CVE-2014-5405 is a hard-coded cleartext password (CWE-259) that Hospira MedNet up to version 5.8 uses for SQL database authorization. A remote user with any valid login who knows the password can get past MedNet's access controls on the database. NVD and CISA score it 9.0 (CVSS v2). MedNet 6.1 removes the hard-coded passwords.
Vulnerability summary
CVSS base score
9
Critical severity
EPSS score (FIRST)
2.3%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2014-5405 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| Hospira MedNet | cpe:2.3:a:hospira:mednet:*:*:*:*:*:*:*:* | Versions up to and including 5.8 (CISA and NVD CPE); fixed in MedNet 6.1 |
What does this mean for affected devices?
The CVE record says "Hospira MedNet before 6.1 uses a hardcoded cleartext password to control SQL database authorization, which allows remote authenticated users to bypass intended access restrictions by leveraging knowledge of this password." CISA classifies it as CWE-259, use of a hard-coded password; NVD files it under CWE-200, information exposure (NVD; the CVE.org record carries the same text).
How severe is it?
NVD's Primary score is CVSS v2 9.0, AV:N/AC:L/Au:S/C:C/I:C/A:C, and ICS-CERT, which assigned the CVE, gives the same score and vector. The attacker needs a single authentication (Au:S) but then gains complete confidentiality, integrity and availability impact on the database. NVD has no CVSS v3 score for this 2015 record.
What is MedNet and what else is in the advisory?
MedNet is Hospira's server software for its infusion pumps; the CVE record for the sibling flaw CVE-2014-5403 describes hard-coded keys protecting data sent from the pumps to MedNet. CISA's ICSA-15-090-03, released on 2015-03-31 and credited to independent researcher Billy Rios, lists four flaws in MedNet 5.8 and earlier: CVE-2014-5401, vulnerable JBoss Enterprise Application Platform components that may allow unauthenticated code execution (CVSS v2 10.0); this CVE-2014-5405 (9.0); CVE-2014-5403, hard-coded cryptographic keys (6.8); and CVE-2014-5400, cleartext passwords in configuration files (6.8). CISA says MedNet is used worldwide in the healthcare and public health sector and that no public exploit was known at release.
What is the fix?
Hospira released MedNet 6.1, which, per the advisory, no longer uses hard-coded passwords or hard-coded cryptographic keys and no longer stores passwords in clear text; existing installations can be upgraded to 6.1. For the JBoss issue Hospira published two knowledge base articles in August 2014, "Improving Security in Hospira MedNet 5.5" and "... 5.8", and directs customers to its technical support (CSAF). Upgrade to 6.1 or later, keep the MedNet server on a segmented clinical network, and limit who holds accounts on it. The hard-coded credentials fix guide explains the coding fix, and the IMDRF legacy medical device guide covers older clinical systems. Other Hospira pages: CVE-2015-7909 (infusion pump CE buffer overflow) and CVE-2015-3955 (LifeCare PCA pump).
Is CVE-2014-5405 exploited?
It is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.02327, the 84.23rd percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).


Evidence
- NVD (read 2026-11-27): published 2015-04-03, status Modified; NVD Primary CVSS v2 9.0 AV:N/AC:L/Au:S/C:C/I:C/A:C; ICS-CERT Secondary CVSS v2 9.0 same vector; CWE-200 (NVD), CWE-259 (ICS-CERT); CPE hospira mednet up to and including 5.8. NVD description of CVE-2014-5403: hardcoded keys protect data transmission from infusion pumps. — NVD
- CISA ICSA-15-090-03 CSAF (read 2026-11-27): Hospira MedNet <=5.8; CVE-2014-5401 JBoss CWE-94 10.0; MedNet 6.1 no longer uses hard-coded passwords or keys and no longer stores passwords in clear text; existing versions can be upgraded; KB articles Improving Security in Hospira MedNet 5.5 and 5.8 (August 2014); technical support 1-800-241-4002. — CISA ICSA-15-090-03
- CISA advisory page (read 2026-11-27 via a page fetch): released 2015-03-31, last revised 2018-08-23; researcher Billy Rios; healthcare and public health sector; worldwide; CVSS v2: CVE-2014-5401 10.0, 5405 9.0, 5403 6.8, 5400 6.8; no known public exploits. — CISA
- Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
- EPSS 0.02327, percentile 0.84232, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which MedNet version fixes CVE-2014-5405?
MedNet 6.1. Per CISA ICSA-15-090-03 it no longer uses hard-coded passwords or cryptographic keys and no longer stores passwords in clear text. MedNet 5.8 and earlier are affected, and existing installations can be upgraded to 6.1.
Does an attacker need a login?
Yes, one. The CVSS v2 vector from NVD and CISA is Au:S: a remote user with a valid account who knows the hard-coded database password can bypass MedNet's intended access restrictions, with complete impact on the database.
What other MedNet flaws did CISA publish at the same time?
CVE-2014-5401 (vulnerable JBoss components, CVSS v2 10.0), CVE-2014-5403 (hard-coded cryptographic keys, 6.8) and CVE-2014-5400 (cleartext passwords in configuration files, 6.8), all in MedNet 5.8 and earlier.
Is CVE-2014-5405 in the CISA KEV catalog?
No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.02327, the 84.23rd percentile, in the 2026-11-26 dataset.