Security Advisory

CVE-2021-27410: Welch Allyn Device Tools Out-of-Bounds Write

Last updated:
  • CVE-2021-27410

CVE-2021-27410 is an out-of-bounds write (CWE-787) in Hillrom Welch Allyn vital signs monitors and the software that manages them, including the Connex Spot Monitor, Connex Vital Signs Monitor and Connex Central Station. It may corrupt data or let an attacker run code. NVD scores it 9.8 and CISA 5.9. Hillrom released fixed versions for every affected product.

Vulnerability summary

CVSS base score

9.8

Critical severity

EPSS score (FIRST)

1.9%

Probability of exploitation in the wild in the next 30 days

CISA KEV catalog

No

Not listed

Published

Vulnerability ID: CVE-2021-27410 · Source: NVD

CVSS v3.1 severity rating scale: None 0.0, Low 0.1 to 3.9, Medium 4.0 to 6.9, High 7.0 to 8.9, Critical 9.0 to 10.0
How a CVSS v3.1 base score maps to a severity rating. Source: FIRST CVSS v3.1 specification, section 5.

Affected components

ComponentCPEVersion range
Welch Allyn Connex Spot Monitor (CSM)cpe:2.3:a:hillrom:connex_spot_monitor:*:*:*:*:*:*:*:*before 1.52; fix 1.52
Welch Allyn Connex Vital Signs Monitor (CVSM)cpe:2.3:a:hillrom:connex_vital_signs_monitor:*:*:*:*:*:*:*:*before 2.43.02; fix 2.43.02
Welch Allyn Connex Integrated Wall System (CIWS)cpe:2.3:a:hillrom:connex_integrated_wall_system:*:*:*:*:*:*:*:*before 2.43.02; fix 2.43.02
Welch Allyn Connex Central Station (CS)cpe:2.3:a:hillrom:connex_central_station:*:*:*:*:*:*:*:*CISA Update B: before 1.8.4 Service Pack 01, fixed in 1.8.4 SP01 (released November 2022); NVD CPE: before 1.8.6
Welch Allyn Spot Vital Signs 4400 / Spot 4400 Vital Signs Extended Care Devicecpe:2.3:a:hillrom:spot_vital_signs_4400:*:*:*:*:-:*:*:*before 1.11.00; fix 1.11.00 (released October 2021)
Welch Allyn Connex Device Integration Suite – Network Connectivity Engine (NCE)cpe:2.3:a:hillrom:connex_device_integration_suite_network_connectivity_engine:*:*:*:*:*:*:*:*before 5.3; fix 5.3 (released September 2021)
Welch Allyn Service Tool, Service Monitor and Software Development Kitcpe:2.3:a:hillrom:service_tool:*:*:*:*:*:*:*:*Service Tool before 1.10, Service Monitor before 1.7.0.0, SDK before 3.2; fixes 1.10, 1.7.0.0 and 3.2

What does this mean for affected devices?

The CVE record from CISA says the product "is vulnerable to an out-of-bounds write, which may result in corruption of data or code execution on the Welch Allyn medical device management tools". NVD's Primary score is CVSS v3.1 9.8 (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H). CISA's advisory scores it much lower, CVSS v3.0 5.9 (AV:A/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:L): the attacker must be on an adjacent network, needs a low-privilege foothold, and CISA notes the flaws have high attack complexity. Both classify it as CWE-787, out-of-bounds write. NVD published the record on 2021-06-11 (NVD, CVE record).

Which Welch Allyn products are affected?

CISA's medical advisory ICSMA-21-152-01 was released on 2021-06-01 and last updated (Update B) on 2021-12-14. It covers nine Hillrom Welch Allyn products in the Healthcare and Public Health sector, deployed worldwide. The devices are the Connex Spot Monitor, Connex Vital Signs Monitor, Connex Integrated Wall System and Spot Vital Signs 4400. The software is the Connex Central Station, the Connex Device Integration Suite Network Connectivity Engine (NCE), the Service Tool, the Service Monitor and the Welch Allyn SDK. The same advisory has a second flaw, an out-of-bounds read (CVE-2021-27408, CVSS 5.9). Itamar Cohen-Matalon of Medigate Research Labs reported both to Hillrom. CISA reported no known public exploits.

How do you fix CVE-2021-27410?

Hillrom recommends upgrading to the latest version of each product. The fixed versions in CISA's advisory are:

  • Connex Spot Monitor 1.52; Connex Vital Signs Monitor and Connex Integrated Wall System 2.43.02;
  • Spot Vital Signs 4400 and the Extended Care Device 1.11.00 (released October 2021);
  • Connex Central Station 1.8.4 Service Pack 01 (released November 2022);
  • NCE 5.3 (released September 2021), Service Tool 1.10, Service Monitor 1.7.0.0 and SDK 3.2.

NVD's CPE configuration still names Central Station versions before 1.8.6, which differs from CISA's Update B; follow the vendor's release notes. Until devices are updated, Hillrom advises network and physical security controls, authentication for server access, and data execution prevention where applicable. Hillrom's disclosure page now redirects to Baxter's product security page, which hosts the Welch Allyn bulletins.

Other medical device flaws on this site are CVE-2023-30560 in BD Alaris and CVE-2023-0888 in a B. Braun pump module. The memory corruption fix guide covers the root cause, and the FDA postmarket cybersecurity guide explains how manufacturers handle fixes like these.

Is CVE-2021-27410 exploited?

CVE-2021-27410 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.01859, the 79.75th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS).

Three signals on every advisory: CVSS says how severe the flaw is, EPSS gives the probability of exploitation in the wild in the next 30 days, and a CISA KEV listing means the flaw has been exploited in the wild
CVSS rates severity, EPSS estimates the likelihood of exploitation, and CISA KEV records exploitation seen in the wild. Source: FIRST EPSS; CISA KEV catalog.
How to fix this class of issueThis advisory's weakness class is Memory corruption. Our practical guide covers how to detect, fix, verify and disclose it: How to fix memory corruption in embedded firmware.
From advisory to verified fix in four steps: read the advisory, find the affected component in the firmware SBOM, apply the fixed version or vendor mitigation, then re-scan the image to confirm the vulnerable version is gone
Read the advisory, find the component in your firmware SBOM, apply the fix, then re-scan to confirm it.

Evidence

  • NVD (read 2026-11-27): status Modified; published 2021-06-11; NVD Primary CVSS v3.1 9.8 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H; NVD Primary CVSS v2 7.5; CWE-787 from ICS-CERT; CPEs hillrom connex_central_station <1.8.6, connex_device_integration_suite_network_connectivity_engine <5.3, connex_integrated_wall_system <2.43.02, connex_spot_monitor <1.52, connex_vital_signs_monitor <2.43.02, service_monitor <1.7.0.0, service_tool <1.10, software_development_kit <3.2, spot_vital_signs_4400 <1.11.00 (standard and extended_care). — NVD
  • CISA ICSMA-21-152-01 Update B CSAF (initial 2021-06-01, current 2021-12-14, read 2026-11-27): CVE-2021-27410 CWE-787 CVSS v3.0 5.9 AV:A/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:L; sibling CVE-2021-27408 CWE-125 5.9; fixed versions: Service Tool 1.10, SDK 3.2, Central Station 1.8.4 SP01 (released November 2022), NCE 5.3 (September 2021), Spot 4400 1.11.00 (October 2021), Service Monitor 1.7.0.0, CVSM 2.43.02, CIWS 2.43.02, CSM 1.52; mitigations: network and physical security controls, authentication for server access, DEP; reported by Itamar Cohen-Matalon, Medigate Research Labs; Healthcare and Public Health; worldwide; HQ United States; no known public exploits, high attack complexity. — CISA ICSMA-21-152-01
  • CVE record (checked 2026-11-27): state PUBLISHED, CNA ICS-CERT, published 2021-06-11; the record carries no CNA CVSS metric; affected product Welch Allyn medical device management tools; reference is the CISA advisory. — CVE.org
  • www.hillrom.com/en/responsible-disclosures/ redirects to Baxter's product security page, which lists Baxter and Welch Allyn bulletins with PDFs hosted under the hillrom asset library (checked 2026-11-27). — Baxter product security
  • Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
  • EPSS 0.01859, percentile 0.79745, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories

Sources

  1. NVD
  2. CISA ICSMA-21-152-01: Hillrom Medical Device Management (Update B) (CSAF)
  3. CISA ICSMA-21-152-01 advisory page
  4. Baxter product security (Hillrom disclosures)
  5. CVE.org record CVE-2021-27410
  6. CISA KEV
  7. FIRST EPSS

Frequently asked questions

Which versions fix CVE-2021-27410?

Per CISA ICSMA-21-152-01 Update B: Connex Spot Monitor 1.52, Connex Vital Signs Monitor and Integrated Wall System 2.43.02, Spot Vital Signs 4400 1.11.00, Connex Central Station 1.8.4 SP01, NCE 5.3, Service Tool 1.10, Service Monitor 1.7.0.0 and SDK 3.2.

Why does NVD score CVE-2021-27410 9.8 while CISA gives 5.9?

NVD assumes network access, no privileges and full impact. CISA assumes an adjacent-network attacker with low privileges and high attack complexity. Hospitals that keep these devices on segmented clinical networks are closer to CISA's case.

What can hospitals do before updating?

Hillrom advises network and physical security controls, authentication for server access, and data execution prevention where the platform supports it.

Is CVE-2021-27410 in the CISA KEV catalog?

No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.01859 (79.75th percentile) in the 2026-11-26 dataset.