Security Advisory

CVE-2023-1966: Illumina UCS Excess Privilege on Sequencers

Last updated:
  • CVE-2023-1966

CVE-2023-1966 is an excessive-privilege flaw in the Illumina Universal Copy Service, software that runs on Illumina DNA sequencers and array scanners such as MiSeq, NextSeq, NovaSeq and iScan, including the Dx models MiSeqDx and NextSeq 550Dx. An unauthenticated remote actor could upload and execute code at operating system level. NVD scores it 9.8; Illumina publishes a per-instrument fix table.

Vulnerability summary

CVSS base score

9.8

Critical severity

EPSS score (FIRST)

1.0%

Probability of exploitation in the wild in the next 30 days

CISA KEV catalog

No

Not listed

Published

Vulnerability ID: CVE-2023-1966 · Source: NVD

CVSS v3.1 severity rating scale: None 0.0, Low 0.1 to 3.9, Medium 4.0 to 6.9, High 7.0 to 8.9, Critical 9.0 to 10.0
How a CVSS v3.1 base score maps to a severity rating. Source: FIRST CVSS v3.1 specification, section 5.

Affected components

ComponentCPEVersion range
Illumina NextSeq 550Dx / MiSeqDx (Operating Software)cpe:2.3:o:illumina:nextseq_550dx_firmware:*:*:*:*:-:*:*:*NextSeq 550Dx Operating Software 1.0.0-1.3.1 and 1.3.3 and later; MiSeqDx Operating Software 4.0.1 and later; Control Software 4.0 in RUO mode (CISA ICSMA-23-117-01); fix: Illumina per-instrument software update
Illumina MiSeq, MiniSeq, iSeq 100, NextSeq 500/550/1000/2000, NovaSeq 6000, iScancpe:2.3:o:illumina:novaseq_6000_firmware:*:*:*:*:*:*:*:*MiSeq Control Software 4.0; MiniSeq 2.0 and later; iSeq 100 all; NextSeq 500/550 4.0; NextSeq 1000/2000 1.4.1 and earlier; NovaSeq 6000 1.8 and earlier; iScan 4.0.0 and 4.0.5 (CISA ICSMA-23-117-01); fix: configure UCS credentials or update per Illumina instructions

What does this mean for affected devices?

The CVE record from CISA says: "Instruments with Illumina Universal Copy Service v1.x and v2.x contain an unnecessary privileges vulnerability. An unauthenticated malicious actor could upload and execute code remotely at the operating system level, which could allow an attacker to change settings, configurations, software, or access sensitive data on the affected product." NVD's Primary analysis scores it 9.8 (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H) and lists CWE-250, execution with unnecessary privileges, and CWE-269, improper privilege management. CISA, the CNA, scored it 7.4 (CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H): CISA treated the attack as needing physical access and low privileges, while NVD rated it network-reachable with no privileges (NVD, CVE record).

Which Illumina instruments are affected?

CISA's medical advisory ICSMA-23-117-01, released on 2023-04-27, lists iScan Control Software 4.0.0 and 4.0.5, iSeq 100 (all versions), MiniSeq Control Software 2.0 and later, MiSeq Control Software 4.0, MiSeqDx Operating Software 4.0.1 and later, NextSeq 500/550 Control Software 4.0, NextSeq 550Dx Control Software 4.0 and Operating Software 1.0.0 to 1.3.1 and 1.3.3 and later, NextSeq 1000/2000 Control Software 1.4.1 and earlier, NovaSeq 6000 Control Software 1.7 and earlier, and NovaSeq Control Software 1.8. CISA names the Healthcare and Public Health sector and worldwide deployment, with Illumina headquartered in the United States; Illumina reported the issues itself. The same advisory carries CVE-2023-1968, binding to an unrestricted IP address (CWE-1327, CISA 10.0), which lets UCS listen on all network interfaces.

How do you fix CVE-2023-1966?

Illumina's UCS vulnerability support page gives the action per instrument and software version. In summary, read on 2026-11-27:

  • iSeq 100, MiniSeq 2.0 and later, iScan 4.0.0 and NovaSeq 6000 1.7 and earlier: configure UCS account credentials;
  • iScan 4.0.5: contact Illumina Technical Support to upgrade to v4.2.1;
  • NextSeq 1000/2000 1.4.1: update with the NextSeq 1000/2000 Control Software Suite v1.5.0 installer;
  • NovaSeq 6000 1.8: install NovaSeq Software Suite v1.8.1.59, configure UCS account credentials and close the firewall port;
  • MiSeq, MiSeqDx, NextSeq 500/550 and NextSeq 550Dx: update software per Illumina's instrument-specific instructions.

Illumina states that software versions not in its table are not impacted. CISA adds the usual network controls: keep instruments off the internet, behind firewalls and isolated from business networks. Lab and biomedical engineering teams should track this like any other device patch; the FDA postmarket cybersecurity guide covers manufacturer duties and the patch management guide the operator side. Other medical device pages on this site include CVE-2023-0888 in a B. Braun infusion pump module and CVE-2023-40150 in MedDream PACS.

Is CVE-2023-1966 exploited?

CVE-2023-1966 is not in the CISA KEV catalog as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS gives it 0.01012, the 62.23th percentile, in the 2026-11-26 dataset (CISA KEV, FIRST EPSS). CISA's advisory states that no known public exploits specifically target these vulnerabilities.

Three signals on every advisory: CVSS says how severe the flaw is, EPSS gives the probability of exploitation in the wild in the next 30 days, and a CISA KEV listing means the flaw has been exploited in the wild
CVSS rates severity, EPSS estimates the likelihood of exploitation, and CISA KEV records exploitation seen in the wild. Source: FIRST EPSS; CISA KEV catalog.
How to fix this class of issueNo dedicated SoftWSP guide covers this weakness class; the fix is usually the vendor patch or firmware update referenced above. Our related guide explains how to plan and verify it: NIST SP 800-40 Rev. 4: enterprise patch management planning, explained.
From advisory to verified fix in four steps: read the advisory, find the affected component in the firmware SBOM, apply the fixed version or vendor mitigation, then re-scan the image to confirm the vulnerable version is gone
Read the advisory, find the component in your firmware SBOM, apply the fix, then re-scan to confirm it.

Evidence

  • NVD (read 2026-11-27): status Modified; published 2023-04-28; NVD Primary CVSS v3.1 9.8 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H; ICS-CERT Secondary 7.4 CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H; CWE-250, CWE-269; CPEs for iScan, iSeq 100, MiniSeq, MiSeq, MiSeqDx, NextSeq 500/550/550Dx/1000/2000 and NovaSeq 6000 firmware. — NVD
  • CISA ICSMA-23-117-01 (page release date 2023-04-27, read 2026-11-27): UCS v1.x and v2.x on listed Illumina instruments; CVE-2023-1966 CWE-250 7.4; sibling CVE-2023-1968 CWE-1327 10.0; vendor fix: UCS Vulnerability Instructions Guide per configuration; sector Healthcare and Public Health; worldwide; HQ United States; reported by Illumina; no known public exploits. — CISA ICSMA-23-117-01
  • CVE record (checked 2026-11-27): state PUBLISHED, CNA ICS-CERT, published 2023-04-28, CNA CVSS v3.1 7.4; affected iScan, MiniSeq, iSeq 100, MiSeq and other control software versions. — CVE.org
  • Illumina support page (read 2026-11-27): per-instrument impacted versions and actions (configure UCS credentials; iScan 4.0.5 upgrade to v4.2.1; NextSeq 1000/2000 Control Software Suite v1.5.0; NovaSeq Software Suite v1.8.1.59); versions not listed are not impacted. — Illumina UCS vulnerability support page
  • Not in the CISA KEV catalog, catalog version 2026.11.25, 1,805 entries, checked 2026-11-27. — CISA KEV
  • EPSS 0.01012, percentile 0.62234, dataset 2026-11-26, read 2026-11-27. — FIRST EPSS

Related advisories

Sources

  1. NVD
  2. CISA ICSMA-23-117-01: Illumina Universal Copy Service (CSAF)
  3. CISA ICSMA-23-117-01 advisory page
  4. CVE.org record CVE-2023-1966
  5. Illumina Universal Copy Service vulnerability support page
  6. CISA KEV
  7. FIRST EPSS

Frequently asked questions

Which Illumina instruments does CVE-2023-1966 affect?

Instruments running UCS v1.x or v2.x with the listed software: iScan, iSeq 100, MiniSeq, MiSeq, MiSeqDx, NextSeq 500/550, NextSeq 550Dx, NextSeq 1000/2000 and NovaSeq 6000 (CISA ICSMA-23-117-01).

What does the fix look like?

It depends on the instrument: some only need UCS account credentials configured, others need a software update such as NextSeq 1000/2000 Control Software Suite v1.5.0 or NovaSeq Software Suite v1.8.1.59, per Illumina's UCS support page.

Why does NVD score 9.8 when CISA scores 7.4?

CISA scored the attack as needing physical access and low privileges. NVD scored it as network-reachable with no privileges, matching the description of an unauthenticated remote actor.

Is CVE-2023-1966 in the CISA KEV catalog?

No, not as of 2026-11-27 (catalog version 2026.11.25, 1,805 entries). FIRST EPSS rates it 0.01012 (62.23th percentile) in the 2026-11-26 dataset.