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
- EPSS 0.01112, percentile 0.65608, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which version fixes CVE-2015-7909?
Communication Engine (CE) version 1.2 or later. Pumps running CE 1.0 or earlier are affected (CISA ICSA-15-337-02).
Which Hospira pumps are affected?
The LifeCare PCA Infusion System 5.07 and the Plum A+ and Plum A+3 Infusion Systems 13.40, each with CE 1.0 or earlier.