CVE-2018-14786 is a missing authentication flaw in BD Alaris Plus syringe pumps (Alaris GS, GH, CC and TIVA, versions 2.3.6 and earlier). When a pump's serial port is connected to a terminal server, a remote attacker can gain unauthorized access and affect the intended operation of the pump. NVD and CISA score it 9.4; BD's remedy is configuration, not new software.
Vulnerability summary
CVSS base score
9.4
Critical severity
EPSS score (FIRST)
2.9%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2018-14786 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| BD Alaris GS syringe pump | cpe:2.3:o:bd:alaris_gs_firmware:*:*:*:*:*:*:*:* | 2.3.6 and earlier (CISA ICSMA-18-235-01, NVD CPE); mitigations only |
| BD Alaris GH syringe pump | cpe:2.3:o:bd:alaris_gh_firmware:*:*:*:*:*:*:*:* | 2.3.6 and earlier; mitigations only |
| BD Alaris CC syringe pump | cpe:2.3:o:bd:alaris_cc_firmware:*:*:*:*:*:*:*:* | 2.3.6 and earlier; mitigations only |
| BD Alaris TIVA syringe pump | cpe:2.3:o:bd:alaris_tiva_firmware:*:*:*:*:*:*:*:* | 2.3.6 and earlier; mitigations only |
What does this mean for affected devices?
The CVE record from CISA says the BD Alaris Plus syringe pumps "are affected by an improper authentication vulnerability where the software does not perform authentication for functionality that requires a provable user identity, where it may allow a remote attacker to gain unauthorized access to various Alaris Syringe pumps and impact the intended operation of the pump when it is connected to a terminal server via the serial port." NVD's Primary score is 9.4 (CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H, CVSS 2.0 7.5), the same vector CISA calculated. The weakness is CWE-287, improper authentication, and NVD's CPE configuration covers Alaris GS, GH, CC and TIVA firmware up to and including 2.3.6 (NVD, CVE record).
Which BD Alaris pumps are affected?
CISA's medical advisory ICSMA-18-235-01, released on 2018-08-23, covers the Alaris GS, Alaris GH, Alaris CC and Alaris TIVA syringe pumps, versions 2.3.6 and earlier. BD determined that these products are not sold in the United States; CISA lists the European Union as the deployment area and healthcare and public health as the sector. Elad Luz of CyberMDX discovered the flaw and BD reported it. BD states that the attack does not work when the pump is docked in an Alaris Gateway Workstation, that an attacker cannot switch the pump on remotely, and that no PHI or PII can be accessed through this flaw. The risk sits with the infusion itself: CISA's summary is that an attacker could affect the intended operation of the pump.
How do you mitigate CVE-2018-14786?
BD did not announce new pump software for this CVE. CISA's advisory lists BD's mitigations and compensating controls:
- do not connect the pumps through terminal servers; BD states that terminal server use is not supported;
- run the pumps in a segmented network environment or as stand-alone devices;
- connect them through the Alaris Gateway Workstation docking station, which deactivates the remote control feature.
Hospital biomedical teams should check each pump's serial connection against that list. The authentication bypass fix guide covers authenticating remote-control functions, and the IMDRF legacy medical device guide covers managing devices whose risk is controlled by configuration. Later BD Alaris flaws are CVE-2020-25165 (PC Unit session authentication) and CVE-2023-30560 (PC Unit configuration change without authentication).
Is CVE-2018-14786 exploited?
CVE-2018-14786 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.02942, the 87.87th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS). CISA reported no known public exploits specifically targeting this vulnerability at publication.


Evidence
- NVD (read 2026-11-27): status Modified; published 2018-08-23; [email protected] Primary CVSS 3.0 9.4 AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H, CVSS 2.0 7.5; CWE-287 (Primary and ICS-CERT); CPE alaris_gs/gh/cc/tiva_firmware up to and including 2.3.6. — NVD
- CISA ICSMA-18-235-01 (2018-08-23): Alaris GS, GH, CC, TIVA versions 2.3.6 and prior; CVE-2018-14786 CWE-287 CVSS 3.0 9.4; not sold in the US, deployed in the EU; healthcare sector; HQ United States; flaw not exploitable when docked in an Alaris Gateway Workstation, no remote power-on, no PHI/PII; mitigations: terminal servers unsupported, segmented or stand-alone use, Gateway Workstation docking; discovered by Elad Luz (CyberMDX); no known public exploits. — CISA CSAF
- CVE.org record (CNA icscert): published 2018-08-23; Becton, Dickinson and Company Alaris GS, GH, CC and TIVA version 2.3.6 and prior; CWE-287; no CVSS in the CNA container. — CVE.org
- CISA KEV catalog version 2026.11.25, 1,805 entries: CVE-2018-14786 not listed; checked 2026-11-27. — CISA KEV
- EPSS 0.02942, percentile 0.87869, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which BD Alaris pumps are affected by CVE-2018-14786?
Alaris Plus syringe pumps Alaris GS, Alaris GH, Alaris CC and Alaris TIVA, versions 2.3.6 and earlier. BD states these products are not sold in the United States (CISA ICSMA-18-235-01).
Is there a software update for CVE-2018-14786?
CISA's advisory lists no new pump version. BD's mitigations are to stop using terminal servers (not supported), run pumps segmented or stand-alone, and dock them in the Alaris Gateway Workstation, which deactivates remote control.
Can CVE-2018-14786 expose patient data?
BD states that no PHI or PII can be accessed through this flaw and that an attacker cannot switch the pump on remotely. The risk is to the intended operation of a running pump.
Is CVE-2018-14786 in the CISA KEV catalog?
No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.02942, the 87.87th percentile, in the 2026-11-26 dataset.