CVE-2013-2808 is a heap-based buffer overflow in the Connect broker of Philips Xper Information Management systems, the physiomonitoring, vascular monitoring and Flex Cardio components. A crafted HTTP request to TCP port 6000 can let a remote attacker run code. NVD scores it CVSS v2 9.3, and Philips fixed it in XperConnect 1.5.4.053 SP2.
Vulnerability summary
CVSS base score
9.3
Critical severity
EPSS score (FIRST)
4.3%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2013-2808 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| Philips Xper Information Management (Physiomonitoring 5, Vascular Monitoring 5, Flex Cardio servers and workstations) with XperConnect | cpe:2.3:a:philips:xperconnect:*:*:*:*:*:*:*:* | XperConnect before 1.5.4.053 SP2 (NVD CPE: xperconnect up to and including 1.5.4.053, on Xper Information Management Physiomonitoring 5, Vascular Monitoring 5 and Xper Flex Cardio); fixed in XperConnect 1.5.4.053 SP2 |
What does this mean for affected devices?
NVD's description reads: "Heap-based buffer overflow in Xper in Philips Xper Information Management Physiomonitoring 5 components, Xper Information Management Vascular Monitoring 5 components, and Xper Information Management servers and workstations for Flex Cardio products before XperConnect 1.5.4.053 SP2 allows remote attackers to execute arbitrary code via a crafted HTTP request to the Connect broker on TCP port 6000." NVD's Primary score is CVSS v2 9.3 (AV:N/AC:M/Au:N/C:C/I:C/A:C) with CWE-119; NVD has no CVSS v3 score. ICS-CERT assigned the record with vendor and product n/a and no CNA score; NVD published it on 2013-10-05 (NVD, CVE record).
Which Philips Xper systems are affected?
CISA published this as medical advisory ICSA-13-277-01 (CSAF). It names three product groups: Xper Information Management Physiomonitoring 5 system components, Xper Information Management Vascular Monitoring 5 system components, and Xper Information Management servers and workstations of the Flex Cardio product line. NVD's CPE configuration pairs each of those with xperconnect up to and including 1.5.4.053, so units on 1.5.4.053 without SP2 count as affected. CISA's CSAF classifies the flaw as CWE-122, heap-based buffer overflow, and gives the same CVSS v2 9.3 vector as NVD.
The vulnerable code is the Connect broker, a network service listening on TCP port 6000. The overflow happens while the broker handles an incoming HTTP request, so the CVSS vector requires no authentication. ICSA-13-277-01 covers only this CVE; there are no sibling CVEs in the advisory.
How do I fix CVE-2013-2808?
CISA reports that "Philips has released an update, XperConnect 1.5.4.053 SP2, that mitigates this vulnerability," and that Philips tested the update and verified that it resolves the flaw. CISA pointed customers to Philips' InCenter customer portal for the download. Where a system cannot be updated at once, block TCP port 6000 at the network boundary so that only the Xper workstations and servers that need the broker can reach it, and keep the clinical network separate from general hospital and internet traffic, in line with CISA's recommended practices.
For hospitals that still run Xper systems from this era, the FDA postmarket cybersecurity guide explains the manufacturer's duty to fix and disclose, and the IMDRF legacy device guide covers compensating controls for systems near end of support. The memory corruption fix guide explains the bounds checking that prevents this class of bug. Other Philips pages: CVE-2018-10597 (IntelliVue monitors) and CVE-2023-40704 (Vue PACS).
As of 2026-11-27, CVE-2013-2808 is not in the CISA KEV catalog (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.04345, the 92.0th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).


Evidence
- NVD (read 2026-11-27): status Modified; published 2013-10-05; NVD Primary CVSS v2 9.3 AV:N/AC:M/Au:N/C:C/I:C/A:C; no v3 metric; CWE-119; CPEs: Xper Information Management Physiomonitoring 5, Vascular Monitoring 5 and Xper Flex Cardio, each with xperconnect up to and including 1.5.4.053; reference ICSA-13-277-01. — NVD
- CISA ICSA-13-277-01 CSAF (read 2026-11-27): CVE-2013-2808 CWE-122 Heap-based Buffer Overflow; CVSS v2 9.3 AV:N/AC:M/Au:N/C:C/I:C/A:C; affected: Physiomonitoring 5 system components, Vascular Monitoring 5 system components, Flex Cardio servers and workstations; Philips released XperConnect 1.5.4.053 SP2, tested and verified to resolve the vulnerability, via the Philips InCenter site. — CISA ICSA-13-277-01
- CVE record (checked 2026-11-27): state PUBLISHED, CNA icscert, published 2013-10-05, affected n/a / n/a, no CNA CVSS. — CVE.org
- Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
- EPSS 0.04345, percentile 0.91992, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which XperConnect version fixes CVE-2013-2808?
XperConnect 1.5.4.053 SP2. Philips tested the update and verified that it resolves the vulnerability; NVD's CPE marks XperConnect up to and including 1.5.4.053 as affected, so a system on 1.5.4.053 without SP2 still needs the update (per CISA ICSA-13-277-01 and NVD).
Which network port is exposed?
TCP port 6000, where the Xper Connect broker listens; the overflow is reached through a crafted HTTP request to that service, and the CVSS vector requires no authentication (per NVD).
Which Philips products are covered by the advisory?
Xper Information Management Physiomonitoring 5 system components, Vascular Monitoring 5 system components, and the servers and workstations of the Flex Cardio product line (per CISA ICSA-13-277-01).
Is CVE-2013-2808 in the CISA KEV catalog?
No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.04345, the 92.0th percentile, in the 2026-11-26 dataset (per CISA KEV and FIRST EPSS).
Questions about this page?