CVE-2015-3955 is a stack-based buffer overflow in the Hospira LifeCare PCA Infusion System, version 5.0 and earlier, that can let a remote attacker execute code on the pump. NVD scores it 10.0 (CVSS v2), CISA 7.6. Hospira developed version 7.0 to address the findings in CISA's advisory, and version 5.0 pumps can be upgraded to it.
Vulnerability summary
CVSS base score
10
Critical severity
EPSS score (FIRST)
6.6%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2015-3955 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| Hospira LifeCare PCA Infusion System (PCA3, PCA5) firmware | cpe:2.3:o:hospira:lifecare_pcainfusion_firmware:*:*:*:*:*:*:*:* | NVD CPE: up to and including 5.0 (hardware LifeCare PCA3 and PCA5); CISA ICSA-15-125-01B: version 5.0 and earlier; fix: version 7.0 |
What does this mean for affected devices?
NVD's description reads: "Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors." NVD's Primary score is CVSS v2 10.0 (AV:N/AC:L/Au:N/C:C/I:C/A:C) with CWE-119. Its CPE configuration matches LifeCare PCA infusion firmware up to and including 5.0 on the LifeCare PCA3 and PCA5 hardware. ICS-CERT assigned the record with vendor and product n/a and no CNA score; it was published on 2015-07-06 (NVD, CVE record).
How does CISA's advisory score it?
CISA's advisory ICSA-15-125-01B (CSAF) maps the flaw to CWE-121, stack-based buffer overflow, and scores it CVSS v2 7.6 (AV:N/AC:H/Au:N/C:C/I:C/A:C): the higher attack complexity is the only difference from NVD's vector. The same advisory covers six more findings on the pump: CVE-2014-5406 (unvalidated drug library, software update and configuration traffic), CVE-2015-1011 (hard-coded credentials), CVE-2015-1012 (wireless keys stored in plain text), CVE-2015-3459 (root Telnet without authentication on the communication module), CVE-2015-3957 (stored private keys and certificates) and CVE-2015-3958 (denial of service that forces a manual reboot).
What is the fix?
CISA reports that Hospira developed version 7.0 of the LifeCare PCA Infusion System to address the identified findings, with port 20 (FTP) and port 23 (Telnet) closed by default. According to the advisory the new version is meant to block unauthorized remote access, stop unauthorized changes to the medication library, remove hard-coded passwords, encrypt stored wireless network keys and drop the vulnerable AppWeb versions. Pumps on version 5.0 can be upgraded to 7.0; Hospira said it would retire versions 2 and 3 by the end of 2015. At the time of the advisory, version 7.0 depended on FDA 510(k) clearance. Hospitals still running older units should keep them on an isolated clinical network segment, block FTP and Telnet to the pumps, and follow CISA's practices of firewalls and no internet exposure.
The memory corruption fix guide covers the bug class, and the FDA postmarket cybersecurity guide and IMDRF legacy device guide cover how manufacturers and hospitals handle fielded pumps. Other infusion device pages: CVE-2020-12040 and CVE-2018-14786.
CVE-2015-3955 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.0662, a percentile of 94.6, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).


Evidence
- NVD (read 2026-11-27): status Modified; published 2015-07-06; NVD Primary CVSS v2 10.0 AV:N/AC:L/Au:N/C:C/I:C/A:C; CWE-119; CPE hospira lifecare_pcainfusion_firmware up to and including 5.0, with hardware lifecare_pca3 and lifecare_pca5. — NVD
- CISA ICSA-15-125-01B CSAF (read 2026-11-27): CVE-2015-3955 CWE-121, CVSS v2 7.6 AV:N/AC:H/Au:N/C:C/I:C/A:C; product LifeCare PCA Infusion System <=5.0; siblings CVE-2014-5406, 2015-1011, 2015-1012, 2015-3459, 2015-3957, 2015-3958; remediation: Hospira developed Version 7.0 that addresses the identified vulnerabilities, Port 20/FTP and Port 23/TELNET closed by default; mitigates unauthorized remote access, unauthorized medication library changes, removes hard-coded passwords, encrypts wireless keys, no vulnerable AppWeb; 5.0 can be upgraded to 7.0; versions 2 and 3 retired by end of 2015; 510(k) under FDA review. — CISA ICSA-15-125-01B
- CVE record (checked 2026-11-27): state PUBLISHED, CNA icscert; published 2015-07-06, affected n/a / n/a; no CNA CVSS. — CVE.org
- Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
- EPSS 0.06622, percentile 0.94638, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which Hospira pumps are affected by CVE-2015-3955?
The LifeCare PCA Infusion System version 5.0 and earlier, per NVD and CISA ICSA-15-125-01B. NVD's CPE data ties the firmware to the LifeCare PCA3 and PCA5 hardware.
What does version 7.0 change?
According to CISA, version 7.0 closes FTP (port 20) and Telnet (port 23) by default, removes hard-coded passwords, encrypts stored wireless keys, blocks unauthorized drug library changes and no longer uses the vulnerable AppWeb versions. Version 5.0 pumps can be upgraded to it.
What should a hospital do with pumps it cannot upgrade?
Keep them on an isolated clinical network segment, block FTP and Telnet to the pumps at the firewall, and make sure they cannot be reached from the internet or the general hospital network, in line with CISA's recommended practices.
Is CVE-2015-3955 in the CISA KEV catalog?
No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.0662 (percentile 94.6) in the 2026-11-26 dataset.
Questions about this page?