CVE-2019-10950 is an unauthenticated telnet service on Fujifilm CR-IR 357 computed radiography readers (FCR Capsula X, FCR Carbon X and FCR XC-2), which can give a network attacker access to the underlying operating system. NVD and CISA both score it 9.8. No firmware fix is listed; Fujifilm's Secure Host setting and network segmentation are the mitigations.
Vulnerability summary
CVSS base score
9.8
Critical severity
EPSS score (FIRST)
3.6%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2019-10950 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| Fujifilm FCR Carbon X (CR-IR 357) | cpe:2.3:o:fujifilm:cr-ir_357_fcr_carbon_x_firmware:-:*:*:*:*:*:*:* | All firmware (NVD CPE without version); no firmware fix listed, Fujifilm Secure Host configuration and network segmentation (CISA ICSMA-19-113-01) |
| Fujifilm FCR XC-2 (CR-IR 357) | cpe:2.3:o:fujifilm:cr-ir_357_fcr_xc-2_firmware:-:*:*:*:*:*:*:* | All firmware (NVD CPE without version); no firmware fix listed, Fujifilm Secure Host configuration and network segmentation (CISA ICSMA-19-113-01) |
| Fujifilm FCR Capsula X (CR-IR 357) | cpe:2.3:o:fujifilm:cr-ir_357_fcr_capsula_x_firmware:-:*:*:*:*:*:*:* | All firmware (NVD CPE without version); no firmware fix listed, Fujifilm Secure Host configuration and network segmentation (CISA ICSMA-19-113-01) |
What does this mean for affected devices?
CVE-2019-10950 is a missing authentication flaw in Fujifilm's CR-IR 357 computed radiography reader units, sold as FCR Capsula X, FCR Carbon X and FCR XC-2. The CVE record from ICS-CERT says the devices "provide insecure telnet services that lack authentication requirements" and that an attacker who exploits this "may be able to access the underlying operating system." NVD's Primary score is 9.8, CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, with a CVSS 2.0 score of 10.0. NVD classifies the weakness as CWE-306, missing authentication for a critical function; ICS-CERT mapped it to CWE-284, improper access control (NVD, CVE record).
Which Fujifilm devices are affected?
CISA's medical advisory ICSMA-19-113-01, released on 2019-04-23 (the CSAF copy has a single revision with that date), lists the CR-IR 357 model in three products: FCR XC-2, FCR Capsula X and FCR Carbon X. NVD's CPE configuration names the firmware of CR-IR 357 FCR Carbon X, CR-IR 357 FCR XC-2 and CR-IR 357 FCR Capsula X with no version boundary, so every firmware release of those units is treated as affected. The products serve the Healthcare and Public Health sector, are deployed worldwide, and the vendor is headquartered in Japan. Marc Ruef and Rocco Gagliardi of scip AG reported the flaws.
What is the impact on an imaging department?
CISA's risk evaluation says attackers "could gain unauthorized access to the underlying operating system, allowing arbitrary code execution", and that the advisory's flaws could cause a denial of service in the cassette reader units, with potential image loss or device unavailability. Because the telnet service asks for no credentials, anyone who can reach the reader on the network can reach that service.
How do you fix CVE-2019-10950?
CISA lists no firmware update for this advisory; the remedy is configuration and network control. Fujifilm stated that the CR-IR 357 can be configured with what it calls Secure Host functionality, which makes the reader ignore all network traffic except from the IP address of the Fujifilm image acquisition console. The trade-off is that only one acquisition console can then use the reader unit. Sites that do not share a reader between consoles can ask Fujifilm to enable Secure Host; sites that do share one should contact Fujifilm to discuss options. Fujifilm also recommends securing the network as the first line of defense: only authorised devices and staff on it, public or guest networks segmented or separated by VLAN, plus administrative and technical controls. The authentication bypass fix guide covers why unauthenticated maintenance services should not ship enabled, and the IMDRF legacy medical device guide covers running imaging equipment that will not get a code fix. Imaging networks are also covered in the DICOM and PACS security guide.
What else is in ICSMA-19-113-01?
The same advisory covers CVE-2019-10948, uncontrolled resource consumption (CWE-400): a flood of TCP packets can put the reader into a denial-of-service state that requires a manual reboot. CISA scores it 7.5. Another root-level telnet exposure on this site is the hard-coded telnet credential in Secomea GateManager, CVE-2020-14510.
Do NVD and CISA score CVE-2019-10950 the same?
Yes on the number: CISA's advisory gives 9.8 in CVSS 3.0 with the same AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H metrics as NVD's CVSS 3.1 score. The CVE record itself carries no CVSS score. They differ only on the CWE (NVD CWE-306, ICS-CERT CWE-284).
Is CVE-2019-10950 exploited?
It is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25). FIRST EPSS gives it 0.03633, the 90.32nd percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS). At publication CISA reported no known public exploits specifically targeting these vulnerabilities.


Evidence
- NVD (read 2026-11-27): status Modified; published 2019-04-30; [email protected] Primary CVSS 3.1 9.8 AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, CVSS 2.0 10.0; NVD CWE-306 (Primary), ICS-CERT CWE-284 (Secondary); CPEs cr-ir_357_fcr_carbon_x, fcr_xc-2, fcr_capsula_x firmware without version. — NVD
- CVE record (CNA ICS-CERT, published 2019-04-30): insecure telnet services that lack authentication requirements; attacker may access the underlying operating system; CWE-284; no CVSS in the record. — CVE.org
- CISA ICSMA-19-113-01 (CSAF, initial release 2019-04-23, one revision): FCR XC-2, FCR Capsula X, FCR Carbon X CR-IR 357; CVE-2019-10950 CWE-284 CVSS 3.0 9.8; CVE-2019-10948 CWE-400 7.5 (TCP overflow DoS, manual reboot); risk: DoS in cassette reader units, image loss, OS access and arbitrary code execution; worldwide, HQ Japan; reported by Marc Ruef and Rocco Gagliardi (scip AG); Fujifilm Secure Host functionality, network securing, segmentation/VLAN; no known public exploits. — CISA CSAF
- CISA KEV catalog version 2026.11.25, 1,805 entries: CVE-2019-10950 not listed; checked 2026-11-27. — CISA KEV
- EPSS 0.03633, percentile 0.90316, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which Fujifilm devices are affected by CVE-2019-10950?
The CR-IR 357 reader units in FCR Capsula X, FCR Carbon X and FCR XC-2. NVD's CPEs carry no version boundary, so all firmware of these units is treated as affected (per CISA ICSMA-19-113-01 and NVD).
Is there a firmware update for CVE-2019-10950?
CISA lists none. Fujifilm's mitigation is its Secure Host configuration, which makes the reader ignore all traffic except from the image acquisition console's IP address, combined with securing and segmenting the network (per CISA ICSMA-19-113-01).
What does Secure Host cost in practice?
With Secure Host enabled, more than one image acquisition console can no longer share the CR-IR 357 reader unit. Sites that share readers should contact Fujifilm to discuss options (per CISA ICSMA-19-113-01).
Is CVE-2019-10950 in the CISA KEV catalog?
No, not as of 2026-11-27 (catalog version 2026.11.25). FIRST EPSS rates it 0.03633, the 90.32nd percentile, in the 2026-11-26 dataset (per CISA KEV and FIRST EPSS).