CVE-2025-61940 is a client-side authentication flaw in Mirion Medical NMIS/BioDose V22.02 and earlier, nuclear medicine tracking software: the login check runs in the client, while every client uses one shared SQL Server account with full database access. NVD scores it 8.8. The fix is V23.0, which adds Windows authentication to the database.
Vulnerability summary
CVSS base score
8.8
High severity
EPSS score (FIRST)
0.7%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2025-61940 · Source: NVD
Affected components
| Component | CPE | Version range |
|---|---|---|
| Mirion Medical EC2 Software NMIS/BioDose (nuclear medicine inventory and dose tracking) | cpe:2.3:a:mirion:biodose\/nmis:*:*:*:*:*:*:*:* | V22.02 and earlier (all versions before 23.0) per CISA ICSMA-25-336-01 and NVD; fixed in V23.0 or later |
What does this mean for affected devices?
CVE-2025-61940 is a use of client-side authentication (CWE-603) in Mirion Medical's EC2 Software NMIS/BioDose. NVD's description reads: "NMIS/BioDose V22.02 and previous versions rely on a common SQL Server user account to access data in the database. User access in the client application is restricted by a password authentication check in the client software but the underlying database connection always has access." The login screen protects the application, not the data: anyone who can reach the database with the shared account bypasses it (NVD).
NVD's Primary score is CVSS v3.1 8.8 (CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H), with the CPE cpe:2.3:a:mirion:biodose\/nmis before 23.0. CISA, the CNA, scores it 8.3 in CVSS v3.1 (availability Low rather than High) and 8.7 in CVSS v4 (CVE record).
What is NMIS/BioDose used for?
ec2 Software, which CISA lists under Mirion Medical, describes BioDose/NMIS as tracking software that lets nuclear medicine departments "keep track of inventory, doses, patient information and billing" (ec2 Software). A database that holds patient and dose records sits on a hospital network, usually shared by several client workstations.
How does V23.0 fix CVE-2025-61940?
NVD's description says the latest version "introduces an option to use Windows user authentication with the database, which would restrict this database connection." The fix in the CSAF file is: "Mirion Medical recommends users update to V23.0 or later." Customers with a support contract update through the software or contact Mirion Medical support. After upgrading, switch the database connection to Windows authentication so that each user's database rights follow their own account, and remove the shared SQL login.
What else is in ICSMA-25-336-01?
CISA's medical advisory ICSMA-25-336-01, published on 2025-12-02 and credited to Joe Dillon, lists five findings in NMIS/BioDose versions before 23.0: CVE-2025-61940 (client-side authentication, CWE-603), CVE-2025-64778 (hard-coded passwords, CWE-798), and three incorrect permission issues (CWE-732): CVE-2025-62575 (the default SQL account has the sysadmin role), CVE-2025-64298 (the embedded SQL Server Express directory is exposed in the Windows share used by networked clients) and CVE-2025-64642 (installation directories that let client workstation users modify executables). The sector is Healthcare and Public Health, deployed worldwide. The hard-coded password finding has its own page, CVE-2025-64778. The authentication bypass fix guide explains why authorisation must be enforced on the server, and the FDA postmarket cybersecurity guide covers how manufacturers handle such fixes. Another ICS medical advisory with credential exposure is CVE-2025-54156 in Sante PACS Server.
How urgent is it?
CVE-2025-61940 is not in the CISA Known Exploited Vulnerabilities catalog as of 2026-11-26 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.00673, the 47.2th percentile, in the 2026-11-25 dataset (CISA KEV, FIRST EPSS).
Evidence
- NMIS/BioDose V22.02 and earlier rely on a common SQL Server account; user access is restricted only by a client-side password check. — NVD
- NVD Primary CVSS v3.1 8.8, CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H; CISA CNA 8.3 (v3.1) and 8.7 (v4). — NVD
- Mirion Medical recommends updating to V23.0 or later; sector Healthcare and Public Health. — CISA CSAF ICSMA-25-336-01
- BioDose/NMIS tracks inventory, doses, patient information and billing for nuclear medicine departments. — ec2 Software
- Not in CISA KEV as of 2026-11-26; EPSS 0.00673 (47.2th percentile) in the 2026-11-25 dataset. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which NMIS/BioDose version fixes CVE-2025-61940?
V23.0 or later. CISA ICSMA-25-336-01 lists all versions before 23.0 as affected, and NVD notes that the latest version adds an option to use Windows user authentication for the database connection (per the CISA CSAF and NVD).
Why is a client-side password check a problem?
In NMIS/BioDose V22.02 and earlier the password check happens in the client application, but every client connects to SQL Server with a common account that always has access, so anyone who reaches the database with that account bypasses the application login (per NVD).
How do NVD and CISA score CVE-2025-61940?
NVD's Primary CVSS v3.1 score is 8.8 with availability impact High. CISA, the CNA, scores 8.3 in CVSS v3.1 with availability Low, and 8.7 in CVSS v4 (per NVD).
Is CVE-2025-61940 exploited in the wild?
It is not in the CISA KEV catalog as of 2026-11-26 (catalog version 2026.11.25). FIRST EPSS gives it 0.00673, the 47.2th percentile, in the 2026-11-25 dataset (per CISA KEV and FIRST EPSS).
Questions about this page?