CVE-2014-5422 is a hard-coded service password in the hardware test tool software of CareFusion Pyxis SupplyStation 8.1, the automated supply cabinets hospitals use to release medical supplies. A remote attacker who knows it can gain access to the cabinet system. NVD and CISA score it 9.7. CareFusion fixed it in test tool version 1.0.16 by adding a second, dynamic login credential.
Vulnerability summary
CVSS base score
9.7
Critical severity
EPSS score (FIRST)
2.1%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2014-5422 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| CareFusion Pyxis SupplyStation system 8.1 (hardware test tool software) | cpe:2.3:h:carefusion:pyxis_supplystation:8.1:*:*:*:*:*:*:* | Hardware test tool 1.0.15 and earlier (CISA ICSA-14-288-01); fixed in 1.0.16 |
What does this mean for affected devices?
The CVE record says CareFusion Pyxis SupplyStation 8.1 with hardware test tool before 1.0.16 "has a hardcoded service password, which makes it easier for remote attackers to obtain access via unspecified vectors." NVD files it under CWE-255, credentials management; CISA's CSAF uses CWE-1392, use of default credentials, and files the sibling CVE-2014-5420 and CVE-2014-5421 under CWE-798, hard-coded credentials (NVD, CVE record).
How severe is CVE-2014-5422?
NVD's Primary score is CVSS v2 9.7, AV:N/AC:L/Au:N/C:C/I:C/A:P, and CISA's advisory gives the same 9.7; NVD has no CVSS v3 score. CISA's ICSA-14-288-01 says exploitation "may allow an attacker to remotely compromise the Pyxis SupplyStation system", and that an attacker with physical access to a compromised cabinet could remove medical supplies. The researcher was Billy Rios. CISA describes the SupplyStation as automated supply cabinets deployed in hospitals in the Healthcare and Public Health sector, with CareFusion operating in 20 countries.
What fixes it?
CareFusion released hardware test tool software version 1.0.16, which adds two-factor authentication: on top of the fixed password, each user needs a dynamic password that is specific to that user and changes often. Per CISA, CareFusion installed 1.0.16 on affected cabinets for customers with a current remote support service agreement; other customers should contact CareFusion. The same release fixes CVE-2014-5420 (hard-coded application password) and CVE-2014-5423 (debugging files). CVE-2014-5421, a hard-coded database password that needs local access to the cabinet, is still present in 1.0.16; CareFusion said it would address it in a later version. CISA also advises keeping the cabinets off the internet and behind firewalls, separate from the hospital business network.
The hard-coded credentials fix guide covers how fixed passwords are removed from device software, and the FDA postmarket cybersecurity guide covers how hospitals and makers handle fielded devices. Related pages: CVE-2014-2363, hard-coded credentials in the Morpho Itemiser 3 that Billy Rios also reported, and CVE-2018-14786, missing authentication in BD Alaris Plus syringe pumps.
Is CVE-2014-5422 exploited?
CVE-2014-5422 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries), and CISA's advisory noted no known public exploit targeting it. FIRST EPSS gives it 0.02088, the 82.25th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).


Evidence
- NVD (read 2026-11-27): status Modified; published 2014-10-19; NVD Primary CVSS v2 9.7 AV:N/AC:L/Au:N/C:C/I:C/A:P; no CVSS v3; CWE-255; CPE carefusion:pyxis_supplystation:8.1 (hardware). — NVD
- CISA ICSA-14-288-01 CSAF (read 2026-11-27): Pyxis SupplyStation system 8.1 hardware test tool <=1.0.15; CVE-2014-5422 CWE-1392 CVSS v2 9.7; 5420/5421 CWE-798; 1.0.16 adds two-factor authentication, installed for customers with a remote support agreement; database password (5421) not addressed in 1.0.16. — CISA ICSA-14-288-01
- CISA advisory page (read 2026-11-27 via a page fetch): researcher Billy Rios; automated supply cabinets in hospitals, Healthcare and Public Health sector; CareFusion in 20 countries; remote compromise; physical access could remove supplies; scores 9.7/6.4/5.5/1.7. — CISA
- CVE.org record via the CVE Services API (read 2026-11-27): CNA ICS-CERT; product and problem type n/a; no CNA CVSS in the record. — CVE.org
- Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
- EPSS 0.02088, percentile 0.82255, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which version fixes CVE-2014-5422?
Hardware test tool software 1.0.16 for Pyxis SupplyStation system 8.1. CareFusion installed it for customers with a current remote support service agreement (CISA ICSA-14-288-01).
Does 1.0.16 remove the hard-coded password?
It adds a second required credential: a dynamic, per-user password that changes often. The fixed service password alone no longer gives access.
Are other Pyxis CVEs from the same advisory still open after 1.0.16?
Yes, CVE-2014-5421, a hard-coded database password that needs local access to the cabinet. CareFusion did not address it in 1.0.16 and said a later version would.
Is CVE-2014-5422 in the CISA KEV catalog?
No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.02088 (82.25th percentile) in the 2026-11-26 dataset.