CVE-2022-25251 is a missing authentication check in PTC's Axeda agent and Axeda Desktop Server, remote-service software built into medical, IoT and embedded devices. A remote attacker who reaches a specific port can read and change the agent's configuration without logging in. NVD and CISA both score it 9.8. PTC's fix is agent 6.9.2 or 6.9.3 set to listen on localhost only.
Vulnerability summary
CVSS base score
9.8
Critical severity
EPSS score (FIRST)
1.6%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2022-25251 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| PTC Axeda agent | cpe:2.3:a:ptc:axeda_agent:*:*:*:*:*:*:*:* | All versions (CISA ICSA-22-067-01, CVE.org); NVD: before 6.9.1. PTC recommends 6.9.2 build 1049 or 6.9.3 build 1051 with loopback-only configuration (6.9.1 or later required) |
| PTC Axeda Desktop Server for Windows | cpe:2.3:a:ptc:axeda_desktop_server:*:*:*:*:*:windows:*:* | All versions (CISA, CVE.org); NVD: before 6.9.215. PTC recommends Version 6.9 build 215 |
What does this mean for affected devices?
The CVE record from CISA says: "When connecting to a certain port Axeda agent (All versions) and Axeda Desktop Server for Windows (All versions) may allow an attacker to send certain XML messages to a specific port without proper authentication. Successful exploitation of this vulnerability could allow a remote unauthenticated attacker to read and modify the affected product's configuration." NVD and CISA both map it to CWE-306, missing authentication for a critical function, and both score it 9.8 with CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H. NVD published the record on 2022-03-16; its configuration marks Axeda agent before 6.9.1 and Axeda Desktop Server for Windows before 6.9.215 (NVD, CVE record).
Which devices carry the Axeda agent?
CISA's ICSA-22-067-01, released on 2022-03-08 and updated to Update C on 2022-03-31, covers all versions of both products. The CISA advisory page states the flaws can affect medical, IoT and embedded devices that depend on Axeda, and links advisories from the device makers that ship it: Accuray, Agilent, BD, Bayer, Beckman Coulter, Elekta, General Electric, Hologic, Roche Diagnostics, Smiths Medical and Varian. Update C added Hologic. So the practical question for a hospital or plant is whether any imaging, laboratory or therapy system on its network runs an Axeda remote-service agent; the device maker's bulletin then gives the model-specific fix. CyberMDX and Vedere Labs reported the flaws to PTC.
What does PTC recommend?
Per the CISA CSAF, PTC recommends Axeda agent 6.9.2 build 1049 or 6.9.3 build 1051, configured so the agent and Axeda Desktop Server listen only on the local host interface; that loopback-only mode needs 6.9.1 or later. PTC also says to set a unique Desktop Server password per unit, never use the ERemoteServer component in production and delete it from the device, remove leftover installation packages, allow ERemoteServer connections only from trusted hosts, and upgrade Axeda Desktop Server to 6.9 build 215. Where a device maker controls the software, apply its bulletin and keep the device on a segmented network with no inbound internet access. See the authentication bypass fix guide, the FDA postmarket guidance and the IMDRF legacy device guide for devices that cannot be updated.
Which other CVEs are in the advisory?
Six siblings, all in the same products: CVE-2022-25246 (hard-coded credentials, 9.8), CVE-2022-25247 (missing authentication, 9.8), CVE-2022-25248 (information exposure, 5.3), CVE-2022-25249 (path traversal, 7.5), CVE-2022-25250 (missing authentication, 7.5) and CVE-2022-25252 (improper handling of exceptional conditions, 7.5). Another medical-device flaw without authentication is CVE-2023-30560 (BD Alaris PCU).
Is CVE-2022-25251 exploited?
CVE-2022-25251 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.01612, the 76.33rd percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).


Evidence
- NVD (read 2026-11-27): status Modified; published 2022-03-16; NVD Primary CVSS v3.1 9.8 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, ICS-CERT Secondary 9.8; CWE-306 (NVD and ICS-CERT); CPEs axeda_agent before 6.9.1, axeda_desktop_server (Windows) before 6.9.215. — NVD
- CISA ICSA-22-067-01 CSAF (released 2022-03-08, Update C 2022-03-31, read 2026-11-27): Axeda agent and Axeda Desktop Server for Windows, all versions; CVE-2022-25251 CWE-306 9.8; PTC: agent 6.9.2 build 1049 or 6.9.3 build 1051, loopback-only listening (6.9.1+), unique password per unit, no ERemoteServer in production, ADS 6.9 build 215; device maker links Accuray, Agilent, BD, Bayer, Beckman Coulter, Elekta, GE, Hologic, Roche Diagnostics, Smiths Medical, Varian; siblings CVE-2022-25246/25247/25248/25249/25250/25252; reported by CyberMDX and Vedere Labs; sectors Multiple, worldwide, HQ United States. — CISA ICSA-22-067-01
- CISA advisory page (read 2026-11-27): Update C adds Hologic; states the vulnerabilities can affect medical, IoT and embedded devices dependent on the affected product. — CISA
- CVE record (read 2026-11-27): state PUBLISHED, CNA ICS-CERT, published 2022-03-16; affected Axeda agent and Axeda Desktop Server for Windows, all versions; CNA CVSS v3.1 9.8; solution text matches the CISA remediations. — CVE.org
- Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
- EPSS 0.01612, percentile 0.76328, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which Axeda version fixes CVE-2022-25251?
PTC recommends Axeda agent 6.9.2 build 1049 or 6.9.3 build 1051, configured to listen only on the local host interface (available from 6.9.1), and Axeda Desktop Server 6.9 build 215 (CISA ICSA-22-067-01).
How do I know if my devices use the Axeda agent?
Check the bulletins of the device makers CISA lists: Accuray, Agilent, BD, Bayer, Beckman Coulter, Elekta, General Electric, Hologic, Roche Diagnostics, Smiths Medical and Varian. Each names its affected models and the fix for them.
What can I do before a device maker ships an update?
Keep the device on a segmented network with no inbound internet access, allow remote-service connections only from trusted hosts, and ask the maker whether the agent can be set to listen on localhost only.
Is CVE-2022-25251 in the CISA KEV catalog?
No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.01612 (76.33rd percentile) in the 2026-11-26 dataset.
Questions about this page?