CVE-2023-40704 is a default credentials weakness in Philips Vue PACS, a medical imaging archive and viewing system, before version 12.2.8.410. Installation does not force unique, complex passwords, so a leaked Philips default password could open the PACS database. NVD scores it 9.8 but CISA 6.8, and Philips rates exploitability low. Customers can ask Philips to change the database password.
Vulnerability summary
CVSS base score
9.8
Critical severity
EPSS score (FIRST)
0.5%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2023-40704 · Source: NVD
Affected components
| Component | CPE | Version range |
|---|---|---|
| Philips Vue PACS | cpe:2.3:a:philips:vue_pacs:*:*:*:*:*:*:*:* | Before 12.2.8.410 (CISA ICSMA-24-200-01, Philips advisory, CVE record, NVD CPE); mitigation: Philips can update the database password(s) on request |
What does this mean for affected devices?
NVD carries CISA's description: "The product does not require unique and complex passwords to be created during installation. Using Philips's default password could jeopardize the PACS system if the password was hacked or leaked. An attacker could gain access to the database impacting system availability and data integrity." It was published on 2024-07-18 (NVD, CVE record).
Which Vue PACS versions are affected?
CISA's medical advisory ICSMA-24-200-01 lists Philips Vue PACS versions before 12.2.8.410, and NVD's CPE data uses the same bound. Philips published its own advisory on 2024-07-18 and updated it on 2024-11-26 (Philips). CISA's Update A of 2024-11-21 removed several vulnerabilities after further Philips analysis showed they did not affect the device or had no security impact; CVE-2023-40704 and CVE-2021-28165 remain. CISA credits TAS Health NZ and Camiel van Es, and lists the Healthcare and Public Health sector.
Why is NVD's score so much higher than CISA's?
NVD's Primary score is CVSS v3.1 9.8 (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H), a remote attack with no privileges. CISA, the CNA, scores 6.8 (CVSS:3.1/AV:A/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H): adjacent network and high privileges, and 5.7 in CVSS v4 with a local vector. CISA's advisory says these vulnerabilities are not exploitable remotely. CISA maps the flaw to CWE-1392, use of default credentials; NVD lists NVD-CWE-Other.
What does Philips recommend?
For CVE-2023-40704, Philips "recommends no action needed due to low risk of exploitability, but customers can request that Philips update database password(s)." For the other CVE, Philips points to its ports and services guide and says Vue PACS 12.2.8.410, released in October 2023, prevents it. Managed-services customers get new releases subject to resources and country regulations, through their Philips representative or the Philips Informatics Support portal. Hospitals should still request the password change, keep the PACS on a segmented clinical network, and track it in their device inventory. The DICOM and PACS security guide covers that segmentation, the FDA postmarket cybersecurity guide covers how manufacturers handle such fixes, and CVE-2025-54156 is a credential weakness in another PACS server.
CVE-2023-40704 is not in the CISA KEV catalog as of 2026-11-26 (catalog version 2026.11.25, 1,805 entries), and CISA had no reports of public exploitation when it published the advisory. FIRST EPSS gives it 0.00534, the 38.6th percentile, in the 2026-11-25 dataset (CISA KEV, FIRST EPSS).
Evidence
- NVD (read 2026-11-26): status Modified; published 2024-07-18; NVD Primary CVSS v3.1 9.8 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H; CISA Secondary v3.1 6.8 CVSS:3.1/AV:A/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H and v4 5.7 (AV:L); CWE-1392 (CISA), NVD-CWE-Other (NVD); CPE philips vue_pacs before 12.2.8.410. — NVD
- CISA ICSMA-24-200-01 (initial 2024-07-18, Update A 2024-11-21 removed vulnerabilities after Philips analysis), CSAF read 2026-11-26: Vue PACS <12.2.8.410; remaining CVEs CVE-2021-28165 and CVE-2023-40704; Philips recommends no action for CVE-2023-40704 but customers can request database password update; 12.2.8.410 released October 2023; not exploitable remotely; credit TAS Health NZ, Camiel van Es; sector Healthcare and Public Health. — CISA ICSMA-24-200-01
- Philips product security 2024 page (read 2026-11-26): Philips VuePACS advisory, publication 2024-07-18, update 2024-11-26; versions prior to 12.2.8.410; for CVE-2023-40704 no action needed due to low risk of exploitability, customers can request that Philips update database password(s). — Philips
- CVE record (checked 2026-11-26): PUBLISHED, CNA icscert, published 2024-07-18, affected < 12.2.8.410, CNA CVSS v3.1 6.8 and v4 5.7. — CVE.org
- Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-26. — CISA KEV
- EPSS 0.00534, percentile 0.38595, dataset 2026-11-25, read 2026-11-26. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Do Vue PACS customers need to do anything about CVE-2023-40704?
Philips says no action is needed due to low risk of exploitability, but customers can request that Philips update the database password(s). Requesting the change and keeping the PACS on a segmented network removes the default-password risk.
Which versions are affected?
Vue PACS versions before 12.2.8.410, according to CISA, Philips, the CVE record and NVD.
Can it be exploited over the internet?
CISA says the vulnerabilities in ICSMA-24-200-01 are not exploitable remotely and scores CVE-2023-40704 with an adjacent-network vector (6.8). NVD's 9.8 uses a network vector.
Is CVE-2023-40704 in the CISA KEV catalog?
No, not as of 2026-11-26 (catalog version 2026.11.25). FIRST EPSS rates it 0.00534 (38.6th percentile) in the 2026-11-25 dataset.
Questions about this page?