CVE-2025-64778 is a hard-coded credentials flaw in Mirion Medical NMIS/BioDose V22.02 and earlier, nuclear medicine tracking software: the program binaries contain plain-text passwords that could give unauthorized access to the application and its database. NVD scores it 7.8 (local access needed). The fix is V23.0 or later.
Vulnerability summary
CVSS base score
7.8
High severity
EPSS score (FIRST)
0.2%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2025-64778 · Source: NVD
Affected components
| Component | CPE | Version range |
|---|---|---|
| Mirion Medical EC2 Software NMIS/BioDose (nuclear medicine inventory and dose tracking) | cpe:2.3:a:mirion:biodose\/nmis:*:*:*:*:*:*:*:* | V22.02 and earlier (all versions before 23.0) per CISA ICSMA-25-336-01 and NVD; fixed in V23.0 or later |
What does this mean for affected devices?
CVE-2025-64778 is a use of hard-coded credentials (CWE-798) in Mirion Medical's EC2 Software NMIS/BioDose. NVD's description reads: "NMIS/BioDose software V22.02 and previous versions contain executable binaries with plain text hard-coded passwords. These hard-coded passwords could allow unauthorized access to both the application and database" (NVD).
NVD's Primary score is CVSS v3.1 7.8 (CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H): a local attacker with a low-privileged account on a machine holding the binaries can read the passwords. CISA, the CNA, scores it 7.3 in CVSS v3.1 (availability Low) and 8.4 in CVSS v4. NVD matches the CPE cpe:2.3:a:mirion:biodose\/nmis before 23.0 (CVE record).
What is NMIS/BioDose used for?
ec2 Software, which CISA lists under Mirion Medical, describes BioDose/NMIS as tracking software that lets nuclear medicine departments "keep track of inventory, doses, patient information and billing" (ec2 Software). A database that holds patient and dose records sits on a hospital network, usually shared by several client workstations. A password compiled into a program is the same in every copy of that version, so it cannot be changed per site.
What should NMIS/BioDose sites do?
The fix in the CSAF file is: "Mirion Medical recommends users update to V23.0 or later." Customers with a support contract update through the software or contact Mirion Medical support. Because the passwords sat in plain text in the program files, treat them as known: after upgrading, confirm that the database and application accounts no longer accept the old credentials, restrict who can log on to client workstations and the server, and keep the database off networks outside the department, following CISA's standard practices in the same advisory.
What else is in ICSMA-25-336-01?
CISA's medical advisory ICSMA-25-336-01, published on 2025-12-02 and credited to Joe Dillon, lists five findings in NMIS/BioDose versions before 23.0: CVE-2025-61940 (client-side authentication, CWE-603), CVE-2025-64778 (hard-coded passwords, CWE-798), and three incorrect permission issues (CWE-732): CVE-2025-62575 (the default SQL account has the sysadmin role), CVE-2025-64298 (the embedded SQL Server Express directory is exposed in the Windows share used by networked clients) and CVE-2025-64642 (installation directories that let client workstation users modify executables). The sector is Healthcare and Public Health, deployed worldwide. The client-side authentication finding has its own page, CVE-2025-61940. The hard-coded credentials fix guide explains how to remove embedded secrets and provision per-site credentials, and the FDA postmarket cybersecurity guide covers manufacturer obligations for fixes like this one.
How urgent is it?
CVE-2025-64778 is not in the CISA Known Exploited Vulnerabilities catalog as of 2026-11-26 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.00194, the 5.2th percentile, in the 2026-11-25 dataset (CISA KEV, FIRST EPSS).
Evidence
- NMIS/BioDose V22.02 and earlier contain executable binaries with plain-text hard-coded passwords allowing unauthorized access to the application and database. — NVD
- NVD Primary CVSS v3.1 7.8, CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H; CISA CNA 7.3 (v3.1) and 8.4 (v4). — NVD
- Mirion Medical recommends updating to V23.0 or later; sector Healthcare and Public Health. — CISA CSAF ICSMA-25-336-01
- Not in CISA KEV as of 2026-11-26; EPSS 0.00194 (5.2th percentile) in the 2026-11-25 dataset. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which NMIS/BioDose version fixes CVE-2025-64778?
V23.0 or later. CISA ICSMA-25-336-01 lists all versions before 23.0 as affected, and Mirion Medical recommends updating to V23.0 or later (per the CISA CSAF and the CVE record).
Does an attacker need network access to exploit CVE-2025-64778?
NVD's Primary CVSS v3.1 vector uses a local attack vector with low privileges (AV:L/PR:L): the passwords are read from the executable binaries on a machine where the software is installed. They could then be used against the application and the database (per NVD).
How do NVD and CISA score CVE-2025-64778?
NVD's Primary CVSS v3.1 score is 7.8 with availability impact High. CISA, the CNA, scores 7.3 in CVSS v3.1 with availability Low, and 8.4 in CVSS v4 (per NVD).
Is CVE-2025-64778 exploited in the wild?
It is not in the CISA KEV catalog as of 2026-11-26 (catalog version 2026.11.25). FIRST EPSS gives it 0.00194, the 5.2th percentile, in the 2026-11-25 dataset (per CISA KEV and FIRST EPSS).
Questions about this page?