CVE-2023-46687 is a command injection (CWE-77) in Emerson Rosemount GC370XA, GC700XA and GC1500XA gas chromatographs, firmware 4.1.5 and prior. An unauthenticated user with network access could run arbitrary commands as root. NVD scores it 9.8 critical. Emerson has updated firmware through its GC technical support; keep the analyzers off the internet.
Vulnerability summary
CVSS base score
9.8
Critical severity
EPSS score (FIRST)
1.5%
Probability of exploitation in the wild in the next 30 days
CISA KEV catalog
No
Not listed
Published
Vulnerability ID: CVE-2023-46687 · Source: NVD
Affected components
| Component | CPE | Version range |
|---|---|---|
| Emerson Rosemount GC370XA | cpe:2.3:o:emerson:gc370xa_firmware:4.1.5:*:*:*:*:*:*:* | Version 4.1.5 and all prior revisions (Emerson notification, CISA ICSA-24-030-01); updated firmware from Emerson GC technical support |
| Emerson Rosemount GC700XA | cpe:2.3:o:emerson:gc700xa_firmware:4.1.5:*:*:*:*:*:*:* | Version 4.1.5 and all prior revisions (Emerson notification, CISA ICSA-24-030-01); updated firmware from Emerson GC technical support |
| Emerson Rosemount GC1500XA | cpe:2.3:o:emerson:gc1500xa_firmware:4.1.5:*:*:*:*:*:*:* | Version 4.1.5 and all prior revisions (Emerson notification, CISA ICSA-24-030-01); updated firmware from Emerson GC technical support |
What does this mean for affected devices?
NVD describes CVE-2023-46687 as follows: "In Emerson Rosemount GC370XA, GC700XA, and GC1500XA products, an unauthenticated user with network access could execute arbitrary commands in root context from a remote computer." The weakness is CWE-77, command injection, in both NVD's and CISA's analysis. NVD's own Primary CVSS v3.1 score is 9.8 (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H); CISA ICS, the CNA, and Emerson give the same score. NVD's CPE configuration names firmware 4.1.5 for the three models (NVD, CVE record).
Which Rosemount gas chromatographs are affected?
Emerson's cyber security notification for ICSA-24-030-01 (revision 2, 8 February 2024) lists the GC370XA, GC700XA and GC1500XA, each at version 4.1.5 and all prior revisions. These are process gas chromatographs; CISA advisory ICSA-24-030-01 (2024-01-30, revised 2025-06-10 to update CWEs) lists the Energy and Chemical sectors and worldwide deployment. The notification covers four CVEs found by Vera Mens of Claroty Research: this command injection, a login bypass through the password reset mechanism (CVE-2023-49716), an authorization flaw (CVE-2023-51761) and a second command injection that needs authentication (CVE-2023-43609). CISA says that together they could let an attacker run arbitrary commands, access sensitive information, cause a denial of service and bypass authentication to acquire admin capabilities.
How do you fix CVE-2023-46687?
Emerson recommends updating the firmware on the affected products; it states that a new software release addressing the issues is available and that users get update information from GC technical support ([email protected]). Neither Emerson nor CISA publishes the fixed version number. Emerson also advises against connecting the analyzers directly to the internet and asks users to make sure they sit on a well-protected network (CSAF file). Our Emerson DeltaV, Ovation and ROC guide covers Emerson's CISA advisories, fixed versions and hardening steps across its control systems, and the command injection fix guide explains the bug class for firmware teams. Another Emerson field device flaw on this site is the WirelessHART gateway issue CVE-2020-12030.
Is CVE-2023-46687 exploited?
It is absent from the CISA KEV catalog as of 2026-11-26 (catalog version 2026.11.25), and CISA reported no known public exploitation at publication. FIRST EPSS gives 0.01534, the 75.11th percentile, in the 2026-11-25 dataset (FIRST EPSS).
Evidence
- NVD: Modified; Primary CVSS v3.1 9.8 AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H; CISA ICS 9.8 same vector; CWE-77 (NVD and CISA); CPE emerson gc370xa/gc700xa/gc1500xa firmware 4.1.5; published 2024-02-09; read 2026-11-26. — NVD
- CVE record: PUBLISHED by CISA ICS 2024-02-09, title Emerson Rosemount GC370XA, GC700XA, GC1500XA Command Injection; affected <= 4.1.5; CVSS v3.1 9.8; credit Vera Mens of Claroty Research. — CVE.org
- ICSA-24-030-01 (2024-01-30, revised 2025-06-10 CWEs): GC370XA/GC700XA/GC1500XA <= 4.1.5; CVE-2023-46687, 49716, 51761, 43609; Energy, Chemical; Emerson recommends firmware update via [email protected], well-protected network, no internet connection; no known public exploitation reported. — CISA ICS advisory
- Emerson notification EMR.MSOL23001 rev 2, 8 Feb 2024: GC370XA/GC700XA/GC1500XA version 4.1.5 and all prior revisions; CVE-2023-46687 9.8; four CVEs found by Vera Mens of Claroty Research; new software release available, contact GC Technical support; do not connect to the internet. — Emerson
- Not in the KEV catalog, version 2026.11.25, checked 2026-11-26. — CISA KEV
- EPSS 0.01534, percentile 0.75114, dataset 2026-11-25. — FIRST EPSS

Related advisories
Sources
Frequently asked questions
Which Emerson gas chromatographs are affected by CVE-2023-46687?
Rosemount GC370XA, GC700XA and GC1500XA, version 4.1.5 and all prior revisions (per Emerson's notification and CISA ICSA-24-030-01).
How do I get the firmware fix for CVE-2023-46687?
Emerson says a new software release that addresses the issue is available and that update information comes from Emerson GC technical support at [email protected]. The fixed version number is not published (per Emerson and CISA).
What other vulnerabilities did Emerson fix with CVE-2023-46687?
Three more: CVE-2023-49716 (login bypass through password reset), CVE-2023-51761 (authorization not enforced) and CVE-2023-43609 (authenticated command injection), all reported by Claroty Research (per Emerson and CISA ICSA-24-030-01).
Is CVE-2023-46687 in the CISA KEV catalog?
No. It is absent from KEV catalog version 2026.11.25 as of 2026-11-26, and FIRST EPSS gives 0.01534 (75.11th percentile) in the 2026-11-25 dataset (per CISA KEV and FIRST EPSS).
Questions about this page?