CVE-2017-5149 is a missing endpoint check in the St. Jude Medical Merlin@home transmitter, the bedside unit that sends data from implanted cardiac devices to the Merlin.net site: versions before 8.2.2 do not verify who is at either end of that channel, so a man-in-the-middle could read or influence it. NVD and CISA score it 8.9. Version 8.2.2 is delivered by automatic update.
Vulnerability summary
CVSS base score
8.9
High severity
EPSS score (FIRST)
1.2%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2017-5149 · Source: NVD

Affected components
| Component | CPE | Version range |
|---|---|---|
| St. Jude Medical Merlin@home RF transmitter EX1150 | cpe:2.3:o:abbott:merlin\@home_firmware:*:*:*:*:*:*:*:* | before 8.2.2; fixed in 8.2.2 by automatic update (CISA ICSMA-17-009-01A; the NVD CPE range ends at 8.0) |
| St. Jude Medical Merlin@home inductive transmitter EX1100 (with or without MerlinOnDemand) | cpe:2.3:o:abbott:merlin\@home_firmware:*:*:*:*:*:*:*:* | before 8.2.2; fixed in 8.2.2 by automatic update (CISA ICSMA-17-009-01A) |
What does this mean for affected devices?
What is CVE-2017-5149?
NVD describes it as follows: "An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints." NVD's Primary score is CVSS v3.1 8.9 (high), vector CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H, with CVSS v2 6.8 (NVD). St. Jude Medical is now part of Abbott, and NVD's CPE uses the abbott:merlin\@home_firmware name, with a range that ends at 8.0 rather than at CISA's fixed 8.2.2.
Why does the weakness mapping differ?
CISA, the CNA, classifies the flaw as CWE-300, Channel Accessible by Non-Endpoint, which matches the description: the transmitter talks to Merlin.net without verifying the other side. NVD's own entry maps it to CWE-476, NULL Pointer Dereference, which does not fit the description; use CISA's CWE-300 for triage. Both give the same 8.9 score and vector, so the severity is not in dispute. The CVE record itself labels the problem type as a man-in-the-middle issue.
What does the Merlin@home transmitter do, and which models are affected?
CISA released ICSMA-17-009-01 on 2017-01-09 and Update A on 2017-02-06. CISA describes the product: "The Merlin@home transmitter allows for remote care management of patients with implanted cardiac devices through scheduled transmissions, patient-initiated transmissions, and daily monitoring." Its CISA CSAF record lists three model groups before version 8.2.2: the RF model EX1150, the inductive model EX1100 and the inductive EX1100 with MerlinOnDemand capability. The advisory names the Healthcare and Public Health sector and credits MedSec Holdings with identifying the flaw. Update A added that a third-party security firm had verified the mitigation and that MerlinOnDemand units are about 0.1 percent of transmitters. At release, CISA knew of no public exploits that specifically targeted the flaw.
How is CVE-2017-5149 fixed?
CISA quotes the vendor: the new transmitter software, version 8.2.2, "will be automatically updated over a period of several months, when all models of the Merlin@home transmitters are connected to an Ethernet, WiFi, cellular network, or a landline." St. Jude Medical recommends that patients keep the transmitter powered and connected at all times so that it receives this and future updates. No action on the implanted device itself is part of this fix. Clinics and hospital biomedical teams can confirm the version through the vendor's support channels.
Verifying the server certificate and pinning the expected endpoint is the class fix covered in our weak cryptography fix guide. For how regulators expect manufacturers to patch fielded devices like this, see our FDA postmarket cybersecurity guide and the IMDRF legacy medical device guide. A later home cardiac monitor flaw on this site is the Medtronic MyCareLink patient monitor empty-password account (CVE-2025-4395).
Is CVE-2017-5149 exploited?
CVE-2017-5149 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25). FIRST EPSS gives it 0.0124, the 69th percentile, in the 2026-11-26 dataset, as of 2026-11-27 (CISA KEV, FIRST EPSS).


Evidence
- St. Jude Medical Merlin@home before 8.2.2 (RF EX1150, inductive EX1100, EX1100 with MerlinOnDemand): endpoints of the channel to Merlin.net not verified, man-in-the-middle may access or influence communications; NVD Primary CVSS v3.1 8.9 AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H; v2 6.8; NVD CWE-476; CPE abbott merlin@home_firmware up to 8.0. — NVD
- ICSMA-17-009-01 released 2017-01-09, Update A 2017-02-06; CVSS v3 8.9 same vector, CWE-300; product description quoted; models before 8.2.2; 8.2.2 applied automatically over several months when connected; keep transmitters powered and connected; Healthcare and Public Health; MedSec Holdings credited; Update A: third-party verification, MerlinOnDemand about 0.1 percent; no known public exploits. — CISA ICSMA-17-009-01A
- Not in the KEV catalog, version 2026.11.25, checked 2026-11-27. — CISA KEV
- EPSS 0.0124, 69th percentile, 2026-11-26 dataset. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
How do patients get the CVE-2017-5149 fix?
Automatically. Version 8.2.2 is installed over a period of several months when the Merlin@home transmitter is connected to Ethernet, Wi-Fi, a cellular network or a landline, so the transmitter should stay powered and connected (per CISA ICSMA-17-009-01A).
Which Merlin@home models are affected?
The RF model EX1150 and the inductive model EX1100, with or without MerlinOnDemand capability, in versions before 8.2.2 (per CISA ICSMA-17-009-01A).
What could an attacker do with CVE-2017-5149?
Sit between the transmitter and the Merlin.net site and access or influence their communication, because the transmitter does not verify the endpoints. The attack complexity is rated high (per NVD and CISA).
Is CVE-2017-5149 in the CISA KEV catalog?
No. It is not listed in catalog version 2026.11.25, checked 2026-11-27, and FIRST EPSS gives it 0.0124, the 69th percentile, in the 2026-11-26 dataset (per CISA KEV and FIRST EPSS).
Questions about this page?